Akira Morino1, Masahiro Shida1, Masashi Tanaka1, Kimihiro Sato1, Toshiaki Seko1, Shunsuke Ito1, Shunichi Ogawa1, Naoaki Takahashi2.
Abstract
[Purpose] The present study aimed to examine the test-retest reliability of expiratory abdominal compression with a handheld dynamometer in patients with prolonged mechanical ventilation.Entities:
Keywords: Bland-Altman analysis; Expiratory abdominal compression; Reliability
Year: 2015 PMID: 26311946 PMCID: PMC4540841 DOI: 10.1589/jpts.27.2163
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Characteristics of the 18 patients
| Variables | Values |
| Age, yr | 69.7 ± 15.9 |
| Gender M/F, No. | 9/9 |
| Height, cm | 156.3 ± 11.2 |
| Weight, kg | 47.6 ± 9.4 |
| BMI, kg/m2 | 19.5 ± 3.6 |
| Disease, No. (%) | |
| Hypoxic-ischemic encephalopathy | 7 (38.9) |
| Chronic respiratory failure | 4 (22.2) |
| Cerebrovascular disease | 3 (16.7) |
| Spinal cord injury | 2 (11.1) |
| Neurodegenerative disease | 2 (11.1) |
| Glasgow coma scale† | 4.0 (3.0–6.0) |
| Days of ventilator support, days | 830.4 ± 745.3 |
| Setting of ventilator | |
| PC (cmH2O) | 8.9 ± 2.2 |
| PS (cmH2O) | 7.1 ± 2.9 |
| PEEP (cmH2O) | 5.7 ± 1.6 |
| FIO2 | 0.23 ± 0.02 |
Values are expressed as the mean ± SD unless otherwise indicated. †Median (interquartile range). BMI: body mass index; PC: pressure control; PS: pressure support; PEEP: positive end-expiratory pressure; FIO2: fraction of inspired oxygen
The results of the ICC and Bland-Altman analysis
| ICC | Bland-Altman analysis | MDC95 | ||||
|---|---|---|---|---|---|---|
| ICC (1,1) | 95% CI | Fixed bias | Proportional bias | |||
| 95% CI | Bias | R | Bias | |||
| 0.987 | [0.965, 0.995] | [−13.156, 8.156] | No | −0.073 | No | 42.0 |
ICC: intraclass correlation coefficient; CI: confidence interval; R: correlation coefficient; MDC95: minimal detectable change at the 95% confidence level