| Literature DB >> 26404301 |
Rui Yatabe1, Junpei Noda2, Yusuke Tahara3, Yoshinobu Naito4, Hidekazu Ikezaki5, Kiyoshi Toko6,7.
Abstract
It is possible to evaluate the taste of foods or medicines using a taste sensor. The taste sensor converts information on taste into an electrical signal using several lipid/polymer membranes. A lipid/polymer membrane for bitterness sensing can evaluate aftertaste after immersion in monosodium glutamate (MSG), which is called "preconditioning". However, we have not yet analyzed the change in the surface structure of the membrane as a result of preconditioning. Thus, we analyzed the change in the surface by performing contact angle and surface zeta potential measurements, Fourier transform infrared spectroscopy (FTIR), X-ray photon spectroscopy (XPS) and gas cluster ion beam time-of-flight secondary ion mass spectrometry (GCIB-TOF-SIMS). After preconditioning, the concentrations of MSG and tetradodecylammonium bromide (TDAB), contained in the lipid membrane were found to be higher in the surface region than in the bulk region. The effect of preconditioning was revealed by the above analysis methods.Entities:
Keywords: FTIR; GCIB-TOF-SIMS; XPS; lipid; surface; taste sensor
Mesh:
Substances:
Year: 2015 PMID: 26404301 PMCID: PMC4610540 DOI: 10.3390/s150922439
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1The result of measurements for bitterness sample (0.01 vol% iso-α-acid) depending on the preconditioning time. The five measurements were carried out at each day. The error bar is standard deviation of the data. (a) CPA value; (b) The electrode potential (vs. Ag/AgCl at room temperature) when the electrode was immersed in the standard solution.
Figure 2The dependence of the surface zeta potential during preconditioning. For example, the red graph is data of the membrane without the preconditioning. The dark blue graph is data of the membrane with the preconditioning for 10 days.
Figure 3The contact angle of a water droplet with the lipid/polymer membrane.
Figure 4FTIR-RAS with transfer method for samples with and without the preconditioning for 7 days.
Figure 5The chemical structures of TDAB, NPOE, PVC and MSG.
The samples for XPS measurement. The first and second numbers are the amount of TDAB and NPOE, respectively. The last number is the preconditioning time. For example, sample 100-1-7 means that the amount of TDAB is 100 times of sample 1-1-7, which is used as a bitterness sensor.
| Sample Name | TDAB (times) | NPOE (times) | Preconditioning Time [ day ] |
|---|---|---|---|
| 0-0-0 | 0x | 0x | 0 |
| 0-0-7 | 0x | 0x | 7 |
| 0-1-0 | 0x | 1x | 0 |
| 0-1-7 | 0x | 1x | 7 |
| 1-1-0 | 1x | 1x | 0 |
| 1-1-7 | 1x | 1x | 7 |
| 100-1-0 | 100x | 1x | 0 |
| 100-1-7 | 100x | 1x | 7 |
| 1000-0-0 | 1000x | 0x | 0 |
Figure 6The result of XPS measurement. The upper graph is the result of wide scan. The lower graph is the result of narrow scan for signal of nitrogen 1s. The red lines are the results of samples without preconditioning. The blue lines are the result of samples with preconditioning. (a) Wide scan; (b) Narrow scan of Nitrogen 1s.
Figure 7The result of GCIB-TOF-SIMS measurement. Four lines indicate the data of samples with and without the preconditioning using positive and negative ion. ATTENTION: The intensities cannot compare with each other samples because of no normalization. (a) 69°C48H100N+ and 81Br of TDAB; (b) 23Na+ and 128C5H6NO3 of MSG; (c) 123C6H5NO2+ and 138C6H4NO3 of NPOE; (d) 252C14H22NO3+ and 251C14H21NO3 of NPOE.