Literature DB >> 26372465

In Vitro Dosing Performance of the ELLIPTA® Dry Powder Inhaler Using Asthma and COPD Patient Inhalation Profiles Replicated with the Electronic Lung (eLung™).

Melanie Hamilton1, Richard Leggett1, Cheng Pang1, Stephen Charles1, Ben Gillett1, David Prime1.   

Abstract

BACKGROUND: To evaluate the in vitro dose delivery characteristics of approved asthma and chronic obstructive pulmonary disease (COPD) therapies delivered via the ELLIPTA(®) dry powder inhaler across inhalation endpoints representative of the target patient population, using the Electronic Lung (eLung™) to replicate inhaler-specific patient inhalation profiles that were previously recorded in vivo.
METHODS: Selected profiles, representative of the range of inhalation endpoints achieved by patients with all severities of asthma and COPD, were replicated using the eLung breathing simulator in conjunction with an oropharyngeal cast. A Next Generation Impactor was coupled to the eLung to determine the aerodynamic particle size distribution of the ex-throat dose (ETD) of asthma and COPD therapies delivered via the ELLIPTA inhaler. Delivered dose (DD), ETD, and fine particle dose (FPD; defined as a mass of active substance less than 5 μm) were determined for fluticasone furoate (FF)/vilanterol (VI) 100/25 μg and 200/25 μg (asthma and COPD), umeclidinium (UMEC)/VI 62.5/25 μg (COPD only), FF 100 μg and 200μg monotherapy (asthma only), and UMEC 62.5 μg monotherapy (COPD only).
RESULTS: Inhalation profiles replicated by eLung covered a wide range of peak inspiratory flow rates (41.6-136.9 L/min), pressure drops (1.2-13.8 kPa), and inhaled volumes through the inhaler (0.7-4.2L). DD was consistent across the range of patient representative inhalation parameters for all components (FF, VI, and UMEC) of each therapy assessed; although ETD and FPD were also generally consistent, some small variation was observed. Dose delivery was consistent for each of the components, whether delivered as mono- or combination therapy.
CONCLUSIONS: The in vitro performance of the ELLIPTA inhaler has been demonstrated for the delivery of FF/VI, UMEC/VI, FF monotherapy, and UMEC monotherapy. Across a range of inspiratory profiles, DD was consistent, while ETD and FPD showed little flow dependency.

Entities:  

Keywords:  COPD; asthma; delivered dose; dry powder inhaler; electronic lung; fine particle dose; inhalation profiles

Mesh:

Substances:

Year:  2015        PMID: 26372465      PMCID: PMC4685503          DOI: 10.1089/jamp.2015.1225

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


Introduction

Dry powder inhalers (DPIs) are used for the delivery of inhaled medications in the management of both asthma and chronic obstructive pulmonary disease (COPD).( The ELLIPTA® DPI (GSK, Hertfordshire, UK; ELLIPTA® is a trademark of the GSK group of companies) is available in either a single-strip or two-strip configuration, both of which have moderate resistance to inspiratory effort. The two-strip configuration (typical specific resistance: 0.0286 kPa0.5[L/min]−1) is currently used for the delivery of the asthma and COPD combination therapy fluticasone furoate (FF)/vilanterol (VI), and for the COPD combination therapy umeclidinium (UMEC)/VI. The single-strip configuration (typical specific resistance: 0.0258 kPa0.5[L/min]−1) is used for the delivery of FF monotherapy (asthma only), and UMEC monotherapy (COPD only). To use any DPI efficiently, patients must be able to apply sufficient inspiratory effort to emit the dose from the inhaler, and to disperse the powder formulation to create a sufficient mass of particles within the respirable size range (fine particle dose [FPD]; typically considered as mass of particles <5 μm aerodynamic diameter).( Some asthma and COPD patients may not be able to achieve the required inspiratory flow rate to use some DPIs efficiently,( particularly those with severely impaired lung function, for example, in very-severe COPD.( It is therefore important to assess the inhaler-specific inhalation characteristics of the target population to confirm whether patients can achieve the minimum inspiratory effort required to use the inhaler efficiently and consistently. Furthermore, consistent performance across the range of inhalation parameters of the target population ensures that these variations do not influence or control the dose delivery and the resultant respirable mass of particles, as it has been suggested that there are correlations between FPD determined in vitro, lung deposition, and clinical response.( The Electronic Lung (eLung™: a trademark of the GSK group of companies) is a breathing simulator designed for the characterization of inhalation devices,( and is capable of replicating patient inhalation profiles recorded in vivo using an inhalation profile recorder. It can be used in combination with an oropharyngeal cast, and coupled with a Cascade Impactor to allow dose delivery characteristics (including aerodynamic particle size distribution) to be assessed in vitro, using more patient-relevant conditions than standard impactor testing. To determine whether asthma and COPD patients are capable of achieving the required inspiratory flow to use the ELLIPTA inhaler, the maximal effort inhaler-specific inhalation profiles for the target population have been characterized, and these are described separately (Prime et al., J Aerosol Med Pulm Drug Deliv, in press). In this second publication, the objective was to evaluate the in vitro dose delivery characteristics of approved therapies delivered via the ELLIPTA inhaler, using the eLung to replicate patient-specific inhalation profiles representative of patients with asthma or COPD, categorized into subgroups of varying disease severity.

Materials and Methods

Maximal effort inhaler-specific inhalation profiles were recorded and characterized from patients with mild to severe asthma or mild to very-severe COPD in two studies (Study 1: GSK study RES113817, clinicaltrials.gov NCT01345266; Study 2: RES117178, NCT02076269). Patient demographics and the results of these evaluations are described separately (Prime et al., submitted in parallel). Generally, the patients with asthma tended to be younger than those with COPD (median age range of 40.9–46.2 vs. 61.6–64.4, respectively), while those with COPD were more likely to be male than female, but no significant differences in terms of age or BMI were seen within each of the asthma or COPD groups across the disease severities. The aerodynamic particle size distribution of the fraction of the dose passing beyond the throat (ex-throat dose [ETD]) was assessed for each of the following approved therapies, where the stated dose is equal to the nominal blister content: FF/VI 100/25 μg (label claim delivered dose: 92/22 μg), FF/VI 200/25 μg (label claim delivered dose: 184/22 μg),[11] UMEC/VI 62.5/25 μg (label claim delivered dose: 55/22 μg),[12] FF 100 μg monotherapy (label claim delivered dose: 90 μg), FF 200 μg monotherapy (label claim delivered dose: 182 μg),[13] or UMEC 62.5 μg monotherapy (label claim delivered dose: 55 μg).[14] FF/VI and UMEC/VI were delivered using the two-strip configuration of the ELLIPTA inhaler, whereas FF monotherapy and UMEC monotherapy were delivered using the single-strip configuration. For each therapy, five inhalation profiles were sourced for the relevant device configuration and disease state(s) by sorting the appropriate full dataset of profiles by peak inspiratory flow rate (PIFR), then selecting individual profiles to represent the absolute minimum and maximum PIFRs, plus three intermediate values at the median and interquartile points. These inhalation profiles were replicated using the eLung. The eLung was used in combination with a specific throat cast based on a previous imaging study (PP2),[15] which was considered more representative of human anatomy than the standard metal throat that would typically be used during quality control testing of inhaled products. While the general population will vary in oropharyngeal anatomical structure, the throat cast selected was deemed suitable for use in these experiments as the average cross-sectional area closely matched the median observed in imaging studies (RES113817, RES117178; GSK data on file) of healthy individuals, as well as asthma and COPD patients. The inlet was modified with a rubber mouthpiece adaptor to enable coupling of the ELLIPTA inhaler to the cast. Prior to use, the internal surface of the anatomical cast was coated with an ethanolic Brij solution in glycerol. This was done by placing a volume of coating solution within the throat, sealing the inlet and exit, inverting in all orientations to ensure absolute coverage, and then allowing draining of any surplus solution from both inlet and exit ends prior to use. A Next Generation Impactor (NGI) was coupled to the eLung, to enable the aerodynamic particle size distribution of the ETD to be characterized. NGI cups were coated with a polyethylene glycol solution to prevent particle bounce. In the companion report to this one (Grant et al, J Aerosol Med Pulm Drug Deliv, in press), a constant flow rate of 30, 60, or 90 L/min (4 L inhaled volume) was used in a standard impactor set-up. In this study, it was desirable to replicate complete patient inhalation profiles and to use the NGI just to characterize the resultant ETD. In order to maintain NGI stage effective cut-off diameters, the impactor was operated at a constant flow rate of 60 L/min; using this set-up, the eLung equipment permits separation of the flow conditions at which dose emission and dose characterization take place. Dose emission takes place under variable flow conditions, representative of the patient inhaling through the inhaler, whilst constant flow is maintained through the impactor, which is a requirement of impactors to maintain effective cut-off diameters of the impactor stages. A flow rate of 60 L/min through the impactor was selected to be consistent with the GSK registered flow rate used during standard aerodynamic particle size distribution testing. Each dose was actuated from the ELLIPTA inhaler as the eLung breathing simulator replicated the selected inhalation profile. Multiple doses (six for FF monotherapy and FF/VI; five for UMEC monotherapy and UMEC/VI) were aerosolized into the eLung set-up during one experiment replicate. To assess the total delivered dose (DD), ETD, and FPD, individual parts of the eLung/impactor apparatus were rinsed with a suitable solvent to recover the deposited mass. The solutions were assessed using high performance liquid chromatography (HPLC), performed on Agilent 1100 Series equipment. This process was repeated in triplicate for each product and inhalation profile combination. FPD was interpolated from NGI stage deposition data using a validated template.

Results

There was wide variation between the inhalation characteristics in the selected inhalation profiles: the nominal PIFRs across both the single-strip and two-strip configurations of the ELLIPTA inhaler ranged from 41.6 (very-severe COPD) to 136.9 L/min (moderate asthma); peak pressure drops ranged from 1.2 (very-severe COPD) to 13.8 kPa (moderate asthma); and nominal inhaled volume ranged from 0.7 L (very-severe COPD) to 4.2 L (mild COPD). Overall, inhalation characteristics for the profiles that were recorded from patients within the same disease severity subgroups were similar between the single-strip and two-strip configurations where these comparisons could be made (Fig. 1; Table 1). Although the data sets for each product and profile combination were relatively small (n = 3), low variability was observed.

Flow rate versus time for each patient representative maximal effort inhalation profile replicated for the assessments of (a) FF/VIa, (b) FFb, (c) UMEC/VIa, and (d) UMECb. COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. aDelivered via the two-strip ELLIPTA inhaler. bDelivered via the single-strip ELLIPTA inhaler.

1.

Summary of Maximal Effort Inhalation Profiles Used for the Assessment of Each Therapy

 Profile 1 (minimum inspiratory effort)Profile 2Profile 3 (median inspiratory effort)Profile 4Profile 5 (maximum inspiratory effort)
FF/VI delivered via the two-strip ELLIPTA inhaler
 Nominal PIFR (L/min)43.567.482.599.9129.9
 Peak pressure drop (kPa)1.63.75.68.213.8
 Nominal inhaled volume (L)0.81.82.12.43.2
 Nominal inhalation time (s)1.62.72.52.32.5
 Nominal average inhalation flow rate (L/min)29.340.450.663.075.8
 Nominal acceleration rate (L/min/s)158152289312647
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
FF delivered via the single-strip ELLIPTA inhaler
 Nominal PIFR (L/min)71.291.8104.2120.5136.9
 Peak pressure drop (kPa)3.45.67.29.712.5
 Nominal inhaled volume (L)2.02.12.12.73.7
 Nominal inhalation time (s)2.72.41.91.82.9
 Nominal average inhalation flow rate (L/min)44.652.464.388.476.1
 Nominal acceleration rate (L/min/s)223351482420705
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
UMEC/VI delivered via the two-strip ELLIPTA inhaler
 Nominal PIFR (L/min)43.563.982.4103.9123.6
 Peak pressure drop (kPa)1.63.35.68.812.5
 Nominal inhaled volume (L)0.80.92.33.54.2
 Nominal inhalation time (s)1.61.53.13.23.3
 Nominal average inhalation flow rate (L/min)29.339.044.566.977.3
 Nominal acceleration rate (L/min/s)149228375689468
Subgroup sourceVery-severe COPDSevere COPDSevere COPDModerate COPDMild COPD
UMEC delivered via the single-strip ELLIPTA inhaler
 Nominal PIFR (L/min)41.663.284.8107.3128.9
 Peak pressure drop (kPa)1.22.74.87.711.1
 Nominal inhaled volume (L)0.71.12.03.34.1
 Nominal inhalation time (s)1.61.82.33.73.2
 Nominal average inhalation flow rate (L/min)25.835.752.253.677.9
 Nominal acceleration rate (L/min/s)91177226641379
Subgroup sourceVery-severe COPDModerate COPDMild COPDModerate COPDMild COPD

COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

Flow rate versus time for each patient representative maximal effort inhalation profile replicated for the assessments of (a) FF/VIa, (b) FFb, (c) UMEC/VIa, and (d) UMECb. COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. aDelivered via the two-strip ELLIPTA inhaler. bDelivered via the single-strip ELLIPTA inhaler. Summary of Maximal Effort Inhalation Profiles Used for the Assessment of Each Therapy COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

DD, ETD, and FPD

Mean DD, ETD, and FPD (% nominal blister content) for each component of the therapies tested are presented in Tables 2, 3, and 4, respectively. There was little evidence of trends in DD, ETD, or FPD for each component across the inhalation profiles tested, with DD close to that of the stated label claims (Fig. 2). For the FF and UMEC components, DD, ETD, and FPD were consistent regardless of whether the component was delivered as part of a combination therapy or as a monotherapy. Similarly, the DD, ETD, and FPD of VI appeared to be unaffected regardless of whether it was delivered as a component of FF/VI or UMEC/VI. The DD (as % nominal blister content) of FF and VI tended to be slightly higher than those of UMEC, but this did not translate to differences in FPD; the FPDs for UMEC and VI were similar, while the FPD of FF was slightly lower than for either UMEC or VI.
2.

Mean DD for Each Component of Each of the Therapies Investigated

 DD (% nominal blister content)
Component expressed as nominal blister contentProfile 1Profile 2Profile 3Profile 4Profile 5
FF/VI 100/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 100 μg85.7 (85.5–85.9)86.5 (85.6–87.6)87.7 (86.4–88.8)87.2 (85.8–88.2)88.5 (86.9–91.2)
 VI 25 μg83.5 (80.3–87.3)86.3 (84.6–87.2)88.0 (85.3–90.1)87.3 (85.2–88.4)87.4 (83.3–89.9)
FF/VI 200/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 200 μg85.3 (84.4–86.2)88.1 (87.3–89.0)87.8 (86.5–88.8)88.5 (87.6–90.2)89.7 (88.0–91.7)
 VI 25 μg84.2 (81.8–87.2)85.8 (83.2–87.2)87.7 (84.7–89.2)86.8 (84.9–88.0)86.4 (84.2–88.0)
UMEC/VI 62.5/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.563.982.4103.9123.6
Subgroup sourceVery-severe COPDSevere COPDSevere COPDModerate COPDMild COPD
 UMEC 62.5 μg77.8 (75.5–79.8)82.8 (79.8–85.7)82.3 (81.3–83.2)79.9 (75.5–83.2)82.6 (81.5–84.3)
 VI 25 μg82.1 (80.1–83.3)82.3 (81.0–84.2)82.2 (81.3–82.8)81.8 (79.5–83.6)85.0 (83.7–86.1)
FF 100μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 100 μg84.0 (82.1–85.6)85.4 (84.3–86.8)87.1 (83.3–89.1)83.9 (82.3–87.0)84.1 (83.3–85.2)
FF 200μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 200 μg87.0 (85.6–89.3)88.8 (87.3–90.0)86.6 (85.4–87.3)88.2 (87.3–89.0)89.0 (87.8–89.8)
UMEC 62.5μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)41.663.284.8107.3128.9
Subgroup sourceVery-severe COPDModerate COPDMild COPDModerate COPDMild COPD
 UMEC 62.5 μg76.9 (75.0–78.3)79.3 (78.4–80.5)79.1 (78.4–79.7)80.8 (79.2–82.6)80.8 (78.6–83.1)

All data are mean (range) unless otherwise stated.

COPD, chronic obstructive pulmonary disease; DD, delivered dose; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

3.

Mean ETD for Each Component of Each of the Therapies Investigated

 ETD (% nominal blister content)
Component expressed as nominal blister contentProfile 1Profile 2Profile 3Profile 4Profile 5
FF/VI 100/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 100 μg28.0 (26.7–28.8)27.0 (25.5–27.7)26.6 (26.1–27.0)25.5 (25.1–25.8)23.0 (22.5–23.3)
 VI 25 μg33.4 (31.0–34.9)33.8 (32.0–36.1)35.0 (33.2–36.7)40.2 (38.4–41.2)40.1 (39.2–41.4)
FF/VI 200/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 200 μg27.7 (27.3–28.1)28.0 (27.9–28.1)26.3 (25.4–26.8)25.5 (24.9–26.1)23.7 (23.5–23.9)
 VI 25 μg34.5 (32.5–35.6)35.2 (34.3–36.5)38.6 (35.3–40.9)39.8 (38.9–41.5)40.8 (39.1–41.8)
UMEC/VI 62.5/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.563.982.4103.9123.6
Subgroup sourceVery-severe COPDSevere COPDSevere COPDModerate COPDMild COPD
 UMEC 62.5 μg38.0 (37.6–38.6)42.5 (41.7–44.1)38.6 (38.0–39.6)38.7 (38.0–39.9)40.3 (38.9–41.2)
 VI 25 μg32.6 (32.3–33.3)37.0 (36.4–37.9)34.3 (33.1–35.0)39.5 (38.5–40.6)42.1 (40.8–43.0)
FF 100 μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 100 μg24.7 (23.5–26.0)24.4 (23.9–24.8)23.3 (22.6–24.1)23.6 (22.5–24.8)22.2 (21.5–23.4)
FF 200 μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 200 μg25.4 (24.4–26.7)25.7 (25.1–26.8)23.1 (22.9–23.2)24.8 (23.6–25.4)22.0 (21.5–22.3)
UMEC 62.5 μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)41.663.284.8107.3128.9
Subgroup sourceVery-severe COPDModerate COPDMild COPDModerate COPDMild COPD
 UMEC 62.5 μg41.4 (40.2–43.3)42.4 (41.7–43.5)36.8 (35.5–38.9)31.5 (30.1–32.5)39.6 (38.8–40.0)

All data are mean (range) unless otherwise stated.

COPD, chronic obstructive pulmonary disease; ETD, ex-throat dose; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

4.

Mean FPD for Each Component of Each of the Therapies Investigated

 FPD (% nominal blister content)
Component expressed as nominal blister contentProfile 1Profile 2Profile 3Profile 4Profile 5
FF/VI 100/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 100 μg21.6 (20.2–22.4)20.5 (16.7–22.5)21.9 (21.7–22.1)20.2 (19.6–21.1)19.8 (19.2–20.1)
 VI 25 μg29.2 (27.2–30.8)28.8 (27.2–30.8)32.0 (30.4–33.2)35.2 (34.4–35.6)37.2 (36.4–38.8)
FF/VI 200/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.567.482.599.9129.9
Subgroup sourceVery-severe COPDModerate asthmaMild asthmaSevere asthmaModerate asthma
 FF 200 μg21.5 (21.0–21.9)22.2 (22.1–22.5)21.3 (21.2–21.6)21.2 (20.9–21.5)19.3 (18.2–20.2)
 VI 25 μg30.4 (28.8–31.2)32.0 (31.2–33.2)35.2 (32.0–36.8)37.2 (36.4–38.8)36.8 (34.0–38.8)
UMEC/VI 62.5/25 μg delivered via the two-strip ELLIPTA inhaler
Nominal PIFR (L/min)43.563.982.4103.9123.6
 UMEC 62.5 μg30.6 (29.9–31.2)36.0 (35.0–37.6)32.3 (31.7–33.3)31.4 (30.7–32.0)31.5 (29.8–32.3)
 VI 25 μg29.6 (28.8–30.4)34.8 (34.0–36.0)31.6 (30.0–32.4)32.8 (31.6–34.0)34.4 (32.0–35.6)
FF 100μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 100 μg20.8 (19.9–21.6)21.2 (20.9–21.5)20.6 (20.2–21.3)20.4 (19.5–21.5)18.9 (18.3–20.0)
FF 200μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)71.291.8104.2120.5136.9
Subgroup sourceSevere asthmaMild asthmaSevere asthmaModerate asthmaModerate asthma
 FF 200 μg21.3 (20.8–22.3)22.2 (21.5–23.0)20.6 (20.4–20.8)21.6 (20.3–22.3)18.8 (18.6–19.1)
UMEC 62.5μg delivered via the single-strip ELLIPTA inhaler
Nominal PIFR (L/min)41.663.284.8107.3128.9
Subgroup sourceVery-severe COPDModerate COPDMild COPDModerate COPDMild COPD
 UMEC 62.5 μg33.1 (32.2–34.6)35.8 (35.0–37.1)30.6 (29.4–32.6)24.2 (23.2–25.6)30.4 (30.2–30.6)

All data are mean (range) unless otherwise stated.

COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; FPD, fine particle dose; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

DD, ETD, and FPD (% nominal blister content) versus PIFR (L/min) for (a) FF delivered as a component of FF/VIa, (b) VI delivered as a component of FF/VIa, (c) UMEC delivered as a component of UMEC/VIa, (d) VI delivered as a component of UMEC/VIa, (e) FF 100 μg and 200 μgb, and (f) UMEC 62.5 μgb. DD, delivered dose; ETD, ex-throat dose; FF, fluticasone furoate; FPD, fine particle dose; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. aDelivered via the two-strip ELLIPTA inhaler. bDelivered via the single-strip ELLIPTA inhaler.

DD, ETD, and FPD (% nominal blister content) versus PIFR (L/min) for (a) FF delivered as a component of FF/VIa, (b) VI delivered as a component of FF/VIa, (c) UMEC delivered as a component of UMEC/VIa, (d) VI delivered as a component of UMEC/VIa, (e) FF 100 μg and 200 μgb, and (f) UMEC 62.5 μgb. DD, delivered dose; ETD, ex-throat dose; FF, fluticasone furoate; FPD, fine particle dose; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. aDelivered via the two-strip ELLIPTA inhaler. bDelivered via the single-strip ELLIPTA inhaler. Mean DD for Each Component of Each of the Therapies Investigated All data are mean (range) unless otherwise stated. COPD, chronic obstructive pulmonary disease; DD, delivered dose; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. Mean ETD for Each Component of Each of the Therapies Investigated All data are mean (range) unless otherwise stated. COPD, chronic obstructive pulmonary disease; ETD, ex-throat dose; FF, fluticasone furoate; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol. Mean FPD for Each Component of Each of the Therapies Investigated All data are mean (range) unless otherwise stated. COPD, chronic obstructive pulmonary disease; FF, fluticasone furoate; FPD, fine particle dose; PIFR, peak inspiratory flow rate; UMEC, umeclidinium; VI, vilanterol.

Discussion

In this article, we have described how the eLung apparatus and an anatomically representative throat cast were used to assess the dose delivery characteristics of FF/VI, UMEC/VI, FF, and UMEC delivered via the ELLIPTA inhaler. This method enabled these assessments to be performed in a more realistic manner than standard in vitro methodologies allow, and over a greater range of inhalation endpoints, by replicating maximal effort inhalation profiles collected from the target patient population. The set-up also allowed for exploration of a wider range of PIFRs than could be achieved with standard NGI testing, as well as achieving variable flow through the device. These in vitro assessments suggest that patients are capable of achieving the inspiratory effort required for consistent performance of the ELLIPTA inhaler when used for the delivery of the asthma and/or COPD therapies FF/VI, UMEC/VI, FF monotherapy, and UMEC monotherapy, at currently approved doses. The inhalation profiles replicated during the eLung assessments were chosen to represent patients for whom that therapy is indicated. The replicated profiles were selected based on PIFR to represent the absolute range of previously observed inhalation parameters in asthma and COPD of all severities. From the previous assessment of inhalation profiles (Prime et al., submitted in parallel), it was established that very-severe COPD was associated with the largest negative impact on PIFR and other inhalation endpoints, and therefore it is these patients who are most likely not to be able to achieve the required inspiratory effort for efficient and consistent dosing when using a DPI. Therefore, an inhalation profile representative of a patient with very-severe COPD was used in the assessments of all COPD relevant therapies (FF/VI, UMEC/VI, and UMEC monotherapy). For FF/VI, the other four profiles used were derived from patients with asthma, although there is overlap between the inhalation profiles and the disease severity subgroups represented (e.g., a ‘moderate asthma’ profile may also be representative of the inhalation endpoints that could be achieved by a patient with mild COPD). Despite the large differences in PIFR and other inhalation parameters for very-severe COPD patients compared with asthma patients, the DD, ETD, and FPD were similar for each of the profiles. For UMEC/VI and UMEC monotherapy, which are only indicated in COPD, the inhalation profiles were representative of patients with COPD of all severities (including very-severe) so it was not possible to make the same comparison. Despite the trend for PIFR (and other parameters) to significantly decrease with increasing disease severity, no trends indicating that DD, ETD, and FPD decreased with increasing severity were identified for any of the therapies studied. It has been shown that the flow acceleration rate may have a significant effect on the performance of some DPIs, independently of the PIFR achieved.( The flow acceleration rate varied significantly between the profiles used in this study, however, in general no effect on DD, and little effect on ETD or FPD was observed. For the individual FF and UMEC components of each of the therapies tested, the DD, ETD, and FPD were consistent whether they were delivered alone or simultaneously with another compound as part of a combination therapy. Although there was some slight variation between FF, UMEC, and VI for the DD, ETD, and FPD, these differences were not considered to be clinically important; this finding is supported by observations from Phase III clinical trials in which clinical efficacy was demonstrated, even in very-severe COPD patients, for FF/VI,( and UMEC/VI and UMEC monotherapy( delivered via the ELLIPTA inhaler. Efficacy of FF monotherapy has also been demonstrated, but only in asthma patients as this treatment is not indicated in COPD.( Although the data set was not large enough to perform formal statistical analysis, these findings provide a useful indication that the ELLIPTA inhaler performs consistently across a wide range of patient relevant inhalation profiles and should therefore be suitable for the majority of patients. This is supported by the fact that there was no tail off in performance for any of the profiles replicated, including that of the very-severe COPD patient group with the absolute minimum observed PIFR. It is further supported by previous in vitro dose delivery assessments, which showed that the delivered dose from the ELLIPTA inhaler for all products assessed was close to the stated label claim at flow rates of 30, 60, and 90 L/min under standard impactor conditions (Grant et al., J Aerosol Med Pulm Drug Deliv, in press). The results of in vitro dose delivery characteristics of the ELLIPTA inhaler described herein were generated through replication of inhalation profiles previously recorded in a study using a maximal effort instruction for comparison with inhaler-independent lung function assessments (Prime et al., submitted in parallel). It is recognized that the ‘long, steady, deep’ inhalation instruction is described in the patient information leaflet. Despite this difference, our assessments included a wide range of inhalation profile parameters and therefore the results should still be relevant to a wide range of adult patients with mild to severe asthma and mild to very-severe COPD. The selection of inhalation profiles based on PIFR is relevant because of the variety of profiles covered, and the fact that these profiles, which were obtained in vivo, will also cover a range of other inhalation parameters that are affected by disease severity (e.g., inhaled volume and acceleration rate). It should also be noted that this study did not replicate any profiles previously recorded from patients experiencing exacerbations or in remission from an exacerbation. Future studies may focus on representing patients with inhalation parameters that may differ from those already investigated, and may include pediatric subjects for whom these results are not applicable. Finally, it should also be noted that the effectiveness of any inhaled therapy is not just dependent on the dose delivery characteristics of the inhaler, but also on patient inhaler preference, ease of use, and inhaler handling, which have been reported elsewhere for the ELLIPTA inhaler.( As noted previously, this study only performed three replicates for each of the inhalation profiles. This number could perhaps be increased in any future experiments to ensure a wider range of data. In addition, as this study included data for only one example of throat geometry, a wider study exploring the impact of variability in this area may be worthy of future investigation, particularly with regard to the potential effect on throat size of imaging obtained in a supine versus upright position.

Conclusions

The in vitro dosing performance of the ELLIPTA DPI has been established through replication of a range of patients' inhaler-specific maximal effort inhalation profiles, which was achieved using the eLung breathing simulator, in conjunction with an anatomically representative oropharyngeal cast of average size. Using a range of inhalation parameters representative of patients with all severities of asthma or COPD (PIFR: 41.6–136.9 L/min; inhaled volume: 0.7–4.2 L), dose delivery of all therapies was consistent and reproducible with little flow dependency.
  20 in total

1.  Review of dry powder inhalers.

Authors: 
Journal:  Adv Drug Deliv Rev       Date:  1997-06-09       Impact factor: 15.470

Review 2.  New propellant-free technologies under investigation.

Authors:  M Dolovich
Journal:  J Aerosol Med       Date:  1999

Review 3.  Dry powder inhalers: an overview.

Authors:  Paul J Atkins
Journal:  Respir Care       Date:  2005-10       Impact factor: 2.258

4.  Efficacy and safety of once-daily umeclidinium/vilanterol 62.5/25 mcg in COPD.

Authors:  J F Donohue; M R Maleki-Yazdi; S Kilbride; R Mehta; C Kalberg; A Church
Journal:  Respir Med       Date:  2013-07-02       Impact factor: 3.415

5.  Once-daily inhaled fluticasone furoate and vilanterol versus vilanterol only for prevention of exacerbations of COPD: two replicate double-blind, parallel-group, randomised controlled trials.

Authors:  Mark T Dransfield; Jean Bourbeau; Paul W Jones; Nicola A Hanania; Donald A Mahler; Jørgen Vestbo; Andrew Wachtel; Fernando J Martinez; Frank Barnhart; Lisa Sanford; Sally Lettis; Courtney Crim; Peter M A Calverley
Journal:  Lancet Respir Med       Date:  2013-04-12       Impact factor: 30.700

6.  Can all patients with COPD use the correct inhalation flow with all inhalers and does training help?

Authors:  Raid A M Al-Showair; Walid Y Tarsin; Khaled H Assi; Stanley B Pearson; Henry Chrystyn
Journal:  Respir Med       Date:  2007-07-12       Impact factor: 3.415

7.  Efficacy and safety of umeclidinium plus vilanterol versus tiotropium, vilanterol, or umeclidinium monotherapies over 24 weeks in patients with chronic obstructive pulmonary disease: results from two multicentre, blinded, randomised controlled trials.

Authors:  Marc Decramer; Antonio Anzueto; Edward Kerwin; Thomas Kaelin; Nathalie Richard; Glenn Crater; Maggie Tabberer; Stephanie Harris; Alison Church
Journal:  Lancet Respir Med       Date:  2014-05-14       Impact factor: 30.700

8.  Once-daily umeclidinium/vilanterol 125/25 mcg in COPD: a randomized, controlled study.

Authors:  Bartolome Celli; Glenn Crater; Sally Kilbride; Rashmi Mehta; Maggie Tabberer; Chris J Kalberg; Alison Church
Journal:  Chest       Date:  2014-05       Impact factor: 9.410

9.  Qualitative assessment of attributes and ease of use of the ELLIPTA™ dry powder inhaler for delivery of maintenance therapy for asthma and COPD.

Authors:  Henrik Svedsater; Peter Dale; Karl Garrill; Richard Walker; Mark W Woepse
Journal:  BMC Pulm Med       Date:  2013-12-07       Impact factor: 3.317

10.  Efficacy and safety of fluticasone furoate 100 μg and 200 μg once daily in the treatment of moderate-severe asthma in adults and adolescents: a 24-week randomised study.

Authors:  Ashley Woodcock; Jan Lötvall; William W Busse; Eric D Bateman; Sally Stone; Anna Ellsworth; Loretta Jacques
Journal:  BMC Pulm Med       Date:  2014-07-09       Impact factor: 3.317

View more
  11 in total

Review 1.  Inhaled Umeclidinium in COPD Patients: A Review and Meta-Analysis.

Authors:  Roy A Pleasants; Tiansheng Wang; Jinming Gao; Huilin Tang; James F Donohue
Journal:  Drugs       Date:  2016-03       Impact factor: 9.546

2.  The Repeatability of Inspiration Performance Through Different Inhalers in Patients with Chronic Obstructive Pulmonary Disease and Control Volunteers.

Authors:  Tamas Erdelyi; Zsofia Lazar; Balazs Odler; Lilla Tamasi; Veronika Müller
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2020-05-28       Impact factor: 2.849

3.  Efficacy of Umeclidinium/Vilanterol in Elderly Patients with COPD: A Pooled Analysis of Randomized Controlled Trials.

Authors:  Riju Ray; Lee Tombs; Michael J Asmus; Isabelle Boucot; David A Lipson; Chris Compton; Ian Naya
Journal:  Drugs Aging       Date:  2018-07       Impact factor: 3.923

4.  Randomized dose-finding study of batefenterol via dry powder inhaler in patients with COPD.

Authors:  Courtney Crim; Michael L Watkins; Eric D Bateman; Gregory J Feldman; Isabelle Schenkenberger; Edward M Kerwin; Catriona Crawford; Krishna Pudi; Shuyen Ho; Charlotte Baidoo; Ramiro Castro-Santamaria
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2019-03-08

Review 5.  Inhaler Devices for Delivery of LABA/LAMA Fixed-Dose Combinations in Patients with COPD.

Authors:  Anthony D'Urzo; Kenneth R Chapman; James F Donohue; Peter Kardos; M Reza Maleki-Yazdi; David Price
Journal:  Pulm Ther       Date:  2019-03-13

Review 6.  Lung Deposition and Inspiratory Flow Rate in Patients with Chronic Obstructive Pulmonary Disease Using Different Inhalation Devices: A Systematic Literature Review and Expert Opinion.

Authors:  Adolfo Baloira; Araceli Abad; Antonia Fuster; Juan Luis García Rivero; Patricia García-Sidro; Eduardo Márquez-Martín; Marta Palop; Néstor Soler; J L Velasco; Fernando González-Torralba
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-04-19

7.  Peak Inspiratory Flow Rate in COPD: An Analysis of Clinical Trial and Real-World Data.

Authors:  Martin Anderson; Kathryn Collison; M Bradley Drummond; Melanie Hamilton; Renu Jain; Neil Martin; Richard A Mularski; Mike Thomas; Chang-Qing Zhu; Gary T Ferguson
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-04-12

Review 8.  Consistent Pulmonary Drug Delivery with Whole Lung Deposition Using the Aerosphere Inhaler: A Review of the Evidence.

Authors:  Omar S Usmani; Nicolas Roche; Martin Jenkins; Neda Stjepanovic; Peter Mack; Wilfried De Backer
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-01-18

9.  Estimation of inhalation flow profile using audio-based methods to assess inhaler medication adherence.

Authors:  Terence E Taylor; Helena Lacalle Muls; Richard W Costello; Richard B Reilly
Journal:  PLoS One       Date:  2018-01-18       Impact factor: 3.240

10.  Experimental Evaluation of Dry Powder Inhalers during Inhalation and Exhalation Using a Model of the Human Respiratory System (xPULM™).

Authors:  Richard Pasteka; Lara Alina Schöllbauer; Joao Pedro Santos da Costa; Radim Kolar; Mathias Forjan
Journal:  Pharmaceutics       Date:  2022-02-24       Impact factor: 6.321

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.