Lijie He1, Qian Wang2, Daniel Mandler3, Musen Li4, Rabah Boukherroub5, Sabine Szunerits6.
Abstract
The detection of disease markers is considered an important step for early diagnosis of cancer. We design in this work a novel electrochemical sensing platform for the sensitive and selective detection of folic acid protein (FP). The platform is fabricated by electrophoretic deposition (EPD) of reduced graphene oxide (rGO) onto a gold electrode and post-functionalization of rGO with folic acid. Upon FP binding, a significant current decrease can be measured using differential pulse voltammetry (DPV). Using this scheme, a detection limit of 1pM is achieved. Importantly, the method also allows the detection of FP in serum being thus an appealing approach for the sensitive detection of biomarkers in clinical samples.Entities:
Keywords: Biomarker; Differential pulse voltammetry; Folic acid; Folic acid protein; Reduced graphene oxide
Mesh:
Substances:
Year: 2015 PMID: 26342582 DOI: 10.1016/j.bios.2015.08.060
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618