Test-retest reliability of expiratory abdominal compression with a handheld dynamometer in patients with prolonged mechanical ventilation.

Akira Morino1, Masahiro Shida1, Masashi Tanaka1, Kimihiro Sato1, Toshiaki Seko1, Shunsuke Ito1, Shunichi Ogawa1, Naoaki Takahashi2.   

Abstract

[Purpose] The present study aimed to examine the test-retest reliability of expiratory abdominal compression with a handheld dynamometer in patients with prolonged mechanical ventilation.
[Subjects and Methods] We recruited 18 patients with prolonged mechanical ventilation. All patients had impaired consciousness. The mode of the ventilator was synchronized intermittent mandatory ventilation. The abdomen above the navel was vertically compressed using a handheld dynamometer in synchronization with expiration. Expiratory abdominal compression was performed two times. We measured the tidal volume during expiratory abdominal compression. There was an interval of 5 minutes between the first and second measurements. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were performed to examine the test-retest reliability of expiratory abdominal compression with a handheld dynamometer.
[Results] The test-retest reliability of expiratory abdominal compression was excellent (ICC(1, 1): 0.987). Bland-Altman analysis showed that there was no fixed bias and no proportional bias.
[Conclusion] The findings of this study suggest that expiratory abdominal compression with a handheld dynamometer is reliable and useful for patients with respiratory failure and prolonged mechanical ventilation.

Entities:  

Keywords:  Bland-Altman analysis; Expiratory abdominal compression; Reliability

Year:  2015        PMID: 26311946      PMCID: PMC4540841          DOI: 10.1589/jpts.27.2163

Source DB:  PubMed          Journal:  J Phys Ther Sci        ISSN: 0915-5287


INTRODUCTION

Airway clearance techniques, range of motion exercise, and mobilization are performed in pulmonary rehabilitation for mechanically ventilated patients1). Airway clearance prevents ventilator-associated pneumonia. The manual breathing assist technique is widely used for improvement of airway clearance in pulmonary rehabilitation. This technique consists of compression on the chest wall synchronized with physiological movement of the chest wall2). This technique is often performed for patients with respiratory dysfunction (i.e., chronic obstructive lung disease, bronchial asthma, spinal cord injuries, and neuromuscular disease). However, the manual breathing assist technique for patients with prolonged mechanical ventilation (PMV) has a risk of fracture of the ribs because of the osteoporosis caused by long-term bed rest3). Expiratory abdominal compression (EAC) is an airway clearance technique similar to the manual breathing assist technique. The difference between EAC and manual breathing assist is the parts of the body to be compressed. EAC is performed using compression on the abdomen above the navel. Several studies reported application of EAC to patients with respiratory failure such as amyotrophic lateral sclerosis, spinal cord injuries, central nervous system disease, and muscular dystrophy4,5,6,7,8). Another reported EAC with powerful compression, like the Heimlich maneuver, synchronized with coughing for the purpose of assisting coughing9). In contrast, EAC aimed at shifting sputum from the peripheral bronchus to the central bronchus is performed gently and synchronized with expiation. EAC is often used for such patients; nevertheless, little had been reported on EAC. It was considered that EAC increased the cough peak flow, forced vital capacity, and forced expiratory volume in 1 second in those patients. Kaneko et al. revealed that EAC increased the tidal volume (VT) in healthy subjects10). However, the empirical evidence for the manual breathing assist technique is poor, because most studies concerning the manual breathing assist technique did not quantify the pressure of the compression. It is important to establish evaluation methods with good reliability for revealing the effect of the manual breathing assist technique. A handheld dynamometer (HHD) is often used to evaluate peripheral muscle force. An HHD is excellent in terms of portability and conveniently. It is possible that an HHD could be used to examine the pressure of EAC objectivity. In this study, the VT during EAC was measured while quantifying the pressure using an HHD. The aim of this study was to examine the test-retest reliability of EAC in patients with PMV.

SUBJECTS AND METHODS

Subjects

Eighteen inpatients with PMV were enrolled in this study. The patients had required tracheostomy mechanical ventilation for more than a month. A servo s (Fukuda Denshi, Tokyo, Japan) ventilator was used and set in synchronized intermittent mandatory ventilation mode. Individuals who had trouble with the synchronization of the ventilator, rib fracture, presence of a chest tube, hemodynamic instability, and pneumonia within two weeks prior to the measurements of VT during EAC were excluded in this study. A prospective observational study design was utilized. The Institutional Review Board of Heiseikai Hospital approved this study, and the patients or their legally acceptable representative gave written informed consent.

Methods

The characteristics of the patients were collected from their medical records. The setting of the ventilator was confirmed from the display of the ventilator. To quantify the pressure, an HHD (μTas F-1, Anima Corporation, Tokyo, Japan) with sensor dimensions of 56 mm by 56 mm was used. The tester held the pressure sensor of the HHD in a single hand. The abdomen above the navel was vertically compressed through the only whole attachment cover of the sensor. The peak pressure of compression during EAC was 2 kgf in this study. EAC was performed on patients lying in a supine position. The compression was performed in synchronization with expiration and released with inspiration. Prior to measurement, no water in the corrugated tube of the ventilator and no sputum retention in the lung and airway of the patients were confirmed by visual observation and auscultation. EAC was performed two times. There was an interval of 5 minutes between the first and second measurements. The VT during EAC was measured. The display of the ventilator was recorded with a video camera during the measurements. The display of the ventilator during EAC was not checked, so the value of the VT with pressure support was collected after the measurements. EAC was performed by an experienced physiotherapist. Other maneuvers of the breathing technique (i.e., vibration and springing) were not combined with EAC. To examine the test-retest reliability of EAC, the ICC(1, 1) was calculated. Furthermore, Bland-Altman analysis was performed in order to conform whether there was the fixed bias and proportional bias11). Furthermore, the minimal detectable change (MDC) at the 95% confidence level was calculated12). Statistical analyses were performed using R2.8.1, and p < 0.05 was considered to represent statistical significance.

RESULTS

Table 1 shows the characteristics of the 18 patients. The VT values from the first and second measurements were 427.9 mL (SD: 127.4) and 430.4 mL (SD: 128.9). The results of calculation of the ICC and Bland-Altman plot are shown in Table 2. The MDC was 42.0 mL. The Bland-Altman plot revealed that there was no fixed bias and no proportional bias.
Table 1.

Characteristics of the 18 patients

VariablesValues
Age, yr 69.7 ± 15.9
Gender M/F, No. 9/9
Height, cm156.3 ± 11.2
Weight, kg47.6 ± 9.4
BMI, kg/m219.5 ± 3.6
Disease, No. (%)
Hypoxic-ischemic encephalopathy 7 (38.9)
Chronic respiratory failure 4 (22.2)
Cerebrovascular disease3 (16.7)
Spinal cord injury 2 (11.1)
Neurodegenerative disease 2 (11.1)
Glasgow coma scale4.0 (3.0–6.0)
Days of ventilator support, days 830.4 ± 745.3
Setting of ventilator
PC (cmH2O)8.9 ± 2.2
PS (cmH2O)7.1 ± 2.9
PEEP (cmH2O)5.7 ± 1.6
FIO20.23 ± 0.02

Values are expressed as the mean ± SD unless otherwise indicated. †Median (interquartile range). BMI: body mass index; PC: pressure control; PS: pressure support; PEEP: positive end-expiratory pressure; FIO2: fraction of inspired oxygen

Table 2.

The results of the ICC and Bland-Altman analysis

ICCBland-Altman analysisMDC95


ICC (1,1)95% CIFixed biasProportional bias


95% CIBiasRBias
0.987[0.965, 0.995][−13.156, 8.156]No−0.073No42.0

ICC: intraclass correlation coefficient; CI: confidence interval; R: correlation coefficient; MDC95: minimal detectable change at the 95% confidence level

Values are expressed as the mean ± SD unless otherwise indicated. †Median (interquartile range). BMI: body mass index; PC: pressure control; PS: pressure support; PEEP: positive end-expiratory pressure; FIO2: fraction of inspired oxygen ICC: intraclass correlation coefficient; CI: confidence interval; R: correlation coefficient; MDC95: minimal detectable change at the 95% confidence level

DISCUSSION

The present study examined the reliability of EAC in patients with PMV. Based on current knowledge, this is the first report to examine the reliability of EAC. The fact that the reliability of EAC was excellent and there was no fixed bias and no proportional bias indicates that EAC would be useful for the pulmonary rehabilitation in patients with PMV. The ICC(1, 1) was 0.987 in this study. Landis et al. reported that the ICC was almost perfect, above 0.8113). The reliability of EAC in this study was excellent. ICC could evaluate the error including the measured value was an individual difference and a measurement. However, it could not be used to evaluate a systematic error. In contrast, Bland-Altman analysis is a statistical method that can be used to evaluate a systematic error such as a fixed bias or proportional bias. The reliability of EAC was excellent regardless of the magnitude of the VT, because Bland-Altman analysis confirmed that there was no systematic error in this study. The MDC of 42.0 mL meant that the any VT less than that value was considered a random error. These findings might be important in future assessments of the effect of EAC on pulmonary rehabilitation in patients with respiratory failure. The VT in this study was about 430 mL and was less than that reported by Kaneko et al. (1011 mL, SD: 340)10). EAC increased the intra-abdominal pressure, so the diaphragm was pushed up. A previous study reported that the VT during EAC was increased by an increase in the intrathoracic pressure due to pushing up of the diaphragm14). Therefore, the pressure of the compression affected the VT during EAC. Although the pressure of the compression in the study of Kaneko et al. was not quantified10), the pressure in the present study might have been lower than theirs. The subjects in their study were healthy and younger than those in the present study (20.6 years [SD: 0.5] vs. 69.7 years [SD: 15.9]). Because VT decreases with age, the VT during EAC in the present study was less than that in their study. This study has a limitation. EAC in this study was performed through an HHD. The manual breathing technique is generally performed by placing the whole palm against the patient’s body. Therefore, it is necessary to be careful when attempting to generalize. In conclusion, it was found that EAC had excellent reliability, no fixed bias, and no proportional bias. The present findings suggest that EAC with an HHD is useful for evaluation of patients who require EAC in pulmonary rehabilitation objectively.
  10 in total

1.  A dynamic analysis of chest wall motions with MRI in healthy young subjects.

Authors:  T Kondo; I Kobayashi; Y Taguchi; Y Ohta; N Yanagimachi
Journal:  Respirology       Date:  2000-03       Impact factor: 6.424

2.  AARC (American Association for Respiratory Care) clinical practice guideline. Postural drainage therapy.

Authors: 
Journal:  Respir Care       Date:  1991-12       Impact factor: 2.258

3.  Death from food-choking prevented by a new life-saving maneuver.

Authors:  H J Heimlich
Journal:  Heart Lung       Date:  1976 Sep-Oct       Impact factor: 2.210

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The measurement of observer agreement for categorical data.

Authors:  J R Landis; G G Koch
Journal:  Biometrics       Date:  1977-03       Impact factor: 2.571

6.  How respiratory muscle strength correlates with cough capacity in patients with respiratory muscle weakness.

Authors:  Jung Hyun Park; Seong-Woong Kang; Sang Chul Lee; Won Ah Choi; Dong Hyun Kim
Journal:  Yonsei Med J       Date:  2010-05       Impact factor: 2.759

7.  Assisted cough and pulmonary compliance in patients with Duchenne muscular dystrophy.

Authors:  Seong Woong Kang; Yeoun Seung Kang; Jae Ho Moon; Tae Won Yoo
Journal:  Yonsei Med J       Date:  2005-04-30       Impact factor: 2.759

Review 8.  Respiratory treatment of amyotrophic lateral sclerosis.

Authors:  Joshua O Benditt; Louis Boitano
Journal:  Phys Med Rehabil Clin N Am       Date:  2008-08       Impact factor: 1.784

9.  Intravenous pamidronate prevents femoral bone loss and renal stone formation during 90-day bed rest.

Authors:  Yukiko Watanabe; Hiroshi Ohshima; Koh Mizuno; Chiharu Sekiguchi; Masao Fukunaga; Kenjiro Kohri; Jörn Rittweger; Dieter Felsenberg; Toshio Matsumoto; Toshitaka Nakamura
Journal:  J Bone Miner Res       Date:  2004-08-23       Impact factor: 6.741

10.  Respiratory muscle strength and cough capacity in patients with Duchenne muscular dystrophy.

Authors:  Seong-Woong Kang; Yeoun-Seung Kang; Hong-Seok Sohn; Jung-Hyun Park; Jae-Ho Moon
Journal:  Yonsei Med J       Date:  2006-04-30       Impact factor: 2.759

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.