GafChromic(®) EBT3 films for patient specific IMRT QA using a multichannel approach.

Livia Marrazzo1, Margherita Zani2, Stefania Pallotta3, Chiara Arilli4, Marta Casati4, Antonella Compagnucci4, Cinzia Talamonti3, Marta Bucciolini3.   

Abstract

PURPOSE: To evaluate EBT3 for pre-treatment patient specific quality assurance (QA). The method we propose combines the experience gained in our center with the guidelines of the protocol proposed by Lewis et al. in 2012. To compare the multichannel approach with the single channel dosimetry.
METHODS: Gafchromic® EBT3 films were irradiated both at linac and TomoTherapy and calibration curves were obtained. A series of irradiations with simple fields (uniform dose distributions on regular shaped targets) was performed. In a second stage, films were exposed to full clinical plans at linac (step and shoot IMRT and VMAT). At TomoTherapy dose maps were obtained for a clinical plan in three different coronal planes. Films were digitized using an Epson 10000XL scanner and FilmQA™ Pro software was employed for the analysis.
RESULTS: The measured calibration curves suggest that, at least for the two beams taken into account (6 MV linac and TomoTherapy), a single calibration can be successfully adopted for each film lot. The application of the multichannel optimization method strongly improves the results in terms of gamma passing rates of the comparison between measured and calculated maps.
CONCLUSIONS: Up to now EBT films, although attractive, were not preferred for routine patient specific QA due to their complex and time consuming processing and to the challenging work of characterization. The application of the mentioned protocol, together with some additional precautions, and the adoption of the multichannel optimization dosimetry, make this detector a handy and reliable tool for patient specific QA.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  IMRT; Multichannel dosimetry; Patient specific quality assurance; Radiochromic film

Mesh:

Year:  2015        PMID: 26429383     DOI: 10.1016/j.ejmp.2015.08.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  10 in total

1.  Correction of lateral response artifacts from flatbed scanners for dual-channel radiochromic film dosimetry.

Authors:  Yuichi Akino; Hiroya Shiomi; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Takero Hirata; Hirokazu Mizuno; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2021-03-10       Impact factor: 2.724

2.  SBRT/SRS patient-specific QA using GAFchromicTM EBT3 and FilmQATM Pro software.

Authors:  Giuseppe Stella; Nina Cavalli; Elisa Bonanno; Lucia Zirone; Giuseppina Rita Borzì; Martina Pace; Andrea Girlando; Anna M Gueli; Carmelo Marino
Journal:  J Radiosurg SBRT       Date:  2022

3.  Dosimetry with gafchromic films based on a new micro-opto-electro-mechanical system.

Authors:  C Guardiola; A Márquez; M C Jiménez-Ramos; J García López; A Baratto-Roldán; X Muñoz-Berbel
Journal:  Sci Rep       Date:  2021-05-17       Impact factor: 4.379

4.  The combination of the error correction methods of GAFCHROMIC EBT3 film.

Authors:  Yinghui Li; Lixin Chen; Jinhan Zhu; Xiaowei Liu
Journal:  PLoS One       Date:  2017-07-27       Impact factor: 3.240

5.  Assessment of log-based fingerprinting system of Mobius3D with Elekta linear accelerators.

Authors:  Yu-Yun Noh; Jihun Kim; Jin Sung Kim; Han-Back Shin; Min Cheol Han; Tae Suk Suh
Journal:  J Appl Clin Med Phys       Date:  2021-11-27       Impact factor: 2.102

6.  Comparison of patient-specific intensity modulated radiation therapy quality assurance for the prostate across multiple institutions.

Authors:  Kazuki Kubo; Hajime Monzen; Kohei Shimomura; Kenji Matsumoto; Tomoharu Sato; Mikoto Tamura; Kiyoshi Nakamatsu; Kentaro Ishii; Ryu Kawamorita
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-17

7.  Improving Delivery Accuracy of Stereotactic Body Radiotherapy to a Moving Tumor Using Simplified Volumetric Modulated Arc Therapy.

Authors:  Young Eun Ko; Byungchul Cho; Su Ssan Kim; Si Yeol Song; Eun Kyung Choi; Seung Do Ahn; Byongyong Yi
Journal:  PLoS One       Date:  2016-06-22       Impact factor: 3.240

8.  Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array.

Authors:  Juan Fernando Mata Colodro; Alfredo Serna Berná; Vicente Puchades Puchades; David Ramos Amores; Miguel Alcaraz Baños
Journal:  J Med Phys       Date:  2017 Jul-Sep

9.  Retrospective dosimetry study of intensity-modulated radiation therapy for nasopharyngeal carcinoma: measurement-guided dose reconstruction and analysis.

Authors:  Wen-Zhao Sun; Dan-Dan Zhang; Ying-Lin Peng; Li Chen; De-Hua Kang; Bin Wang; Xiao-Wu Deng
Journal:  Radiat Oncol       Date:  2018-03-15       Impact factor: 3.481

10.  A method for time-independent film dosimetry: Can we obtain accurate patient-specific QA results at any time postirradiation?

Authors:  Leon Dunn; Guy Godwin; James Hellyer; Xiaolei Xu
Journal:  J Appl Clin Med Phys       Date:  2022-01-20       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.