| Literature DB >> 26289113 |
Aleksej Krunić1, Jimmy Orjala1.
Abstract
We have investigated and compared a number of sample conditions on different NMR platforms in the search of maximum SNR and optimal experiment time efficiency for structure elucidation and quantitation of natural products. Using restricted volume 3 mm Shigemi microcell assembly in conjunction with a 900 MHz NMR spectrometer equipped with a 5 mm carbon-sensitive inverse cryoprobe, it was possible to achieve a substantial increase in SNR (46-fold) as compared with a conventional room temperature 400 MHz instrument. Switching from standard 5 mm NMR tube to 3 mm Shigemi microcell assembly typically improved SNR by threefold on either 600 or 900 MHz cryoplatform. A quantitation method that relies on a calibrated residual protonated NMR solvent signal as internal standard was developed using the same hardware setup and restricted sample volume tubes. Linearity of the method spans over 3 orders of magnitude, from low microgram to milligram quantities. We successfully applied this method to quantify a low micrgram sample of paclitaxel, verified by a UV/VIS quantitation measurement.Entities:
Keywords: 13C; 15 N; 1H; 900 MHz; NMR; Shigemi microcell assembly; cryoprobe; internal standard; qNMR
Mesh:
Substances:
Year: 2015 PMID: 26289113 PMCID: PMC4752180 DOI: 10.1002/mrc.4304
Source DB: PubMed Journal: Magn Reson Chem ISSN: 0749-1581 Impact factor: 2.447