Abstract
BACKGROUND: Various surface coatings have been developed over the past decades to enhance fixation of cementless total knee arthroplasty (TKA). BIOFOAM(®) (MicroPort Orthopedics Inc., Arlington, TN, USA) is a novel cancellous titanium surface coating intended to increase both initial and long-term fixation. The purpose of this study was to investigate the early functional and radiographic outcomes of this coating used in a TKA application.Entities:
Keywords: ADVANCE; BIOFOAM; Cementless fixation; Total knee arthroplasty; Total knee replacement
Mesh:
Substances:
Year: 2015 PMID: 26410423 PMCID: PMC4819930 DOI: 10.1007/s12306-015-0382-z
Source DB: PubMed Journal: Musculoskelet Surg ISSN: 2035-5114
Fig. 1ADVANCE® BIOFOAM® tibial component is shown. The 100× magnification displays the roughened and porous surface of BIOFOAM. These components feature a Morse taper that allows for the attachment of various stem options
Demographics for included subjects
| Age | 64 years (range 43.0–87.0) |
| Male/female | 44 (42.3 %)/60 (57.7 %) |
| BMI | 32.0 kg/m2 (range 22.8–44.7 kg/m2) |
| Primary indications | |
| Degenerative osteoarthritis | 102 (98.1 %) |
| Avascular necrosis | 2 (1.9 %) |
Fig. 2Tibial component radiographic zones used to evaluate radiolucencies
Mean Knee Society Scores and range of motion values at each follow-up interval
| Mean ± Std. dev. |
| |
|---|---|---|
| Knee Society Scores | ||
| Preoperative | 53.6 ± 15.2 | – |
| 6 months | 87.2 ± 14.3* | <0.001 |
| 12 months | 90.0 ± 10.3* | <0.001 |
| 24 months | 93.3 ± 10.0* | <0.001 |
| Knee Society Function Scores | ||
| Preoperative | 52.6 ± 15.2 | – |
| 6 months | 83.9 ± 17.0* | <0.001 |
| 12 months | 87.0 ± 16.3* | <0.001 |
| 24 months | 87.5 ± 18.5* | <0.001 |
| Maximum flexion (°) | ||
| Preoperative | 110.3 ± 15.1 | – |
| 6 months | 111.5 ± 13.1 | 0.05 |
| 12 months | 114.8 ± 11.4* | 0.008 |
| 24 months | 116.8 ± 11.4* | <0.001 |
* Represents a statistically significant difference from preoperative values (p = 0.05)
Fig. 3Representative anterior–posterior and lateral radiographs of a well-fixed BIOFOAM-coated tibial component