Quantification of Shear-Induced Platelet Activation: High Shear Stresses for Short Exposure Time.

Jun Ding1,2, Zengsheng Chen1, Shuqiong Niu1, Jiafeng Zhang1, Nandan K Mondal1, Bartley P Griffith1, Zhongjun J Wu1.   

Abstract

Thrombosis and thromboembolism are the life-threatening clinical complications for patients supported or treated with prosthetic cardiovascular devices. The high mechanical shear stress within these devices is believed to be the major contributing factor to cause platelet activation (PA) and function alteration, leading to thrombotic events. There have been limited quantitative data on how the high mechanical shear stress causes platelet activation. In this study, shear-induced PA in the ranges of well-defined shear stress and exposure time relevant to cardiovascular devices was quantitatively characterized for human blood using two novel flow-through Couette-type blood shearing devices. Four markers of platelet activation-surface P-selectin (CD62p), platelet-derived microparticles (PMPs), platelet-monocyte aggregation (PMA), and soluble P-selectin-were measured by flow cytometry and enzyme-linked immunosorbent assay (ELISA), respectively. The results indicated that PA induced by high shear stresses with short exposure time could be reliably detected with surface P-selectin, and, to a lesser extent, PMPs rather than soluble P-selectin. It was also verified that PMA can be a highly sensitive indirect marker of platelet activation. The quantitative relationship between percentage of activated platelets indicated by surface P-selectin expression and shear stress/exposure time follows well the power law functional form. The coefficients of the power law models of PA based on surface P-selectin expression were derived.
Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Blood-shearing device; Exposure time; Power law model; Shear stress; Shear-induced platelet activation

Mesh:

Substances:

Year:  2015        PMID: 25808300     DOI: 10.1111/aor.12438

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  16 in total

1.  Are anticoagulant independent mechanical valves within reach-fast prototype fabrication and in vitro testing of innovative bi-leaflet valve models.

Authors:  Lawrence N Scotten; Rolland Siegel
Journal:  Ann Transl Med       Date:  2015-08

2.  Constricted microfluidic devices to study the effects of transient high shear exposure on platelets.

Authors:  Nesreen Z Alsmadi; Sarah J Shapiro; Christopher S Lewis; Vinit M Sheth; Trevor A Snyder; David W Schmidtke
Journal:  Biomicrofluidics       Date:  2017-11-28       Impact factor: 2.800

3.  Shear-induced platelet receptor shedding by non-physiological high shear stress with short exposure time: glycoprotein Ibα and glycoprotein VI.

Authors:  Zengsheng Chen; Nandan K Mondal; Jun Ding; Jingya Gao; Bartley P Griffith; Zhongjun J Wu
Journal:  Thromb Res       Date:  2015-02-07       Impact factor: 3.944

4.  A New Evaluation Q-Factor to Be Calculated for Suction Geometries as a Basis for Smooth Suction in the Operating Field to Ensure the Highest Possible Blood Integrity for Retransfusion Systems.

Authors:  Ireneusz Iwanowski; Jan Böckhaus; Pascal Richardt; Ingo Kutschka; Gunnar G Hanekop; Martin G Friedrich
Journal:  J Extra Corpor Technol       Date:  2022-06

5.  Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit.

Authors:  Patrick Borchers; Patrick Winnersbach; Sandra Kraemer; Christian Beckers; Eva Miriam Buhl; Steffen Leonhardt; Rolf Rossaint; Marian Walter; Thomas Breuer; Christian Bleilevens
Journal:  J Clin Med       Date:  2022-06-15       Impact factor: 4.964

6.  Quantitative Characterization of Shear-Induced Platelet Receptor Shedding: Glycoprotein Ibα, Glycoprotein VI, and Glycoprotein IIb/IIIa.

Authors:  Zengsheng Chen; Steven C Koenig; Mark S Slaughter; Bartley P Griffith; Zhongjun J Wu
Journal:  ASAIO J       Date:  2018 Nov/Dec       Impact factor: 2.872

7.  Impact of high mechanical shear stress and oxygenator membrane surface on blood damage relevant to thrombosis and bleeding in a pediatric ECMO circuit.

Authors:  Wenji Sun; Shigang Wang; Zengsheng Chen; Jiafeng Zhang; Tieluo Li; Katherin Arias; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2020-02-19       Impact factor: 3.094

8.  Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor.

Authors:  Zengsheng Chen; Nandan K Mondal; Jun Ding; Steven C Koenig; Mark S Slaughter; Zhongjun J Wu
Journal:  Artif Organs       Date:  2015-11-18       Impact factor: 3.094

9.  Models of Shear-Induced Platelet Activation and Numerical Implementation With Computational Fluid Dynamics Approaches.

Authors:  Dong Han; Jiafeng Zhang; Bartley P Griffith; Zhongjun J Wu
Journal:  J Biomech Eng       Date:  2022-04-01       Impact factor: 2.097

10.  Shear Stress Accumulation Enhances von Willebrand Factor-Induced Platelet P-Selectin Translocation in a PI3K/Akt Pathway-Dependent Manner.

Authors:  Jinhua Fang; Xiaoxi Sun; Silu Liu; Pu Yang; Jiangguo Lin; Jingjing Feng; Miguel A Cruz; Jing-Fei Dong; Ying Fang; Jianhua Wu
Journal:  Front Cell Dev Biol       Date:  2021-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.