| Literature DB >> 26342055 |
Stephanie J Marin1, Kelly Doyle2, Annie Chang3, Marta Concheiro-Guisan4, Marilyn A Huestis5, Kamisha L Johnson-Davis2.
Abstract
Some amphetamine (AMP) and ecstacy (MDMA) urine immunoassay (IA) kits are prone to false-positive results due to poor specificity of the antibody. We employed two techniques, high-resolution mass spectrometry (HRMS) and an in silico structure search, to identify compounds likely to cause false-positive results. Hundred false-positive IA specimens for AMP and/or MDMA were analyzed by an Agilent 6230 time-of-flight (TOF) mass spectrometer. Separately, SciFinder (Chemical Abstracts) was used as an in silico structure search to generate a library of compounds that are known to cross-react with AMP/MDMA IAs. Chemical formulas and exact masses of 145 structures were then compared against masses identified by TOF. Compounds known to have cross-reactivity with the IAs were identified in the structure-based search. The chemical formulas and exact masses of 145 structures (of 20 chemical formulas) were compared against masses identified by TOF. Urine analysis by HRMS correlates accurate mass with chemical formulae, but provides little information regarding compound structure. Structural data of targeted antigens can be utilized to correlate HRMS-derived chemical formulas with structural analogs.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26342055 PMCID: PMC4731401 DOI: 10.1093/jat/bkv101
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367