Literature DB >> 26362051

Trabecular metal spacers as standalone or with pedicle screw augmentation, in posterior lumbar interbody fusion: a prospective, randomized controlled trial.

Erik Van de Kelft1, Johan Van Goethem2,3.   

Abstract

PURPOSE: This prospective randomized comparative trial compared radiological and clinical outcome of Trabecular Metal™ (TM) spacers in PLIF, used as standalone (SA) devices, to TM spacers in PLIF with pedicle screw fixation (PF), in patients with single-level degenerative disc disease (DDD).
METHODS: Patients (n = 80) with chronic low back pain and single-level degenerative disc were randomly assigned to the SA PLIF (n = 40) or PLIF with PF (n = 40). The primary radiological outcome was the evaluation of a long-term (±6 years; range 6.0-7.7 years) stable construct measured by dynamic X-rays. CT scan does not allow judging the bony bridging between vertebrae, because of Tantalum artefacts. The clinical evaluation (6 weeks, 6, 12 and 24 months) consisted of the Oswestry Disability Index (ODI) score, intensity of low back pain (Visual Analogue Scale) and quality of life (Short Form-36).
RESULTS: At 6-year follow-up, X-rays showed a stable construct in 94 % of patients treated by SA TM-500 spacers and in 97 % of those with additional PF. Neither subsidence nor migration was observed in either the SA or the PF group. The average improvement in ODI scores at 24-month clinical follow-up was 14.4 and 13.8 for the SA and PF group, respectively. The VAS score showed an average improvement of 6.4 (SA) and 6.7 (PF), 2 years after implantation. No significant difference between groups was observed at all the evaluation points.
CONCLUSION: In this study, TM spacers were found to provide a solid construct at more than 6-year follow-up after PLIF for DDD both with and without additional pedicle fixation. The clinical, but also radiological results were not significantly different between both cohorts. Future studies focusing on the differences of SA and PF at L4/5 level should be powered to study differences in post-surgery stability at the long term.

Entities:  

Keywords:  Pedicle screw fixation; Posterior lumbar interbody fusion; Radiological outcome; Standalone; Trabecular metal

Mesh:

Year:  2015        PMID: 26362051     DOI: 10.1007/s00586-015-4229-y

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  51 in total

1.  Fibrous tissue ingrowth and attachment to porous tantalum.

Authors:  S A Hacking; J D Bobyn; K Toh; M Tanzer; J J Krygier
Journal:  J Biomed Mater Res       Date:  2000-12-15

Review 2.  When is a spine fused?

Authors:  Christina Goldstein; Brian Drew
Journal:  Injury       Date:  2010-12-13       Impact factor: 2.586

Review 3.  Trabecular metal in hip reconstructive surgery.

Authors:  James B Stiehl
Journal:  Orthopedics       Date:  2005-07       Impact factor: 1.390

4.  Lumbar interbody fusion using the Brantigan I/F cage for posterior lumbar interbody fusion and the variable pedicle screw placement system: two-year results from a Food and Drug Administration investigational device exemption clinical trial.

Authors:  J W Brantigan; A D Steffee; M L Lewis; L M Quinn; J M Persenaire
Journal:  Spine (Phila Pa 1976)       Date:  2000-06-01       Impact factor: 3.468

5.  Revision strategies for salvaging or improving failed cylindrical cages.

Authors:  P C McAfee; B W Cunningham; G A Lee; C M Orbegoso; C J Haggerty; I L Fedder; S L Griffith
Journal:  Spine (Phila Pa 1976)       Date:  1999-10-15       Impact factor: 3.468

Review 6.  Clinical applications of Trabecular Metal.

Authors:  Michael J Christie
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2002-04

7.  The use of a trabecular metal acetabular component and trabecular metal augment for severe acetabular defects.

Authors:  Scott M Sporer; Wayne G Paprosky
Journal:  J Arthroplasty       Date:  2006-09       Impact factor: 4.757

8.  Acetabular revision using a trabecular metal acetabular component for severe acetabular bone loss associated with a pelvic discontinuity.

Authors:  Scott M Sporer; Wayne G Paprosky
Journal:  J Arthroplasty       Date:  2006-09       Impact factor: 4.757

9.  Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.

Authors:  Fabian Bamberg; Alexander Dierks; Konstantin Nikolaou; Maximilian F Reiser; Christoph R Becker; Thorsten R C Johnson
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

10.  Posterior lumbar interbody fusion using local facet joint autograft and pedicle screw fixation.

Authors:  Yukihiro Kai; Masanobu Oyama; Masaaki Morooka
Journal:  Spine (Phila Pa 1976)       Date:  2004-01-01       Impact factor: 3.468

View more
  3 in total

1.  A new 3D printed titanium metal trabecular bone reconstruction system for early osteonecrosis of the femoral head.

Authors:  Ying Zhang; Leilei Zhang; Ruibo Sun; Yudong Jia; Xiantao Chen; Youwen Liu; Hong Oyang; Lizhi Feng
Journal:  Medicine (Baltimore)       Date:  2018-06       Impact factor: 1.889

2.  Vertebral Endplate Cyst Formation in Relation to Properties of Interbody Cages.

Authors:  Manabu Sasaki; Masao Umegaki; Takanori Fukunaga; Yasukazu Hijikata; Yohei Banba; Katsumi Matsumoto; Yasuyoshi Miyao
Journal:  Neurospine       Date:  2021-03-31

Review 3.  The Clinical Application of Porous Tantalum and Its New Development for Bone Tissue Engineering.

Authors:  Gan Huang; Shu-Ting Pan; Jia-Xuan Qiu
Journal:  Materials (Basel)       Date:  2021-05-18       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.