Tanvir Samra1, Neerja Banerjee2, Arushi Gupta3. 1. Department of Anaesthesia and Intensive Care, Post Graduate Institute of Medical Education and Research, Chandigarh, India. 2. Department of Anaesthesia and Intensive Care, Dr. Ram Manohar Lohia Hospital, New Delhi, India. 3. Department of Anaesthesia and Critical Care, Dr. Ram Manohar Lohia Hospital, New Delhi, India.
Abstract
INTRODUCTION: Caloric intake of critically ill patients are usually calculated using predictive equations. Recent advances in gas exchange measurements have the potential to estimate energy expenditure at the bedside and at different time periods. MATERIALS AND METHODS: Energy needs of critically ill patients were estimated over a period of 3 months using simplistic formula of 25 kcal/kg/day estimated energy expenditure (EEE), Harris-Benedict equation (HBE) (Basal energy expenditure [BEE]) and M-COVX™ metabolic monitor resting energy expenditure (REE) on day 4 of Intensive Care Unit (ICU) admission. Calculations based on HBE were taken as standard, and percentage errors (PE) were calculated for each patient for values derived from simplistic formula and metabolic monitor. Adequacy of nutritional intake in ICU was also assessed. RESULTS: Metabolic monitor could be used in only 20/70 patients. The mean age of patients was 40 years, 65% were males, and average body mass index was 23.69 kg/m(2). Intermittent intolerance to feeds was reported in 50%. Values of REE and EEE were greater than BEE in 70% of patients. A significant difference was reported in values of PE of ≤20% and ≥30%; P = 0.0003 and 0.0001, respectively estimated using REE and EEE. CONCLUSIONS: It is not feasible to use metabolic monitors in all patients. Variability in readings is large and further studies are needed to establish the validity of its measurements. Calculations using simplistic formulas are much closer to values obtained using HBE.
INTRODUCTION: Caloric intake of critically illpatients are usually calculated using predictive equations. Recent advances in gas exchange measurements have the potential to estimate energy expenditure at the bedside and at different time periods. MATERIALS AND METHODS: Energy needs of critically illpatients were estimated over a period of 3 months using simplistic formula of 25 kcal/kg/day estimated energy expenditure (EEE), Harris-Benedict equation (HBE) (Basal energy expenditure [BEE]) and M-COVX™ metabolic monitor resting energy expenditure (REE) on day 4 of Intensive Care Unit (ICU) admission. Calculations based on HBE were taken as standard, and percentage errors (PE) were calculated for each patient for values derived from simplistic formula and metabolic monitor. Adequacy of nutritional intake in ICU was also assessed. RESULTS: Metabolic monitor could be used in only 20/70 patients. The mean age of patients was 40 years, 65% were males, and average body mass index was 23.69 kg/m(2). Intermittent intolerance to feeds was reported in 50%. Values of REE and EEE were greater than BEE in 70% of patients. A significant difference was reported in values of PE of ≤20% and ≥30%; P = 0.0003 and 0.0001, respectively estimated using REE and EEE. CONCLUSIONS: It is not feasible to use metabolic monitors in all patients. Variability in readings is large and further studies are needed to establish the validity of its measurements. Calculations using simplistic formulas are much closer to values obtained using HBE.
Entities:
Keywords:
Energy expenditure; gas exchange; measurement techniques; metabolic monitors; metabolism
Energy needs of critically illpatients are estimated using simplistic formulas (25–30 kcal/kg/day), predictive equations or indirect calorimeters.[1] Metabolic monitors like M-COVX™ (GE Healthcare/Datex-Ohmeda) measure pulmonary oxygen uptake (VO2) and carbon dioxide production (VCO2) and have thus been used in Intensive Care Units (ICUs) for measuring resting energy expenditure (REE), work of breathing and alveolar dead space.[23]Critically illpatients are in an unstable metabolic or hemodynamic state. The energy needs of these patients thus vary from one to another and in a single patient vary from 1-time to another.[4]Calorie intake is calculated in our ICU using a simplistic formula of 25 kcal/kg/day. The aim of this pilot study was to evaluate the feasibility of M-COVX™ metabolic monitor for assessment of nutritional needs of critically illpatients admitted in our multidisciplinary ICU. Values of REE estimated by M-COVX™ were compared with those calculated using Harris-Benedict equation (HBE). Variability in the measurements, the clinical significance of the difference and adequacy of nutritional intake was assessed for each patient.
Materials and Methods
After approval from the Ethics Committee, a pilot study was conducted over a period of 3 months to assess the nutritional status of patients in the multidisciplinary ten bedded ICU of our hospital.
Inclusion criteria
All patients on assisted or control mode of mechanical ventilation and stable hemodynamicsDay 4 of ICU admission.Energy requirements were calculated using HBE (Basal energy expenditure [BEE]), empirical formula of 25 kcal/kg (Estimated energy expenditure [EEE]) and simultaneously recorded from metabolic monitor REE.Value of BEE was considered as a gold standard and all comparisons were made with it.[5]Harris-Benedict formula:Males: BEE (kcal/day) =66 + (13.7 × weight in kg) + (5 × height in cm) − (6.8 × age in years)Females: BEE (kcal/day) =655 + (9.7 × weight in kg) + (1.8 × height in cm) − (4.7 × age in years).M-COVX™ (Engstrom carestation, Datex-Ohmeda Inc., Madison, WI, USA) metabolic module was used in our ICU. It consists of a gas analyzer and a spirometer unit and displays oxygen consumption (VO2) and carbon dioxide production (VCO2). Partial pressures of O2 and CO2 are measured by the rapid paramagnetic analyzer and infrared analyzer, respectively and inspired tidal volume is measured using a pneumotachograph. Modified weir equation (REE = [VO2 (3.941) + VCO2 (1.11)] 1440 kcal/day) is used to calculate REE and respiratory quotient (RQ = VCO2 /VO2).[6] Reliability of displayed value of REE was assessed after considering the concomitant readings of RQ whose normal values range from 0.69 to 0.98. All values of REE with an RQ outside the above mentioned range were discarded.The value of REE was recorded from the metabolic monitor at 700 and 1900 h. The following precautions were taken at the time of measurement:Patients were not disturbed, and no major procedure was performed within 30 min of the test; e.g., tracheal suctioning, physiotherapy, postural changes, radiologic studies, and body washingsConstant minute ventilation was assured for 30 min prior to testTotal of three 1-min gas exchange measurements of REE and RQ were recorded and the mean value recordedLeaks in the ventilator circuit and around tracheal tube cuffs were checkedCare was taken to ensure that there was no water vapor in the circuit.Patients with agitation, chest tubes or bronchopleural fistula, cuff leaks, and increased oxygen requirements (FiO2 >0.4), were excluded. Demographic parameters (age, sex, weight, height, and body mass index [BMI]) clinical diagnosis, Sequential Organ Failure Assessment (SOFA) score, vitals (heart rate, blood pressure, pulse oximetry [SPO2], end tidal carbon dioxide concentrations, and temperature) biochemical variables (blood gas analysis, electrolyte values, serum albumin, hematology, and coagulation studies) were recorded.Intermittent enteral feeds were administered at intervals of 3 h by the nurses through a nasogastric tube. The empirical formula of 25 kcal/kg was used for calculation of caloric intake, and the total intake was administered in six equally divided feeds with 6 h of rest at night. Formula feed [Appendix 1] with calorie density of 1 kcal/ml and carbohydrate to fat ratio of 60:40 was given to all patients (Ensure, Abbott Laboratories BV, Zwolle, The Netherlands, Division of Abbott Laboratories, USA). All checks and precautions were taken to avoid aspiration. Intolerance to feed was defined as an aspirate volume >200 ml or more than half the previous feed after administration of a prokinetic (injection metoclopramide 0.15 mg/kg BID) agent. Dextrose infusions (5%) at 1 ml/kg/h was administered to patients with intolerance to enteral feeding. Total caloric intake (enteral and parenteral) was recorded for each patient. Signs of intolerance such as abdominal distention, diarrhea, and absent bowel sounds were also recorded. Percentage error (PE) was calculated to compare the values of REE (recorded from metabolic monitor) and EEE (calculated using empirical formula) with the gold standard that is, BEE (HBE).Formula used to calculate PE:PE of more than 20% was considered clinically significant. Fischer's exact test was applied to compare the values in each group and calculate the level of significance. P < 0.05 was considered statistically significant.
Results
The metabolic monitor could be used in only 20 patients [Figure 1].
Figure 1
Flowchart of patient enrollment
Flowchart of patient enrollmentThe demographic data is summarized in Table 1. The mean age of the patients was 40 years (range; 22–62 years) and 65% were males. Average BMI was 23.69 kg/m2 (range; 15.22–29.1 kg/m2); only one patient was malnourished; none was obese. About half of the patients had serum albumin levels <3 g% and 75% had < 3.5 g%. Median value of SOFA score was 6; approximately half of the patients had scores below it and half had above it.
Table 1
Demographic data of the 20 patients included in the study
Demographic data of the 20 patients included in the studyEnteral feeding was initiated in 30% (21/70) of patients on the 1st day of ICU admission. It was established in 75% (53/70) of the patients by day 3. Intermittent episodes of intolerance were reported in 50% of the patients. Parenteral supplementation using 5% dextrose and/or intralipid infusions was needed in 7 patients.Estimations of energy expenditure and nutritional intake of patients are summarized in Tables 2 and 3.
Table 2
Nutritional assessment based on three different approaches and total intake in 24 h (day 4 of ICU stay)
Table 3
Calorie intake and PE
Nutritional assessment based on three different approaches and total intake in 24 h (day 4 of ICU stay)Calorie intake and PE
Nutritional intake
70% (14/20) had calorie intake >75%15% (3/20) had an intake of 60–65%15% had intake of 30–40%Four patients had an intake >90% of estimated.The values of REE obtained from metabolic monitors were greater than BEE calculated using HBE in 70% (14/20) of patients.
Percentage error value of resting energy expenditure
≤20% PE in 55% (11/20) of patients20–30% PE in 30% (6/20) ≥30% PE in 15% of (3/20) patientsThe value of EEE was greater than BEE in 70% (14/20) of patients.
Percentage error values of estimated energy expenditure
≤20% in 80% of patients (16/20)20–30% PE in 20% patientsNo patient had a difference of >30% [Table 4].
Table 4
Comparison of values of PE for REE and EEE
Comparison of values of PE for REE and EEE
Discussion
Routine use of metabolic monitors for nutritional assessment is labor intensive.[7] In our study, a dedicated clinician was involved in data collection and nutritional assessment. An intensivist has multiple tasks to perform, e.g., admission/discharge of patients, intubation/extubation, hemodynamic stabilization, ventilator management, etc., and thus it may not be feasible for the clinician to make appropriate use of the metabolic monitor in the ICU. Criteria to be fulfilled prior to data collection are very stringent and lead to erroneous results if not followed precisely. Values cannot be recorded unless a constant minute ventilation is attained. In a multidisciplinary ICU, it is technically very difficult to ensure that the patient is not disturbed for 30 min prior to the test. Only 28% of the patients admitted in our ICU fulfilled the criteria for inclusion in this study.We considered the value computed by HBE as the gold standard. PEREE and PEEEE were then calculated as described above. We recorded a significant difference (P = 0.0003) in the values belonging to group I when comparing PEREE and PEEEE. But error ≤20% (group I) was not considered to cause any significant change clinically. This difference is statistically significant but may not have any clinical relevance. No significant difference was noted in the values belonging to group II (20–30% error). A significant difference (P = 0.0003) was present in the values in group III (≥30% error). But only 15% of patients included for assessment belonged to this group. Thus, in this pilot study the PE obtained using REE were larger than that obtained using EEE and the difference was statistically significant (group III). This questions the reliability of the use of metabolic monitors for nutritional assessment and favors the use of simplistic formula of 25 ml/kg/day.Canadian Clinical Practice Guidelines published in 2013[8] clearly state that there is insufficient evidence to put forward a recommendation for the use of indirect calorimetry over predictive equations for determining energy needs in critically illpatients. Brandi et al.[9] reported several sources of error and technical difficulties in applying indirect calorimetry in critical care settings. However, metabolic monitors may prove beneficial for assessment of energy needs in patients with poor response to nutritional supplementation, patients with single- or multiple-organ dysfunction, patients with prolonged ICU stay and artificial nutritional support. Some more indications enumerated in the American Association for Respiratory Care Clinical Practice guidelines[2] are as follows: Patients with neurological trauma, chronic obstructive pulmonary disease, acute pancreatitis, multiple trauma, amputations, severe sepsis, obesity, and severely hyper metabolic or hypo metabolicpatients.BEE of critically illpatients is calculated using HBE. In critically ill septicpatients or patients with multiple injuries, it is further multiplied with different stress and activity factors. The HBEs has been found to overestimate energy expenditure by 6–15% when compared with measurements done using indirect calorimetry.[10] Intake of 25 kcal/kg is adequate for most patients with normal BMI. In most patients, this value approximates the one calculated from HBE. The poor agreement has been documented by Chioléro et al.[11] between predictive equations and measured values derived from metabolic monitors. Results of our study are similar to these findings. Meyer et al.[12] concluded that the measurements using M-COVX metabolic monitor were not in the clinically acceptable range in nonsedated patients and different ventilation modes.One of the limitations of our study is that we did not assess patients for nutritional status in our ICU. A very rough estimate can be made from the value of BMI, which was less than normal in only one patient (15.22 kg/m2) with a diagnosis of Ca esophagus. Chakravarty et al.[13] reported malnutrition in two-fifth of the patients admitted in their ICU and thus emphasized on the urgent need to develop a comprehensive nutritional care program. They used Subjective Global Nutritional Assessment score to screen their patients.Our study was an observational pilot study with a limited number of patients. We did not categorize the patients on the basis of the severity of illness, sex, BMI, etc., which could have a confounding effect on the measurements. We took three readings each at 700 and 1900 h but only on day 4 of ICU admission. There are day to day variabilities, and thus data collection should be done every day.Hypoalbuminemia was present in three-fourth of our patients (50% had serum albumin levels <3 g%; 75% had <3.5 g%). But it is not a marker of nutritional status. It is used as a marker of systemic inflammatory response and has prognostic importance.Underfeeding (<50% of daily energy requirement) has been reported in 38% of critically illpatients in a previous study.[14] The most common cause of interruptions in tube feedings was the performance of ICU tests and invasive procedures and feed intolerance.Energy expenditure can be measured directly by putting a person in a calorimeter and measuring the amount of heat produced by the body mass.This is expensive and very impractical in the clinical setting. Most valid metabolic cart is Deltatrac TM (Datex-Ohmeda Division, Instrumentarium Corp, Teollisuuskatu 29, Helsinki, FIN 00031, Finland).[15] It has undergone a number of independent laboratory and clinical validations. But it is expensive, requires high technical expertise and is time consuming. New compact modular metabolic monitors such as E-COVX™ (formerly M-COVX™, GE Healthcare/Datex-Ohmeda) are less expensive, simpler, smaller, and automatically calibrated. Kaiyala[16] has concluded that mathematical relationships between whole body oxygen uptake, carbon dioxide release, and metabolic heat production may are complex and thus unpredictable and thus assumptions made by practitioners of respirometric indirect calorimeters may be wrong.We have successfully used Deltatrac II in a previous study conducted at our institute to compare two different regimes of enteral feeding-continuous vs. intermittent in patients with head injuries.[17] The metabolic monitor was also advantageous in diagnosing and managing a child with hypometabolism.[18] In this pilot study, we aimed to assess the feasibility and accuracy of the metabolic monitor as a tool to assess nutritional requirements of patients admitted in a multidisciplinary ICU.We conclude that technical advances in measurement of gas exchange in ICU's have enabled the use of metabolic monitors for estimation of energy expenditure, assessment of pulmonary physiology, oxygen utilization, success of weaning, and measurement of metabolic stress. In order to fully utilize its potential it is essential to train a few intensivists in the measurement techniques and interpretation of data with these modules so that its limitations can be acknowledged and relevant information obtained. A team approach can also be used in the ICU with the involvement of a physician, nutritionist, and respiratory therapist in analyzing the results of the metabolic monitor and then modifying nutritional support. But first and foremost is the performance of further studies to establish the reliability and validity of results obtained with these modules. The impact of metabolic monitors on clinical outcomes such as mortality, ICU length of stay, duration of mechanical ventilation also needs to be established.