Literature DB >> 26283153

Factors Related to Pump Thrombosis With the Heartmate II Left Ventricular Assist Device.

Charles T Klodell1, H Todd Massey2, Robert M Adamson3, David A Dean4, Douglas A Horstmanshof5, John M Ransom6, Christopher T Salerno7, Jennifer A Cowger7, Juan M Aranda8, Leway Chen2, James W Long5, Walter Dembitsky3.   

Abstract

BACKGROUND: Recent reports suggested that HeartMate II (HMII) thrombosis rates may be higher in implants after 2011. We characterize events at HMII centers (>100 HMII implants) whose device thrombosis rates are equivalent or lower than reported by INTERMACS.
METHODS: Seven centers pooled implants from 2011 through June 2013 to examine pump thrombus and identify characteristics and clinical strategies that potentially mitigate the risk. A total of 666 patients (age 59 ± 13 years; 81% male) were studied (support duration: 13.7 ± 8.3 months, cumulative: 759 patient years). Median target INR was 2.25 (range 2.0 to 2.5), and median pump speed was 9200 rpm (range 8600 to 9600). Pump thrombus was suspected with clinical evidence (e.g., hemolysis, positive ramp test) requiring intervention (e.g., anticoagulation therapy, pump exchange) or patient death.
RESULTS: Suspected pump thrombus occurred in 24/666 (3.6%) patients within three months of implant. At six months, 38/666 (5.7%) had suspected pump thrombus including 24 (3.6%) resulting in pump exchange or death. Stroke (hemorrhagic: 0.049, and ischemic: 0.048 events/patient year) and survival (six months: 88 ± 1%; 1 year: 81 ± 2%) were consistent with national averages. Suspected pump thrombus patients were younger (55 ± 13 vs. 59 ± 13, p = 0.046) and had more females (31.6% vs. 18.3%, p = 0.054). There was no difference in indication, etiology of heart failure, or body size.
CONCLUSIONS: This analysis demonstrates low HMII thrombus events. Minimization of risk factors by uniform implant techniques and consistent post-op management may reduce device thrombosis. A larger scale multicenter evaluation may better elucidate the difference in thrombus events between centers.
© 2015 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2015        PMID: 26283153     DOI: 10.1111/jocs.12607

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  3 in total

Review 1.  CSI position statement on management of heart failure in India.

Authors:  Santanu Guha; S Harikrishnan; Saumitra Ray; Rishi Sethi; S Ramakrishnan; Suvro Banerjee; V K Bahl; K C Goswami; Amal Kumar Banerjee; S Shanmugasundaram; P G Kerkar; Sandeep Seth; Rakesh Yadav; Aditya Kapoor; Ajaykumar U Mahajan; P P Mohanan; Sundeep Mishra; P K Deb; C Narasimhan; A K Pancholia; Ajay Sinha; Akshyaya Pradhan; R Alagesan; Ambuj Roy; Amit Vora; Anita Saxena; Arup Dasbiswas; B C Srinivas; B P Chattopadhyay; B P Singh; J Balachandar; K R Balakrishnan; Brian Pinto; C N Manjunath; Charan P Lanjewar; Dharmendra Jain; Dipak Sarma; G Justin Paul; Geevar A Zachariah; H K Chopra; I B Vijayalakshmi; J A Tharakan; J J Dalal; J P S Sawhney; Jayanta Saha; Johann Christopher; K K Talwar; K Sarat Chandra; K Venugopal; Kajal Ganguly; M S Hiremath; Milind Hot; Mrinal Kanti Das; Neil Bardolui; Niteen V Deshpande; O P Yadava; Prashant Bhardwaj; Pravesh Vishwakarma; Rajeeve Kumar Rajput; Rakesh Gupta; S Somasundaram; S N Routray; S S Iyengar; G Sanjay; Satyendra Tewari; Sengottuvelu G; Soumitra Kumar; Soura Mookerjee; Tiny Nair; Trinath Mishra; U C Samal; U Kaul; V K Chopra; V S Narain; Vimal Raj; Yash Lokhandwala
Journal:  Indian Heart J       Date:  2018-06-08

Review 2.  Stroke and Intracranial Hemorrhage in HeartMate II and HeartWare Left Ventricular Assist Devices: A Systematic Review.

Authors:  Sung M Cho; Nader Moazami; Jennifer A Frontera
Journal:  Neurocrit Care       Date:  2017-08       Impact factor: 3.210

3.  Successful continuous-flow left ventricular assist device implantation with adjuvant tricuspid valve repair for advanced heart failure.

Authors:  Chih-Hsien Lee; Jeng Wei
Journal:  Cardiovasc J Afr       Date:  2016-04-12       Impact factor: 1.167

  3 in total