Influence of radiation dose and reconstruction algorithm in MDCT assessment of airway wall thickness: A phantom study.

Daniel Gomez-Cardona1, Scott K Nagle2, Ke Li3, Terry E Robinson4, Guang-Hong Chen3.   

Abstract

PURPOSE: Wall thickness (WT) is an airway feature of great interest for the assessment of morphological changes in the lung parenchyma. Multidetector computed tomography (MDCT) has recently been used to evaluate airway WT, but the potential risk of radiation-induced carcinogenesis-particularly in younger patients-might limit a wider use of this imaging method in clinical practice. The recent commercial implementation of the statistical model-based iterative reconstruction (MBIR) algorithm, instead of the conventional filtered back projection (FBP) algorithm, has enabled considerable radiation dose reduction in many other clinical applications of MDCT. The purpose of this work was to study the impact of radiation dose and MBIR in the MDCT assessment of airway WT.
METHODS: An airway phantom was scanned using a clinical MDCT system (Discovery CT750 HD, GE Healthcare) at 4 kV levels and 5 mAs levels. Both FBP and a commercial implementation of MBIR (Veo(TM), GE Healthcare) were used to reconstruct CT images of the airways. For each kV-mAs combination and each reconstruction algorithm, the contrast-to-noise ratio (CNR) of the airways was measured, and the WT of each airway was measured and compared with the nominal value; the relative bias and the angular standard deviation in the measured WT were calculated. For each airway and reconstruction algorithm, the overall performance of WT quantification across all of the 20 kV-mAs combinations was quantified by the sum of squares (SSQs) of the difference between the measured and nominal WT values. Finally, the particular kV-mAs combination and reconstruction algorithm that minimized radiation dose while still achieving a reference WT quantification accuracy level was chosen as the optimal acquisition and reconstruction settings.
RESULTS: The wall thicknesses of seven airways of different sizes were analyzed in the study. Compared with FBP, MBIR improved the CNR of the airways, particularly at low radiation dose levels. For FBP, the relative bias and the angular standard deviation of the measured WT increased steeply with decreasing radiation dose. Except for the smallest airway, MBIR enabled significant reduction in both the relative bias and angular standard deviation of the WT, particularly at low radiation dose levels; the SSQ was reduced by 50%-96% by using MBIR. The optimal reconstruction algorithm was found to be MBIR for the seven airways being assessed, and the combined use of MBIR and optimal kV-mAs selection resulted in a radiation dose reduction of 37%-83% compared with a reference scan protocol with a dose level of 1 mGy.
CONCLUSIONS: The quantification accuracy of airway WT is strongly influenced by radiation dose and reconstruction algorithm. The MBIR algorithm potentially allows the desired WT quantification accuracy to be achieved with reduced radiation dose, which may enable a wider clinical use of MDCT for the assessment of airway WT, particularly for younger patients who may be more sensitive to exposures with ionizing radiation.

Entities:  

Mesh:

Year:  2015        PMID: 26429266      PMCID: PMC4575323          DOI: 10.1118/1.4930797

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  An automated approach to quantitative air trapping measurements in mild cystic fibrosis.

Authors:  Michael L Goris; Hongyun J Zhu; Francis Blankenberg; Frandics Chan; Terry E Robinson
Journal:  Chest       Date:  2003-05       Impact factor: 9.410

2.  Fully automated system for three-dimensional bronchial morphology analysis using volumetric multidetector computed tomography of the chest.

Authors:  Raman Venkatraman; Raghav Raman; Bhargav Raman; Richard B Moss; Geoffrey D Rubin; Lawrence H Mathers; Terry E Robinson
Journal:  J Digit Imaging       Date:  2006-06       Impact factor: 4.056

Review 3.  State of the Art. A structural and functional assessment of the lung via multidetector-row computed tomography: phenotyping chronic obstructive pulmonary disease.

Authors:  Eric A Hoffman; Brett A Simon; Geoffrey McLennan
Journal:  Proc Am Thorac Soc       Date:  2006-08

4.  A three-dimensional statistical approach to improved image quality for multislice helical CT.

Authors:  Jean-Baptiste Thibault; Ken D Sauer; Charles A Bouman; Jiang Hsieh
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

5.  Relationship between airway wall thickness assessed by high-resolution computed tomography and lung function in patients with asthma and chronic obstructive pulmonary disease.

Authors:  J Kosciuch; R Krenke; K Gorska; M Zukowska; M Maskey-Warzechowska; R Chazan
Journal:  J Physiol Pharmacol       Date:  2009-11       Impact factor: 3.011

6.  Automated CT scan scores of bronchiectasis and air trapping in cystic fibrosis.

Authors:  Emily M DeBoer; Waldemar Swiercz; Sonya L Heltshe; Margaret M Anthony; Paul Szefler; Rebecca Klein; John Strain; Alan S Brody; Scott D Sagel
Journal:  Chest       Date:  2014-03-01       Impact factor: 9.410

7.  CT reconstruction techniques for improved accuracy of lung CT airway measurement.

Authors:  A Rodriguez; F N Ranallo; P F Judy; D S Gierada; S B Fain
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  Accurate airway wall estimation using phase congruency.

Authors:  Raúl San José Estépar; George G Washko; Edwin K Silverman; John J Reilly; Ron Kikinis; Carl-Fredrik Westin
Journal:  Med Image Comput Comput Assist Interv       Date:  2006

9.  Chest computed tomography predicts the frequency of pulmonary exacerbations in children with cystic fibrosis.

Authors:  Don B Sanders; Zhanhai Li; Alan S Brody
Journal:  Ann Am Thorac Soc       Date:  2015-01

10.  Automatic airway analysis on multidetector computed tomography in cystic fibrosis: correlation with pulmonary function testing.

Authors:  Mark O Wielpütz; Monika Eichinger; Oliver Weinheimer; Sebastian Ley; Marcus A Mall; Matthias Wiebel; Arved Bischoff; Hans-Ulrich Kauczor; Claus P Heußel; Michael Puderbach
Journal:  J Thorac Imaging       Date:  2013-03       Impact factor: 3.000

View more
  2 in total

1.  Influence of exposure parameters and iterative reconstruction on automatic airway segmentation and analysis on MDCT-An ex vivo phantom study.

Authors:  Patricia Leutz-Schmidt; Oliver Weinheimer; Bertram J Jobst; Julien Dinkel; Jürgen Biederer; Hans-Ulrich Kauczor; Michael U Puderbach; Mark O Wielpütz
Journal:  PLoS One       Date:  2017-08-02       Impact factor: 3.240

2.  Comparison of Filtered Back Projection, Hybrid Iterative Reconstruction, Model-Based Iterative Reconstruction, and Virtual Monoenergetic Reconstruction Images at Both Low- and Standard-Dose Settings in Measurement of Emphysema Volume and Airway Wall Thickness: A CT Phantom Study.

Authors:  Cherry Kim; Ki Yeol Lee; Chol Shin; Eun-Young Kang; Yu-Whan Oh; Moin Ha; Chang Sub Ko; Jaehyung Cha
Journal:  Korean J Radiol       Date:  2018-06-14       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.