Imaging of prompt gamma rays emitted during delivery of clinical proton beams with a Compton camera: feasibility studies for range verification.

Jerimy C Polf1, Stephen Avery, Dennis S Mackin, Sam Beddar.   

Abstract

The purpose of this paper is to evaluate the ability of a prototype Compton camera (CC) to measure prompt gamma rays (PG) emitted during delivery of clinical proton pencil beams for prompt gamma imaging (PGI) as a means of providing in vivo verification of the delivered proton radiotherapy beams. A water phantom was irradiated with clinical 114 MeV and 150 MeV proton pencil beams. Up to 500 cGy of dose was delivered per irradiation using clinical beam currents. The prototype CC was placed 15 cm from the beam central axis and PGs from 0.2 MeV up to 6.5 MeV were measured during irradiation. From the measured data (2D) images of the PG emission were reconstructed. (1D) profiles were extracted from the PG images and compared to measured depth dose curves of the delivered proton pencil beams. The CC was able to measure PG emission during delivery of both 114 MeV and 150 MeV proton beams at clinical beam currents. 2D images of the PG emission were reconstructed for single 150 MeV proton pencil beams as well as for a 5   ×   5 cm mono-energetic layer of 114 MeV pencil beams. Shifts in the Bragg peak (BP) range were detectable on the 2D images. 1D profiles extracted from the PG images show that the distal falloff of the PG emission profile lined up well with the distal BP falloff. Shifts as small as 3 mm in the beam range could be detected from the 1D PG profiles with an accuracy of 1.5 mm or better. However, with the current CC prototype, a dose of 400 cGy was required to acquire adequate PG signal for 2D PG image reconstruction. It was possible to measure PG interactions with our prototype CC during delivery of proton pencil beams at clinical dose rates. Images of the PG emission could be reconstructed and shifts in the BP range were detectable. Therefore PGI with a CC for in vivo range verification during proton treatment delivery is feasible. However, improvements in the prototype CC detection efficiency and reconstruction algorithms are necessary to make it a clinically viable PGI system.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26317610     DOI: 10.1088/0031-9155/60/18/7085

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  16 in total

1.  Feasibility Studies of a New Event Selection Method to Improve Spatial Resolution of Compton Imaging for Medical Applications.

Authors:  E Draeger; S Peterson; D Mackin; H Chen; S Beddar; J C Polf
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2017-05-10

2.  Acoustic-based proton range verification in heterogeneous tissue: simulation studies.

Authors:  Kevin C Jones; Wei Nie; James C H Chu; Julius V Turian; Alireza Kassaee; Chandra M Sehgal; Stephen Avery
Journal:  Phys Med Biol       Date:  2018-01-11       Impact factor: 3.609

Review 3.  In vivo range verification in particle therapy.

Authors:  Katia Parodi; Jerimy C Polf
Journal:  Med Phys       Date:  2018-11       Impact factor: 4.071

4.  The effects of Compton camera data acquisition and readout timing on PG imaging for proton range verification.

Authors:  Jerimy C Polf; Paul Maggi; Rajesh Panthi; Stephen Peterson; Dennis Mackin; Sam Beddar
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-02-05

5.  3D prompt gamma imaging for proton beam range verification.

Authors:  E Draeger; D Mackin; S Peterson; H Chen; S Avery; S Beddar; J C Polf
Journal:  Phys Med Biol       Date:  2018-01-30       Impact factor: 3.609

Review 6.  Latest developments in in-vivo imaging for proton therapy.

Authors:  Katia Parodi
Journal:  Br J Radiol       Date:  2019-12-12       Impact factor: 3.039

7.  Anteriorly Oriented Beam Arrangements with Daily in Vivo Range Verification for Proton Therapy of Prostate Cancer: Rectal Toxicity Rates.

Authors:  Jerimy C Polf; Michael Chuong; Bin Zhang; Minesh Mehta
Journal:  Int J Part Ther       Date:  2016-03-24

8.  Computational model for detector timing effects in Compton-camera based prompt-gamma imaging for proton radiotherapy.

Authors:  Paul Maggi; Steve Peterson; Rajesh Panthi; Dennis Mackin; Hao Yang; Zhong He; Sam Beddar; Jerimy Polf
Journal:  Phys Med Biol       Date:  2020-06-18       Impact factor: 3.609

9.  Secondary Particle Interactions in a Compton Camera Designed for in vivo Range Verification of Proton Therapy.

Authors:  Rajesh Panthi; Paul Maggi; Stephen Peterson; Dennis Mackin; Jerimy Polf; Sam Beddar
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-10-12

10.  First Images of a Three-Layer Compton Telescope Prototype for Treatment Monitoring in Hadron Therapy.

Authors:  Gabriela Llosá; Marco Trovato; John Barrio; Ane Etxebeste; Enrique Muñoz; Carlos Lacasta; Josep F Oliver; Magdalena Rafecas; Carles Solaz; Paola Solevi
Journal:  Front Oncol       Date:  2016-02-02       Impact factor: 6.244

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.