Francesca Buttini1,2, Gaetano Brambilla3, Diego Copelli3, Viviana Sisti3, Anna Giulia Balducci4, Ruggero Bettini1, Irene Pasquali3.
Abstract
BACKGROUND: European and United States Pharmacopoeia compendial procedures for assessing the in vitro emitted dose and aerodynamic size distribution of a dry powder inhaler require that 4.0 L of air at a pressure drop of 4 kPa be drawn through the inhaler. However, the product performance should be investigated using conditions more representative of what is achievable by the patient population. This work compares the delivered dose and the drug deposition profile at different flow rates (30, 40, 60, and 90 L/min) of Foster NEXThaler(®) (beclomethasone dipropionate/formoterol fumarate), Seretide(®) Diskus(®) (fluticasone propionate/salmeterol xinafoate), and Symbicort(®) Turbohaler(®) (budesonide/formoterol fumarate).Entities:
Keywords: Diskus®; NEXThaler®; NGI flow rate; Turbohaler®; aerodynamic assessment; extra-fine particle mass
Mesh:
Substances:
Year: 2015 PMID: 26355743 PMCID: PMC4841907 DOI: 10.1089/jamp.2015.1220
Source DB: PubMed Journal: J Aerosol Med Pulm Drug Deliv ISSN: 1941-2711 Impact factor: 2.849
Experimental Plan of Doses Collected Inside the NGI for Each of the Two Devices Discharged
| 30 | 18-19-20 |
| 38-39-40 | |
| 58-59-60 | |
| 40 | 13-14-15 |
| 33-34-35 | |
| 53-54-55 | |
| 60 | 8-9-10 |
| 28-29-30 | |
| 48-49-50 | |
| 90 | 3-4-5 |
| 23-24-25 | |
| 43-44-45 |

Delivered dose (% of the label claim) performance of the NEXThaler (A), Diskus (B), and Turbohaler (C) at different flow rates; inhalation volume 4L. Error bars represent standard deviations (n=6).

Delivered dose of formoterol fumarate (A) and beclomethasone dipropionate (B) from NEXThaler as a function of inhalation volume (2L and 4L) and different air flow rates (values are mean (SD), n=10).

Fine particle mass expressed as amount of formoterol (FF) (A) and beclometasone dipropionate (BDP) (B) below 2 μm and below 5 μm when NEXThaler was activated at different flow rates, inhalation volume 4L (n=6).

Fine particle mass expressed as amount of salmeterol (as xinafoate) (SX) (A) and fluticasone propionate (FP) (B) below 2 μm and below 5 μm when Diskus was activated at different flow rates, inhalation volume 4L (n=6).

Fine particle mass expressed as amount of formoterol (FF) (A) and budesonide (BD) (B) below 2 μm and below 5 μm when Turbohaler was activated at different flow rates, inhalation volume 4L (n=6).
Delivered Dose, Fine Particle Mass, and Fine Particle Fraction Values of Drug Microparticles Released by Combination Product Foster® NEXThaler®, Seretide® Diskus® and Symbicort® Turbohaler®
| 30 | 84.4 (5.1) | 4.9 (0.2) | 48.2 (1.1) | 2.6 (0.2) | 57.3 (3.7) | 54.5 (4.5) |
| 40 | 83.1 (7.1) | 4.9 (0.3) | 49.2 (1.1) | 2.9 (0.2) | 59.5 (4.6) | 57.9 (5.5) |
| 60 | 88.0 (2.3) | 4.8 (0.2) | 58.4 (4.1) | 3.4 (0.2) | 66.3 (1.0) | 64.9 (2.0) |
| 90 | 85.6 (4.1) | 5.1 (0.2) | 57.9 (3.3) | 3.4 (0.2) | 67.6 (3.0) | 65.6 (2.3) |
| 30 | 45.4 (1.7) | 220.9 (3.8) | 6.4 (0.3) | 36.5 (1.1) | 14.0 (0.4) | 16.5 (0.6) |
| 40 | 48.4 (1.4) | 232.3 (10.6) | 7.7 (1.0) | 44.3 (4.9) | 16.0 (1.7) | 19.1 (2.0) |
| 60 | 48.1 (0.6) | 231.2 (9.1) | 9.3 (0.6) | 52.9 (2.9) | 19.3 (0.9) | 22.9 (1.2) |
| 90 | 50.2 (0.9) | 237.6 (13.1) | 10.5 (1.0) | 59.2 (3.3) | 20.8 (1.7) | 25.0 (1.6) |
| 30 | 125.8 (3.3) | 3.5 (0.1) | 37.0 (2.2) | 1.0 (0.1) | 29.5 (2.5) | 29.8 (1.8) |
| 40 | 138.4 (3.2) | 3.9 (0.2) | 66.5 (3.5) | 1.9 (0.1) | 48.1 (3.4) | 48.4 (2.9) |
| 60 | 164.0 (3.3) | 4.6 (0.1) | 102.1 (2.9) | 2.9 (0.1) | 62.3 (2.7) | 62.6 (1.4) |
| 90 | 158.6 (4.9) | 4.5 (0.1) | 100.2 (4.4) | 2.8 (0.1) | 63.2 (2.6) | 62.7 (3.0) |
BD, budesonide; BPD, beclomethasone dipropionate; DD, delivered dose; FF, formoterol fumarate; FP, fluticasone proprionate; FPF, fine particle fraction; FPM, fine particle mass; SX, salmeterol xinafoate. Values are mean (SD) (n=6).

Extrafine particle fraction (0–2 μm) expressed as percentage of dose delivered of the long-acting beta agonist (LABA) and corticosteroid released from products at 60 L/min, inhalation volume 4L (n=6).

Stage by stage deposition profiles of formoterol fumarate (A) and beclomethasone dipropionate (B) inside the Andersen Cascade Impactor and Next Generation Impactor after NEXThaler aerosolization at 60 L/min, inhalation volume 2 and 4L (values are mean (SD), n=3).
Fine Particle Mass and Fine Particle Fraction Values of Beclomethasone Dipropionate and Formoterol Fumarate Aerosolized by Foster® NEXThaler® inside the Andersen Cascade Impactor and Next Generation Impactor at 2 or 4 L of Air through the Inhaler
| ACI | 2L | 3.1 (0.1) | 65.1 (2.5) |
| 4L | 3.2 (1.3) | 64.1 (1.0) | |
| NGI | 2L | 3.4 (0.2) | 64.9 (2.0) |
| 4L | 3.4 (0.1) | 66.8 (0.8) | |
ACI, Anderson cascade impactor; FPF, fine particle fraction; FPM, fine particle mass; NGI, Next generation impactor.