Self-Calibration of Cone-Beam CT Geometry Using 3D-2D Image Registration: Development and Application to Task-Based Imaging with a Robotic C-Arm.

S Ouadah1, J W Stayman1, G Gang1, A Uneri2, T Ehtiati3, J H Siewerdsen4.   

Abstract

PURPOSE: Robotic C-arm systems are capable of general noncircular orbits whose trajectories can be driven by the particular imaging task. However obtaining accurate calibrations for reconstruction in such geometries can be a challenging problem. This work proposes a method to perform a unique geometric calibration of an arbitrary C-arm orbit by registering 2D projections to a previously acquired 3D image to determine the transformation parameters representing the system geometry.
METHODS: Experiments involved a cone-beam CT (CBCT) bench system, a robotic C-arm, and three phantoms. A robust 3D-2D registration process was used to compute the 9 degree of freedom (DOF) transformation between each projection and an existing 3D image by maximizing normalized gradient information with a digitally reconstructed radiograph (DRR) of the 3D volume. The quality of the resulting "self-calibration" was evaluated in terms of the agreement with an established calibration method using a BB phantom as well as image quality in the resulting CBCT reconstruction.
RESULTS: The self-calibration yielded CBCT images without significant difference in spatial resolution from the standard ("true") calibration methods (p-value >0.05 for all three phantoms), and the differences between CBCT images reconstructed using the "self" and "true" calibration methods were on the order of 10-3 mm-1. Maximum error in magnification was 3.2%, and back-projection ray placement was within 0.5 mm.
CONCLUSION: The proposed geometric "self" calibration provides a means for 3D imaging on general non-circular orbits in CBCT systems for which a geometric calibration is either not available or not reproducible. The method forms the basis of advanced "task-based" 3D imaging methods now in development for robotic C-arms.

Entities:  

Keywords:  cone-beam CT; image-guided interventions; imaging task; task-based imaging

Year:  2015        PMID: 26388661      PMCID: PMC4574500          DOI: 10.1117/12.2082538

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  5 in total

1.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

2.  Accurate technique for complete geometric calibration of cone-beam computed tomography systems.

Authors:  Youngbin Cho; Douglas J Moseley; Jeffrey H Siewerdsen; David A Jaffray
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

3.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT.

Authors:  M J Daly; J H Siewerdsen; Y B Cho; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

4.  Automatic localization of vertebral levels in x-ray fluoroscopy using 3D-2D registration: a tool to reduce wrong-site surgery.

Authors:  Y Otake; S Schafer; J W Stayman; W Zbijewski; G Kleinszig; R Graumann; A J Khanna; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2012-08-03       Impact factor: 3.609

5.  Robust 3D-2D image registration: application to spine interventions and vertebral labeling in the presence of anatomical deformation.

Authors:  Yoshito Otake; Adam S Wang; J Webster Stayman; Ali Uneri; Gerhard Kleinszig; Sebastian Vogt; A Jay Khanna; Ziya L Gokaslan; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2013-11-18       Impact factor: 3.609

  5 in total
  2 in total

1.  Marker-free motion correction in weight-bearing cone-beam CT of the knee joint.

Authors:  M Berger; K Müller; A Aichert; M Unberath; J Thies; J-H Choi; R Fahrig; A Maier
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

2.  Hybrid spectral CT reconstruction.

Authors:  Darin P Clark; Cristian T Badea
Journal:  PLoS One       Date:  2017-07-06       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.