B Gory1, A M Spiotta2, S Mangiafico3, A Consoli3, A Biondi4, E Pomero4, M Killer-Oberpfalzer5, W Weber6, R Riva7, P E Labeyrie7, F Turjman7.
Abstract
The PulseRider is a novel endovascular device specifically designed to treat bifurcation intracranial aneurysms with wide necks. In an international series, we report the results of PulseRider stent-assisted coiling of 15 patients (9 women and 6 men; mean age, 62.6 years) with 15 unruptured wide-neck (median dome size, 8 mm; median neck size, 5 mm) bifurcation aneurysms. Failure of PulseRider treatment occurred in 1 case, and 1 intraprocedural thromboembolic complication was observed. There was no mortality or neurologic permanent morbidity at discharge and at 1 month. Immediate angiographic outcome showed 12 complete occlusions and 2 neck remnants. Follow-up at 6 months was available for 3 aneurysms and demonstrated 2 complete aneurysm occlusions and 1 growing neck remnant. In this small series of selected patients, PulseRider stent-assisted coiling of wide-neck bifurcation aneurysms was feasible with low procedural complication rates. Angiographic follow-up will be required to evaluate the efficacy of the PulseRider device.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26338920 PMCID: PMC7960199 DOI: 10.3174/ajnr.A4506
Source DB: PubMed Journal: AJNR Am J Neuroradiol ISSN: 0195-6108 Impact factor: 3.825