Julia Purtov1,2, Mareike Frensemeier1,2, Elmar Kroner2. 1. Department of Materials Science and Engineering, Saarland University , Campus D2 2, 66123 Saarbrücken, Germany. 2. INM - Leibniz Institute for New Materials , Campus D2 2, 66123 Saarbrücken, Germany.
Abstract
Suction based attachment systems for pick and place handling of fragile objects like glass plates or optical lenses are energy-consuming and noisy and fail at reduced air pressure, which is essential, e.g., in chemical and physical vapor deposition processes. Recently, an alternative approach toward reversible adhesion of sensitive objects based on bioinspired dry adhesive structures has emerged. There, the switching in adhesion is achieved by a reversible buckling of adhesive pillar structures. In this study, we demonstrate that these adhesives are capable of switching adhesion not only in ambient air conditions but also in vacuum. Our bioinspired patterned adhesive with an area of 1 cm(2) provided an adhesion force of 2.6 N ± 0.2 N in air, which was reduced to 1.9 N ± 0.2 N if measured in vacuum. Detachment was induced by buckling of the structures due to a high compressive preload and occurred, independent of air pressure, at approximately 0.9 N ± 0.1 N. The switch in adhesion was observed at a compressive preload between 5.6 and 6.0 N and was independent of air pressure. The difference between maximum adhesion force and adhesion force after buckling gives a reasonable window of operation for pick and place processes. High reversibility of the switching behavior is shown over 50 cycles in air and in vacuum, making the bioinspired switchable adhesive applicable for handling operations of fragile objects.
Suction based attachment systems for pick and place handling of fragile objects like glass plates or optical lenses are energy-consuming and noisy and fail at reduced air pressure, which is essential, e.g., in chemical and physical vapor deposition processes. Recently, an alternative approach toward reversible adhesion of sensitive objects based on bioinspired dry adhesive structures has emerged. There, the switching in adhesion is achieved by a reversible buckling of adhesive pillar structures. In this study, we demonstrate that these adhesives are capable of switching adhesion not only in ambient air conditions but also in vacuum. Our bioinspired patterned adhesive with an area of 1 cm(2) provided an adhesion force of 2.6 N ± 0.2 N in air, which was reduced to 1.9 N ± 0.2 N if measured in vacuum. Detachment was induced by buckling of the structures due to a high compressive preload and occurred, independent of air pressure, at approximately 0.9 N ± 0.1 N. The switch in adhesion was observed at a compressive preload between 5.6 and 6.0 N and was independent of air pressure. The difference between maximum adhesion force and adhesion force after buckling gives a reasonable window of operation for pick and place processes. High reversibility of the switching behavior is shown over 50 cycles in air and in vacuum, making the bioinspired switchable adhesive applicable for handling operations of fragile objects.
Animals like flies,[1] ants,[2] or beetles[3] have developed versatile attachment systems which
enable them to attach quickly and reversibly to surfaces of varying
chemistry and topography, i.e., smooth and rough surfaces. Their contact
elements are covered with millions of fine fibrils, which, often in
combination with secretions,[4] play a crucial
role in adhesion.[5] One of the most complex
and efficient adhesion systems is found in geckos, the largest known
animals with hairy attachment pads.[6,7] Their attachment
system is a “dry” system and does not rely on adhesion
enhancing secretions. Although some phospholipids were found in gecko
footprints,[8] the function of these lipids
seem to be irrelevant for adhesion. The adhesive interaction of gecko
toe pads with a surface is mainly based on van der Waals forces,[7] likely enhanced by capillary forces due to humidity.[9,10] Geckos can generate large forces, reaching a surprisingly high shear
strength of up to 100 kPa.[6] This performance
is assumed to be related to good adaptability of the hairy attachment
pads to roughness, an improved stress distribution, an increased defect
tolerance, and size effects.[11] Besides
the outstanding adhesive properties, a quick and easy release of the
adhesive pads is crucial for locomotion and, ultimately, the survival
of the gecko. Detachment is controlled by the anisotropy of the adhesive
structures and the biomechanics of the gecko’s motion, which
consists of simultaneous shear and peel movement.[12]The first systematic investigations of the adhesive
mechanisms and the interactions of gecko toe pads with a broad variety
of substrates were made in the early 20th century. Weitlaner performed
adhesion experiments with living and dead geckos to understand whether
the gecko uses a “pneumatic mechanism” for attachment.
Despite his very limited experimental equipment, he found that amputated
and shear loaded gecko feet did not lose their sticking capability
to various surfaces even at reduced air pressure. He concluded that
the, at that time assumed, “pneumatic mechanism” does
not have an essential impact on the extraordinary adhesive properties
of the gecko toe pad but may only have a minor contribution to adhesion.[13] In summary, the gecko’s adhesion system
combines the following properties: high adhesive forces, quick and
easy detachment, dry “residue-free” contact, and operational
in vacuum.Hence, it is not surprising that this attachment
system gains growing attention, not only from the scientific community
but also from industry, especially as its properties may lead to new
artificial attachment devices, which could replace current state of
the art systems such as suction cups. Consequently, artificial bioinspired
adhesive systems have been extensively studied,[11,14−17] and comparably high adhesive performance was reached, even exceeding
the so-called “gecko-limit” of 100 kPa.[18] In extension to Weitlaner’s results on the adhesion
of geckos, recent studies suggest that adhesion of (synthetic) bioinspired
surfaces relies, in addition to van der Waals interactions, to a small
part on suction.[19] It was found experimentally
that a small suction effect is present for mushroom-shaped patterned
adhesives if adhesion is tested in vacuum.[19−21] It has also
been predicted theoretically that suction effects may become relevant
in patterned surfaces as soon as a certain critical contact size is
exceeded.[22,23]Many of these bioinspired systems
have been applied to grip and release objects; most approaches function
close to the directional, shear induced adhesion found for geckos.[12] There, switching adhesion mainly relies on asymmetric
adhesive structures, which exhibit high adhesion if sheared into one
direction, while adhesion drops significantly if sheared in the opposite
direction.[24] The frequently occurring lateral
displacement of the object during attachment and detachment may be
circumvented by gripper designs, where two or more anisotropic adhesive
pads are sheared in opposite directions so that the lateral forces
cancel out.[25] Other approaches combine
electrostatic adhesion and bioinspired adhesives to maintain a compressive
preload on the adhesive structures[26] or
even use biological structures obtained from gecko toes for handling
of small objects.[27] An approach to handle
objects with a more complex geometry is based on a balloon-like gecko
adhesive tape, which can be adapted to curved surface geometries by
“inflating” and “deflating” the balloon.[28]Besides fabrication and characterization
of bioinspired adhesives with high and robust adhesion and adhesion
control using “passive” peeling or shearing, the control
of adhesion by an external stimulus[29−33] has been studied and improved to obtain switchable
adhesives even in extreme environments like outer space.[34] While the complex detachment motion works efficiently
for geckos and has already been mimicked relatively close to the natural
archetype,[35] other approaches have emerged
to switch adhesion by using external triggers. Shape memory materials,[36,37] active polymeric materials such as liquid crystal elastomers,[38] injection of liquids in subsurface microchannels,[29] application of magnetic fields to orient magnetic
structures,[39] or mechanical loading of
rubber elastic patterned samples was applied to obtain switchability.[40] The latter has been investigated in detail and
shows detachment of rubber elastic pillars due to mechanical instability
at high compressive load, leading to a preload responsive switchable
adhesive.[40−43] Due to the simplicity of the operation mode and the fast and reversible
response, this approach shows significant potential for pick and place
processes. In a more recent publication, this approach of pressure
activated switchable adhesion was extended by using structures of
different length to switch between three adhesive states, namely,
low, high, and very low adhesion.[44]To obtain pick and place handling in vacuum conditions, robust adhesion
in vacuum and reliable switchability need to be linked. In the present
study, the adhesive performance of a preload responsive, bioinspired
adhesive was tested both in vacuum and in air, and the influence of
air pressure on adhesion was quantified. Further, pick and place processes
were performed and the reversibility was demonstrated over 50 loading
cycles.
Experimental Section
Mold Preparation
Aluminum molds were fabricated using
a process similar to the one reported in earlier studies.[45,46] An array of holes with 2 mm depth, 0.4 mm width, and a center–center
spacing of 0.8 mm was milled. The geometrical parameters were chosen
to yield samples which possess a mechanical instability at high compressive
loading, known to lead to detachment.[40,42] The array
contained 203 holes and covered an area of approximately 1 cm2. The mold was thoroughly cleaned in acetone, ethanol, isopropanol,
and deionized water in an ultrasonic bath and subsequently silanized.
For this, the mold was placed together with a glass vial containing
10 μL of trichloroperfluorooctysilane (Sigma-Aldrich) into a
desiccator and evacuated to a pressure below 10 mbar for at least
45 min until the silane evaporated completely. Afterward, the mold
was kept in an oven in air at 95 °C for 2 h.
Sample Preparation
Samples were prepared from polydimethylsiloxane
(PDMS, Sylgard 184, Dow Corning) by soft molding of the previously
prepared aluminum molds. The PDMS prepolymer and cross-linker were
mixed in a 10:1 ratio, poured onto the silanized mold, and degassed
in a desiccator. The filled mold was then placed in an oven and cured
at 75 °C for 4 h. After cooling to room temperature, the PDMS
sample was carefully peeled from the mold, resulting in a PDMS array
of cylindrical pillars. The backing layer was approximately 3 mm thick.
Tip Modification
Pillars with mushroom-shaped
tips were reported to significantly increase adhesion compared to
pillars having flat or spherical tips.[18,45,47] Thus, the tips were modified using a variation of
a previously described process.[18] Briefly,
PDMS was mixed and degassed as described above. A thin metal rod was
dipped into the liquid PDMS and gently brought into contact with the
pillars, resulting in deposition of a small droplet of liquid PDMS
on the tip of each pillar. Afterward, the droplet-covered tips were
placed face-down onto a smooth, silanized glass plate (silanization
protocol as above). The tips were squashed, resulting in a flattened
mushroom-shape. The patterned sample was fully cured in an oven at
75 °C for 4 h and, after cooling to room temperature, carefully
removed from the glass plate. Three samples were chosen for adhesion
experiments; an exemplary sample is shown in Figure a.
Figure 1
(a) The photograph shows an exemplary bioinspired
switchable dry adhesive PDMS sample with an array of mushroom-shaped
pillars. The inset exhibits a side view of a single mushroom shaped
tip. (b) The experimental adhesion tester setup is built in a vacuum
oven for experiments at ambient air pressure and at low pressure condition
(<10 mbar).
(a) The photograph shows an exemplary bioinspired
switchable dry adhesive PDMS sample with an array of mushroom-shaped
pillars. The inset exhibits a side view of a single mushroom shaped
tip. (b) The experimental adhesion tester setup is built in a vacuum
oven for experiments at ambient air pressure and at low pressure condition
(<10 mbar).
Adhesion
Testing Setup
An adhesion measurement setup as shown in Figure b, inspired by the
macroscopic adhesion measurement device (MAD),[48] was built in a vacuum oven. It consisted of a linear z-positioning system and a load-cell based force measurement.
Each patterned PDMS sample was fixated to a glass backing by applying
oxygen plasma to the backside of the sample and bringing it into contact
with the smooth cleaned glass plate. The sample was mounted to a load
cell with a stiffness of >100 kN/m. Prior to the adhesion measurements,
the sample was aligned using a manual alignment stage and applying
the alignment process published by Kroner et al.[49] Force and displacement were recorded during all measurements.
Adhesion Measurements
All experiments were
performed with a testing velocity of 80 μm/s. Each sample was
tested at least three times and thereby rotated by ∼120°
along the vertical axis in between the experiments to avoid misalignment.
Adhesion measurements were conducted in ambient atmosphere, called
“air” in the subsequent text, and at reduced pressure
<10 mbar, called “vacuum” in the subsequent text.
A glass plate was used as test substrate which was, depending on the
testing mode, either fixated or loosened. To obtain an equilibrium
surface state, more than 300 contacts were made between a smooth PDMS
sample and the glass plate prior to adhesion measurements.[50,51] The experimental error for all adhesion measurements was ±0.1
N.Different phases occurred during an adhesion experiment with
preload responsive dry-adhesive samples, which are schematically shown
in Figure a and can
be described as follows:
Figure 2
(a) Schematic of different
phases during adhesion experiments; (1) the sample with the rubber
elastic pillars is moved toward the glass slide, (2) attachment occurs
and a preload P < Pb is applied, (3) the sample is moved upward, lifting the glass plate
with the lifting force FL, (4) a preload P > Pb is applied, where
buckling of the pillars occurs, and (5) pull-off from the glass plate
with the force F. The different sequences are schematically
given in (b, c, and d).
Phase (1): The aligned sample is moved toward
the smooth glass plate.Phase (2): The sample forms contact with the glass plate, and a
compressive preload P below the critical buckling
preload Pb is applied.[40] The load is kept for at least three seconds.Phase (3): The sample is retracted
and, due to adhesion, the glass plate is lifted with a force FL. This phase only applies if the glass plate
is loosened.Phase
(4): A preload above the critical buckling preload Pb is applied, causing the structures to buckle and the
structure tips to detach from the glass probe.[40]Phase
(5): The sample is retracted from the glass plate, and the pull-off
force F, defined as the absolute value of the maximum
negative force of the recorded force–time curves, is measured.These phases can be grouped into different
sequences to represent specific adhesion measurements. The following
sequences were applied:Sequence 1 (Figure b): The preload P is chosen so that contact is formed
with the glass plate, but no buckling of the structures occurs.[40] The glass plate is fixated to prevent it from
lifting. This experiment corresponds to the phases (1) (2) (5).Sequence 2 (Figure c): In this sequence,
a preload P above the critical buckling load Pb is applied.[40] The
glass plate is also fixated. This experiment is represented by the
phases (1) (2) (4) (5).Sequence 3 (Figure d): A pick and place process is imitated using a loosened glass plate.
A preload P below the buckling load Pb is applied and the glass plate is lifted, which corresponds
to the phases (1) (2) (3). In the next step, the glass plate is lowered
and detached using a preload P above the buckling
load Pb, which corresponds to the phases
(2) (4) (5). The complete pick and place process is described by phases
(1) (2) (3) (2) (4) (5).(a) Schematic of different
phases during adhesion experiments; (1) the sample with the rubber
elastic pillars is moved toward the glass slide, (2) attachment occurs
and a preload P < Pb is applied, (3) the sample is moved upward, lifting the glass plate
with the lifting force FL, (4) a preload P > Pb is applied, where
buckling of the pillars occurs, and (5) pull-off from the glass plate
with the force F. The different sequences are schematically
given in (b, c, and d).
Applied Measurement Sets
To determine
the adhesive properties of the switchable bioinspired adhesive and
its applicability for pick and place processes, the following measurement
sets and analyses were conducted in air and vacuum conditions:Force–time
curves were recorded for different preloads up to 7 N, allowing determination
of the preload dependent pull-off force behavior of the switchable
adhesive. The pull-off force F (absolute value of
the maximum detachment force) was plotted as a function of preload P, leading to the identification of the buckling preload Pb.[40] These measurements
correspond to sequence 1 for P < Pb and sequence 2 for P > Pb.Exemplary
force–time curves from (i) were analyzed for two selected measurements,
one having a preload P < Pb according to sequence 1, and one having a preload P > Pb according to sequence
2.Adhesion experiments
with sequence 1 directly followed by sequence 2 were repeated 50 times
to test for reversibility.Force–time curves were recorded for a pick and place process
represented by sequence 3. The glass plate with a weight of 65 g was
lifted for at least 10 s during phase (3). Reversibility was again
tested by repeating this sequence for 50 times.
Results
The experimental results of
the different measurement sets are described in the following four
subsections.
Measurement Set (i): Preload Dependent Pull-off
Force
Pull-off forces F were measured as
a function of preload and are given in Figure . The pull-off force was found to be almost
preload independent at low preloads. As soon as a critical preload
was applied, the pull-off force dropped significantly which corresponded
to the optically observed elastic buckling of the pillars. The critical
buckling preload Pb was highly reproducible
for each sample but showed some variation in a range from 5.6 to 6.0
N for different samples. For preloads P < Pb, the pull-off forces were between 2.5 and
2.7 N in air, while experiments in vacuum resulted in pull-off forces
between 1.8 and 1.9 N. It can be clearly seen that, for lower preloads,
the pull-off force depends on the air pressure; adhesion was reduced
by about 30% in vacuum. For P > Pb, the pull-off force was found to be 0.9 N and was independent
of air pressure.
Figure 3
Absolute values of the pull-off forces are plotted as
a function of applied preload, measured in air and in vacuum. At a
critical preload Pb, indicated by the
dashed line within the gray area, buckling of the pillars was observed.
Absolute values of the pull-off forces are plotted as
a function of applied preload, measured in air and in vacuum. At a
critical preload Pb, indicated by the
dashed line within the gray area, buckling of the pillars was observed.
Measurement
Set (ii): Force–Time Curves
Exemplary force–time
curves with preload P < Pb and P > Pb, respectively, measured in air and vacuum, are plotted in Figure .
Figure 4
Representative force–time
curves for adhesion experiments of bioinspired dry adhesives on a
fixated glass plate. Measurements using sequence 1 with a preload P below the buckling preload Pb in (a) air and (b) in vacuum. Measurements using sequence 2 with
a preload P above the buckling preload Pb (c) in air and d) in vacuum. The phases from Figure a are indicated.
Representative force–time
curves for adhesion experiments of bioinspired dry adhesives on a
fixated glass plate. Measurements using sequence 1 with a preload P below the buckling preload Pb in (a) air and (b) in vacuum. Measurements using sequence 2 with
a preload P above the buckling preload Pb (c) in air and d) in vacuum. The phases from Figure a are indicated.Figure a,b depicts exemplary force–time curves
of adhesion measurements with a preload P < Pb, which corresponds to sequence 1. The sample
was moved toward the fixated glass plate, formed contact, and was
loaded, until the preload P was reached. In these
examples, the preload P was 3.7 N for testing in
air (Figure a) and
4.0 N for testing in vacuum (Figure b). The load was applied for at least 3 s. Then, the
sample was retracted again, leading to an adhesive (tensile) force,
ultimately reaching the maximum pull-off force F.
The negative sign indicates the direction of force measurement. For
these measurements, a pull-off force F of 2.6 N was
found in air (Figure a) and of 1.9 N (Figure b) in vacuum.Representative force–time curves,
where a preload P > Pb was applied corresponding to sequence 2, are shown in Figure c,d. The sample was approached,
formed contact with the glass probe, and was loaded. The compressive
force increased during loading until a local maximum occurred at a
critical load Pb of 5.8 N. Subsequently,
the compressive force decreased rapidly with ongoing compression and
buckling of the pillars was optically observed. As the pillars were
bent further with increasing displacement, the compressive force increased
again until the predefined preload P was reached
and kept for at least 3 s. In Figure c, the preload was 6.6 N, and in Figure d, the preload was 6.3 N. The reverse force–time
behavior was observed during retraction; the occurring maximum corresponded
to an optically observed “unbuckling” of the pillars.
A pull-off force F of 0.9 N was recorded during retraction,
both for measurements in air and in vacuum. These force–time
curves for a preload P > Pb are characteristic for the buckling behavior and have been
reported and characterized in earlier studies.[40]The force–time curves are very similar for
experiments in air and in vacuum. A significant difference was found
only in the pull-off force F; experiments with a
preload P < Pb resulted
in a change in pull-off force F from 2.6 to 1.9 N,
which is equal to a loss in adhesion of approximately 30%. For a preload P > Pb, the pull-off force F was substantially lower, reaching only 0.9 N, and did
not differ between air and vacuum condition.
Measurement
Set (iii): Reversibility
To evaluate the reversibility of
the switching behavior between high and low pull-off force, alternating
preloads below and above Pb were applied.
50 cycles of the sequence 1, directly followed by sequence 2, were
performed in air and in vacuum according to the measurement set (iii). Figure shows the recorded
forces, namely, the preload P and the pull-off force F, as a function of testing cycles. Note that the pull-off
force F is not given as absolute value but has a
negative sign for clarity of the diagram. In air (Figure a), the alternating preloads P of 2.9 and 5.8 N resulted in adhesive forces F of 2.6 and 0.9 N, respectively. In vacuum (Figure b), the applied preloads P were 3.3 and 5.8 N, which resulted in adhesive forces F of 1.9 and 0.9 N, respectively. No notable changes in pull-off force
were found within the 50 test cycles.
Figure 5
Reversibility tests of adhesion measurements
with alternating preload below and above the buckling load, described
by sequence 1, directly followed by sequence 2. Each plot shows the
applied preload P and the resulting pull-off force F over 50 test cycles (a) in air and (b) in vacuum. Note
that the pull-off force F is not given as absolute
value here but has a negative sign for clarity of the diagram.
Reversibility tests of adhesion measurements
with alternating preload below and above the buckling load, described
by sequence 1, directly followed by sequence 2. Each plot shows the
applied preload P and the resulting pull-off force F over 50 test cycles (a) in air and (b) in vacuum. Note
that the pull-off force F is not given as absolute
value here but has a negative sign for clarity of the diagram.
Measurement
Set (iv): Pick and Place
A pick and place process, corresponding
to the sequence 3 in Figure d, was simulated. As described in the measurement set (iv),
an alternating preload below and above the buckling load Pb was applied with the glass plate being loosened to allow
lifting. Representative force–time curves in air and in vacuum
are given in Figure a,b, respectively. The graphs show no notable differences, indicating
that the lifting process and the release of the glass plate were comparable
in both air and vacuum conditions. An exemplary video of a pick and
place process is shown in the Supporting Information, using a silicon wafer instead of a glass plate for the sake of
better visibility.
Figure 6
Force–time curves of a pick and place process,
where a glass plate of 65 g is lifted and released again. The experiment
corresponds to the sequence 3; the numbers indicate the respective
phases from Figure . The measurements were performed (a) in air and (b) in vacuum.
Force–time curves of a pick and place process,
where a glass plate of 65 g is lifted and released again. The experiment
corresponds to the sequence 3; the numbers indicate the respective
phases from Figure . The measurements were performed (a) in air and (b) in vacuum.
Discussion
On the basis of the experimental results, the following properties
of the pressure actuated adhesive system in air and vacuum conditions
were analyzed: the adhesive properties, the reversibility of switching,
and the adaptability for pick and place processes.The adhesive
properties, represented by the force–time curves, are given
in Figure . The curves
exhibit a characteristic shape which is typical for patterned bioinspired
adhesives such as the tested samples. Low preload leads to a comparably
high pull-off force, while high preload results in buckling of the
structures at a certain buckling preload Pb, which reduces adhesion significantly.[40] The mechanism of adhesion loss has been investigated in an earlier
study, where it was found that the unbuckling during unloading does
not allow reformation of intimate contact between the pillar tips
and the probe.[40] The lack of intimate contact
between pillar tips and probe causes a reduction in adhesion. The
adhesive behavior in air and vacuum conditions is qualitatively similar,
indicating that the mechanism of adhesion loss by buckling is unaffected
by air pressure. The main influence of air pressure on the adhesive
properties is found in the magnitude of the pull-off force, which
can be seen in Figures and 4a,b (indicated as ΔF). The pull-off force F was found to be between
2.5 and 2.7 N in air and 1.8 and 1.9 N in vacuum, respectively, exhibiting
that the application of vacuum reduces adhesion by ΔF of 0.7 N ± 0.2 N, which corresponds to a loss in
adhesion of approximately 30%. This reduction becomes obvious by considering Figure ; all pull-off forces
obtained in vacuum lie below the ones obtained in air if the preload
was chosen to be below the critical buckling preload Pb.We identified two factors which may be responsible
for the varying adhesion with changing air pressure, namely, humidity
and suction. We tend to exclude humidity and favor suction as the
main mechanism for the change in pull-off force for the following
reasons. PDMS is a hydrophobic material which does not tend to absorb
water in larger quantities. In addition, it was found that no measurable
humidity effect is present at humidity between 2% and 90% for smooth
PDMS surfaces and for pillar arrays with diameters of 25 μm.[9] Huber et al. found an additional adhesion effect
in the presence of humidity and explained it by a smoothening effect
of the water on rough surfaces,[52] but the
tested surfaces in this study are expected to be smooth. These points
indicate that capillarity effects may have a minor influence on the
adhesion in our experiments.In contrast, suction effects on
mushroom shaped pillars are expected from theoretical considerations[23] and were also found in earlier experimental
studies.[19,20] It was shown that suction is present for
adhesive pillars with mushroom-shaped tips and can contribute considerably
to adhesion with up to 10% of the pull-off force.[19] In our case, the suction component even exceeds this percentage,
reaching approximately 30%. While adhesion experiments in air result
in pull-off forces between 2.5 and 2.7 N, the same set of experiments
in vacuum exhibits pull-off forces between 1.8 and 1.9 N. This difference
may be explained by the size of the pillars. Suction based forces
scale with the area of the contact, while adhesion of patterned surfaces
due to van der Waals forces was theoretically and experimentally shown
to scale with length.[53,54]Theoretically, the suction
force Fsuction of a perfect suction cup,
disregarding other adhesive interactions than suction, is given by
the contact area Acontact and the pressure
difference ΔP caused by the suction effect:Consequently, the
pull-off strength of a perfect suction cup is directly proportional
to the pressure difference inside the contact area and outside the
suction cup. For ideal vacuum (0 bar) and atmospheric pressure (∼1
bar), a suction force of ∼10 N/cm2 can be achieved
using eq . Such high
values are usually not obtained using typical suction cups.To compare our experimentally derived pull-off strength values to
the performance of typical suction cups, we have converted given data
from commercial macroscopic silicone suction cups.[55] The performance of the analyzed suction cups with diameters
between 2.6 and 51.4 mm lie between 2.8 and 7.3 N/cm2 if
a compressive stress of 9.0 N/cm2 is applied;[55] see also Table . For comparing these pull-off strength data with our
results, it is important to consider that the strength data from the
present study reflects the apparent contact strength. Thus, reduction
in “real” contact area due to the pillar packing density
of ∼30% has to be taken into account. A comparable pull-off
performance between conventional suction cups and the experiments
from our studies would then result in a corrected strength, which
is calculated by multiplying the given pull-off strength of the commercial
suction cups with the pillar packing density of ∼30% from our
samples. These values are also given in Table .
Table 1
Geometric Parameters
of Commercial Suction Cups and Their Adhesive Performance after Attaching
Them with a Compressive Load of 9.0 N/cm2 a
diameter, mm
contact area, mm2
force, N
strength, N/cm2
corrected strength,
N/cm2
2.6
5.3
0.15
2.8
0.9
3.8
11.3
0.65
5.7
1.7
5.0
19.6
1.3
6.6
2.0
7.0
38.5
2.5
6.5
2.0
9.0
63.6
3.9
6.1
1.8
11.0
95.0
6.9
7.3
2.2
16.5
213.8
11.0
5.1
1.5
22.0
380.1
16.0
4.2
1.3
32.0
804.2
30.0
3.7
1.1
41.0
1320.3
49.0
3.7
1.1
51.4
2075.0
92.0
4.4
1.3
The corrected strength assumes a packing density
of 30%, data after ref (55).
The corrected strength assumes a packing density
of 30%, data after ref (55).As can be seen from the
corrected pull-off strengths, values between 0.9 and 2.2 N/cm2 can be considered as typical for commercial suction cups.
Our experimentally derived suction component to the pull-off force
lies slightly below the lowest suction force of 0.85 N. As the mushroom
shaped structures in our study were not specifically optimized for
suction, these results fit astonishingly well to the values provided
for commercial suction cups.It was reported that suction and
van der Waals interactions have a different size effect.[23] Thus, we expect that the suction effect becomes
more prominent with increasing size of the contact elements, while
reducing the size of the contact elements diminishes the influence
of suction.Consequently, if suction caused the change in adhesion
of the present experiments, it would be strongly influenced by the
contact geometry. During the buckling process, the mushroom tips detach
and the pillars form side contact with the glass plate. This contact
geometry does not allow building up a difference in air pressure,
thereby diminishing the suction component of the pull-off force. Our
experiments show that the application of a load exceeding Pb leads to a pull-off force F of 0.9 N, both in air and in vacuum. This phenomenon is also reflected
in Figure ; while
the pull-off forces at a preload below Pb differ for measurements in air and in vacuum, similar pull-off forces
are found if the buckling preload Pb is
overcome. These observations support the assumption that air pressure
enhances adhesion due to suction in patterned bioinspired surfaces
with structure sizes in the macroscopic range, while detachment events
after buckling of the pillars are not affected by air pressure, since
suction cannot be maintained after buckling has occurred. This leads
to the conclusion that the difference in pull-off force of the bioinspired
adhesive is not a result of changing humidity but is caused most likely
by a suction effect.For a better description of the switching
behavior, a switching efficiency S is introduced
in eq , which is defined
by the ratio of the pull-off forces at a preload above and below the
buckling preload Pb:A value of S = 0 indicates no switching behavior, and S = 1
resembles a perfect switch where adhesion can be completely turned
on and off. If eq is
applied to the obtained experimental data, the switching efficiency S is approximately 0.65 ± 0.07 in air, while a value
of S = 0.50 ± 0.1 is obtained in vacuum. Thus,
applying vacuum reduces the switching efficiency by a mean value of
ΔS = 0.15. These calculated efficiencies indicate
that the switch in adhesion may be further improved. Still, the reached
values allow a significant change in adhesion in air and in vacuum,
opening a sufficiently large window of operation for pick and place
applications.These promising results are promoted by the reversibility
test shown in Figure , which indicates that the switch in adhesion is highly reversible
in air and in vacuum. No change in adhesive performance or damage
of the dry adhesive structures was detected after 50 testing cycles.
Finally, pick and place processes were conducted using a glass plate
with a weight of 65 g. The glass plate was securely lifted and released
in air and in vacuum. No significant difference is observed in the
adhesion curves given in Figure for operation at both air pressure conditions. Hence,
the pick and place process is not notably influenced by the reduction
of air pressure.It should be mentioned that the pick and release
process has two restrictions for operation if a reliable switch in
adhesion is required. First, if the object to be lifted is too light,
the pull-off force after application of a preload above the buckling
preload Pb may be too high for reliable
detachment, representing a minimum weight threshold, and second, if
the object to be lifted is too heavy, it will detach prior to lifting,
representing a maximum weight threshold. It follows that an optimum
range of operation can be defined on the basis of the pull-off forces
measured for a preload below, and above, the buckling preload Pb. For the tested samples, the range of operation
can be determined to be between approximately 0.9 and 2.5 N in air
or between 0.9 and 1.8 N in vacuum. It further has to be considered
that the viscoelasticity of the applied material may have a significant
influence in the buckling of the structures and may shift both the
lower and the higher boundary of the range of operation. This window
of operation may be tuned according to the envisaged application,
for example, by changing the number of pillars, by their packing density,
by further modification of their tip geometry, or by a different choice
of sample material.
Conclusions
The
present study showed that bioinspired switchable adhesion based on
reversible buckling of elastic pillars is applicable in vacuum. At
low compressive load, the pull-off force for samples with sizes of
1 cm2 was between 2.5 and 2.7 N (±0.1 N) in air and
was reduced to 1.8 to 1.9 N (±0.1 N) if measured in vacuum. This
indicates that a suction component was present in the attachment state,
since an influence of humidity may be excluded. Application of a compressive
load above the buckling preload Pb between
5.6 and 6.0 N (±0.1 N) caused a reversible buckling of the pillars
and resulted in pull-off forces of 0.9 N (±0.1 N), which were
similar for experiments in air and in vacuum. This indicates the absence
of a suction component after buckling of the pillars occurred. Our
experiments exhibited that the transition between the two adhesive
states was sharp and the switching behavior was independent of air
pressure. Further, the switch in adhesion exhibited high reversibility;
we showed that the system works reliably in air and in vacuum for
50 pick and place cycles without any signs of wear or change in adhesion
performance. The functionality of the switchable adhesive at low air
pressure makes it applicable for handling operations of fragile objects
in vacuum.
Authors: Gabriela Moreira Lana; Xuan Zhang; Christian Müller; René Hensel; Eduard Arzt Journal: ACS Appl Mater Interfaces Date: 2022-10-04 Impact factor: 10.383
Authors: Julian K A Langowski; Dimitra Dodou; Peter van Assenbergh; Johan L van Leeuwen Journal: Integr Comp Biol Date: 2020-10-01 Impact factor: 3.326