| Literature DB >> 26393591 |
Rong-Chao Peng1,2,3, Wen-Rong Yan4, Ning-Ling Zhang5,6,7, Wan-Hua Lin8,9, Xiao-Lin Zhou10,11, Yuan-Ting Zhang12,13,14.
Abstract
Cardiovascular disease, like hypertension, is one of the top killers of human life and early detection of cardiovascular disease is of great importance. However, traditional medical devices are often bulky and expensive, and unsuitable for home healthcare. In this paper, we proposed an easy and inexpensive technique to estimate continuous blood pressure from the heart sound signals acquired by the microphone of a smartphone. A cold-pressor experiment was performed in 32 healthy subjects, with a smartphone to acquire heart sound signals and with a commercial device to measure continuous blood pressure. The Fourier spectrum of the second heart sound and the blood pressure were regressed using a support vector machine, and the accuracy of the regression was evaluated using 10-fold cross-validation. Statistical analysis showed that the mean correlation coefficients between the predicted values from the regression model and the measured values from the commercial device were 0.707, 0.712, and 0.748 for systolic, diastolic, and mean blood pressure, respectively, and that the mean errors were less than 5 mmHg, with standard deviations less than 8 mmHg. These results suggest that this technique is of potential use for cuffless and continuous blood pressure monitoring and it has promising application in home healthcare services.Entities:
Keywords: blood pressure; cross-validation; heart sound; smartphone; support vector machine
Mesh:
Year: 2015 PMID: 26393591 PMCID: PMC4610503 DOI: 10.3390/s150923653
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Schematic diagram for heart sound acquisition using a smartphone. The stethoscope is used to enhance the acoustic wave and the microphone on the earphone line is fixed in the hollow tube of the stethoscope.
Figure 2The procedure of the cold pressor test.
Figure 3Identification of the second heart sound using the Shannon energy envelope. (a) The normalized heart sound signal; (b) The normalized average Shannon energy; (c) The first heart sounds and the second heart sounds marked with squares and circles, respectively.
Figure 4The second heart sound and its frequency spectrum.
Figure 5Spectral change of the second heart sound with blood pressure variation. (a) Normalized Fourier spectrum of the second heart sound with systolic blood pressure increasing. (b) Normalized Fourier spectrum of the second heart sound when the systolic blood pressure was 140, 150, 160 and 170 mmHg.
Figure 6Variation of predicted values and measured values for systolic blood pressure (SBP), diastolic blood pressure (DBP) and mean blood pressure (MBP) during the cold pressor test. The vertical lines indicate the beginning and the end of the cold stimulus.
Figure 7Correlation analysis between predicted values and measured values for (a) Systolic blood pressure (SBP); (b) Diastolic blood pressure (DBP); (c) Mean blood pressure (MBP). In each plot, the diagonal is the linear regression line; r, Pearson correlation coefficient.
Figure 8Comparison of predicted values and measured values for each subject. (a) Correlation coefficients (CC); The presented data were all statistically significant (p < 0.05); (b) Mean absolute error (MAE); (c) Mean error (ME); (d) Standard deviation (SD). SBP, systolic blood pressure; DBP, diastolic blood pressure; MBP, mean blood pressure.
Distributions of the statistical parameters among all the subjects.
| Parameter | Maximum | Median | Minimum | Mean |
|---|---|---|---|---|
| CCSBP | 0.981 | 0.707 | 0.386 | 0.707 |
| CCDBP | 0.923 | 0.716 | 0.358 | 0.712 |
| CCMBP | 0.996 | 0.742 | 0.567 | 0.748 |
| MAESBP(mmHg) | 7.472 | 3.846 | 1.050 | 4.339 |
| MAEDBP(mmHg) | 5.472 | 3.040 | 1.767 | 3.171 |
| MAEMBP(mmHg) | 6.101 | 3.459 | 0.585 | 3.480 |
| MESBP(mmHg) | 1.231 | −0.108 | −2.494 | −0.204 |
| MEDBP(mmHg) | 0.496 | −0.174 | −1.190 | −0.274 |
| MEMBP(mmHg) | 0.463 | −0.247 | −1.490 | −0.357 |
| SDSBP(mmHg) | 10.708 | 5.452 | 2.815 | 6.121 |
| SDDBP(mmHg) | 7.488 | 4.225 | 2.878 | 4.471 |
| SDMBP(mmHg) | 8.383 | 4.819 | 1.014 | 4.961 |
CC, Pearson correlation coefficient; MAE, mean absolute error; ME, mean error; SD, standard deviation; SBP, systolic blood pressure; DBP, diastolic blood pressure; MBP, mean blood pressure.