Evaluation of Ultrasound-Based Sensor to Monitor Respiratory and Nonrespiratory Movement and Timing in Infants.

Gregory P Heldt, Raymond J Ward.   

Abstract

GOAL: To describe and validate a noncontacting sensor that used reflected ultrasound to separately monitor respiratory, nonrespiratory, and caretaker movements of infants.
METHODS: An in-phase and quadrature (I & Q) detection scheme provided adequate bandwidth, in conjunction with postdetection filtering, to separate the three types of movement. The respiratory output was validated by comparing it to the electrical activity of the diaphragm (Edi) obtained from an infant ventilator in 11 infants. The nonrespiratory movement output was compared to movement detected by miniature accelerometers attached to the wrists, ankles, and heads of seven additional infants. Caretaker movement was compared to visual observations annotated in the recordings.
RESULTS: The respiratory rate determined by the sensor was equivalent to that from the Edi signal. The sensor could detect the onset of inspiration significantly earlier than the Edi signal (23+/-69 ms). Nonrespiratory movement was identified with an agreement of 0.9 with the accelerometers. It potentially interfered with the respiratory output an average of 4.7+/-4.5% and 14.9+/15% of the time in infants not requiring or on ventilatory support, respectively. Caretaker movements were identified with 98% sensitivity and specificity. The sensor outputs were independent of body coverings or position.
CONCLUSION: This single, noncontacting sensor can independently quantify these three types of movement. SIGNIFICANCE: It is feasible to use the sensor as trigger for synchronizing mechanical ventilators to spontaneous breathing, to quantify overall movement, to determine sleep state, to detect seizures, and to document the amount and effects of caretaker activity in infants.

Entities:  

Mesh:

Year:  2015        PMID: 26276983      PMCID: PMC4873773          DOI: 10.1109/TBME.2015.2466633

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  14 in total

1.  Diaphragm electrical activity during expiration in mechanically ventilated infants.

Authors:  Guillaume Emeriaud; Jennifer Beck; Marisa Tucci; Jacques Lacroix; Christer Sinderby
Journal:  Pediatr Res       Date:  2006-05       Impact factor: 3.756

2.  A sensor for monitoring pulse rate, respiration rhythm, and body movement in bed.

Authors:  Yusuke Yamana; Sosuke Tsukamoto; Koji Mukai; Hiromichi Maki; Hidekuni Ogawa; Yoshiharu Yonezawa
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2011

3.  Early prediction of cerebral palsy by computer-based video analysis of general movements: a feasibility study.

Authors:  Lars Adde; Jorunn L Helbostad; Alexander R Jensenius; Gunnar Taraldsen; Kristine H Grunewaldt; Ragnhild Støen
Journal:  Dev Med Child Neurol       Date:  2010-02-24       Impact factor: 5.449

4.  Characterization of neural breathing pattern in spontaneously breathing preterm infants.

Authors:  Jennifer Beck; Maureen Reilly; Giacomo Grasselli; Haibo Qui; Arthur S Slutsky; Michael S Dunn; Christer A Sinderby
Journal:  Pediatr Res       Date:  2011-12       Impact factor: 3.756

5.  A remote monitor of bed patient cardiac vibration, respiration and movement.

Authors:  Koji Mukai; Yoshiharu Yonezawa; Hidekuni Ogawa; Hiromichi Maki; W Morton Caldwell
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

6.  Recombinant human tumor necrosis factor administered as a 24-hour intravenous infusion. A phase I and pharmacologic study.

Authors:  D R Spriggs; M L Sherman; H Michie; K A Arthur; K Imamura; D Wilmore; E Frei; D W Kufe
Journal:  J Natl Cancer Inst       Date:  1988-09-07       Impact factor: 13.506

7.  Response time and reliability of three neonatal patient-triggered ventilators.

Authors:  G Bernstein; J P Cleary; G P Heldt; J F Rosas; L D Schellenberg; F L Mannino
Journal:  Am Rev Respir Dis       Date:  1993-08

8.  Airway leak size in neonates and autocycling of three flow-triggered ventilators.

Authors:  G Bernstein; E Knodel; G P Heldt
Journal:  Crit Care Med       Date:  1995-10       Impact factor: 7.598

9.  Increased and more consistent tidal volumes during synchronized intermittent mandatory ventilation in newborn infants.

Authors:  G Bernstein; G P Heldt; F L Mannino
Journal:  Am J Respir Crit Care Med       Date:  1994-11       Impact factor: 21.405

10.  An ultrasonic contactless sensor for breathing monitoring.

Authors:  Philippe Arlotto; Michel Grimaldi; Roomila Naeck; Jean-Marc Ginoux
Journal:  Sensors (Basel)       Date:  2014-08-20       Impact factor: 3.576

View more
  7 in total

1.  Contact and Remote Breathing Rate Monitoring Techniques: A Review.

Authors:  Mohamed Ali; Ali Elsayed; Arnaldo Mendez; Yvon Savaria; Mohamad Sawan
Journal:  IEEE Sens J       Date:  2021-04-12       Impact factor: 4.325

2.  A Low-Power Thermal-Based Sensor System for Low Air Flow Detection.

Authors:  Akm Arifuzzman; Mohammad Rafiqul Haider; David B Allison
Journal:  Analog Integr Circuits Signal Process       Date:  2016-09-08       Impact factor: 1.337

3.  Feasibility and physiological effects of noninvasive neurally adjusted ventilatory assist in preterm infants.

Authors:  Christopher K Gibu; Phillip Y Cheng; Raymond J Ward; Benjamin Castro; Gregory P Heldt
Journal:  Pediatr Res       Date:  2017-07-12       Impact factor: 3.756

4.  The Ultrasonic Directional Tidal Breathing Pattern Sensor: Equitable Design Realization Based on Phase Information.

Authors:  Arijit Sinharay; Raj Rakshit; Anwesha Khasnobish; Tapas Chakravarty; Deb Ghosh; Arpan Pal
Journal:  Sensors (Basel)       Date:  2017-08-11       Impact factor: 3.576

5.  Ultrasound Sensors for Diaphragm Motion Tracking: An Application in Non-Invasive Respiratory Monitoring.

Authors:  Amirhossein Shahshahani; Carl Laverdiere; Sharmistha Bhadra; Zeljko Zilic
Journal:  Sensors (Basel)       Date:  2018-08-09       Impact factor: 3.576

6.  Smart Vest for Respiratory Rate Monitoring of COPD Patients Based on Non-Contact Capacitive Sensing.

Authors:  David Naranjo-Hernández; Alejandro Talaminos-Barroso; Javier Reina-Tosina; Laura M Roa; Gerardo Barbarov-Rostan; Pilar Cejudo-Ramos; Eduardo Márquez-Martín; Francisco Ortega-Ruiz
Journal:  Sensors (Basel)       Date:  2018-07-03       Impact factor: 3.576

Review 7.  Sensing and Artificial Intelligent Maternal-Infant Health Care Systems: A Review.

Authors:  Saima Gulzar Ahmad; Tassawar Iqbal; Anam Javaid; Ehsan Ullah Munir; Nasira Kirn; Sana Ullah Jan; Naeem Ramzan
Journal:  Sensors (Basel)       Date:  2022-06-09       Impact factor: 3.847

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.