Quantitative assessment of corneal vibrations during intraocular pressure measurement with the air-puff method in patients with keratoconus.

Robert Koprowski1, Renato Ambrósio2.   

Abstract

BACKGROUND: One of the current methods for measuring intraocular pressure is the air-puff method. A tonometer which uses this method is the Corvis device. With the ultra-high-speed (UHS) Scheimpflug camera, it is also possible to observe corneal deformation during measurement. The use of modern image analysis and processing methods allows for analysis of higher harmonics of corneal deflection above 100 Hz.
METHOD: 493 eyes of healthy subjects and 279 eyes of patients with keratoconus were used in the measurements. For each eye, 140 corneal deformation images were recorded during intraocular pressure measurement. Each image was recorded every 230 µs and had a resolution of 200 × 576 pixels. A new, original algorithm for image analysis and processing has been proposed. It enables to separate the eyeball reaction as well as low-frequency and high-frequency corneal deformations from the eye response to an air puff. Furthermore, a method for classification of healthy subjects and patients with keratoconus based on decision trees has been proposed.
RESULTS: The obtained results confirm the possibility to distinguish between patients with keratoconus and healthy subjects. The features used in this classification are directly related to corneal vibrations. They are only available in the proposed software and provide specificity of 98%, sensitivity-85%, and accuracy-92%. This confirms the usefulness of the proposed method in this type of classification that uses corneal vibrations during intraocular pressure measurement with the Corvis tonometer. DISCUSSION: With the new proposed algorithm for image analysis and processing allowing for the separation of individual features from a corneal deformation image, it is possible to: automatically measure corneal vibrations in a few characteristic points of the cornea, obtain fully repeatable measurement of vibrations for the same registered sequence of images and measure vibration parameters for large inter-individual variability in patients.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Air-puff; Corvis ST tonometer; High-speed deformation; Image processing; Keratoconus; Vibrations; measurement automation

Mesh:

Year:  2015        PMID: 26410602     DOI: 10.1016/j.compbiomed.2015.09.007

Source DB:  PubMed          Journal:  Comput Biol Med        ISSN: 0010-4825            Impact factor:   4.589


  6 in total

1.  A new method for detecting the outer corneal contour in images from an ultra-fast Scheimpflug camera.

Authors:  Magdalena Jędzierowska; Robert Koprowski; Sławomir Wilczyński; Katarzyna Krysik
Journal:  Biomed Eng Online       Date:  2019-12-03       Impact factor: 2.819

2.  Corneal Biomechanics in Ectatic Diseases: Refractive Surgery Implications.

Authors:  Renato Ambrósio; Fernando Faria Correia; Bernardo Lopes; Marcella Q Salomão; Allan Luz; Daniel G Dawson; Ahmed Elsheikh; Riccardo Vinciguerra; Paolo Vinciguerra; Cynthia J Roberts
Journal:  Open Ophthalmol J       Date:  2017-07-31

3.  Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects.

Authors:  Robert Koprowski; Lei Tian
Journal:  J Ophthalmol       Date:  2017-01-30       Impact factor: 1.909

4.  In vivo measurement of regional corneal tangent modulus.

Authors:  Ying Hon; Guo-Zhen Chen; Shu-Hao Lu; David Cc Lam; Andrew Kc Lam
Journal:  Sci Rep       Date:  2017-11-02       Impact factor: 4.379

Review 5.  Accuracy of Machine Learning Assisted Detection of Keratoconus: A Systematic Review and Meta-Analysis.

Authors:  Ke Cao; Karin Verspoor; Srujana Sahebjada; Paul N Baird
Journal:  J Clin Med       Date:  2022-01-18       Impact factor: 4.241

6.  Impaired decidualization of human endometrial stromal cells from women with adenomyosis†.

Authors:  Yaoming Peng; Zhixing Jin; Haiou Liu; Congjian Xu
Journal:  Biol Reprod       Date:  2021-05-07       Impact factor: 4.285

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.