| Literature DB >> 26275445 |
Leman Yel1,2, Christopher J Rabbat3, Charlotte Cunningham-Rundles4, Jordan S Orange5, Troy R Torgerson6, James W Verbsky7, Yeong Wang8, Maoyong Fu9, Terry S Robins9, Marc S Edwards9, Jesper Nymann-Andersen9.
Abstract
PURPOSE: To develop an assay to quantify serum immunoglobulin (IgG, IgM, IgA) levels using dried blood spots (DBS) obtained on collection cards to be used as a tool for targeted screening for hypogammaglobulinemia.Entities:
Keywords: DBS; ELISA; Ig-DBS assay; dried blood spots; immunoglobulin; nephelometry; primary immunodeficiency; validation
Mesh:
Substances:
Year: 2015 PMID: 26275445 PMCID: PMC4572045 DOI: 10.1007/s10875-015-0184-y
Source DB: PubMed Journal: J Clin Immunol ISSN: 0271-9142 Impact factor: 8.317
Fig. 1Blood sample collection using a DBS card. Photo, IDBS spot saver card, ID biological systems, PerkinElmer, Greenville, SC, USA
Intra- and inter-assay precision and accuracy for IgG, IgM, and IgA
| QC samples | Intra-assay precision (CV%) | Intra-assay accuracy (% bias) | QC samples | Inter-assay precision (CV%) | Inter-assay accuracy (% bias) |
|---|---|---|---|---|---|
| IgG | IgG | ||||
| LLOQ (3.2 ng/mL) | 12.3 | −17.1 | LLOQ (3.9 ng/mL) | 19.1 | −16.5 |
| Low QC (7.2 ng/mL) | 10.7 | −16.7 | LQC (8.6 ng/mL) | 18.6 | −7.7 |
| Middle QC (14.2 ng/mL) | 9.5 | −8.6 | MQC (15.6 ng/mL) | 19.9 | −16.9 |
| High QC (33.5 ng/mL) | 6.7 | −2.3 | HQC (34.3 ng/mL) | 15.2 | −8.7 |
| ULOQ (57.7 ng/mL) | 17.1 | −9.2 | ULOQ (63.5 ng/mL) | 16.5 | −15.3 |
| IgM | IgM | ||||
| LLOQ (43.4 ng/mL) | 7.1 | 11.8 | LLOQ (49.3 ng/mL) | 14.0 | 5.1 |
| Low QC (91.8 ng/mL) | 4.4 | −8.7 | LQC (100.5 ng/mL) | 7.5 | 7.2 |
| Middle QC (190.8 ng/mL) | 4.2 | −5.3 | MQC (201.6 ng/mL) | 5.9 | 7.5 |
| High QC (371.2 ng/mL) | 3.6 | −5.4 | HQC (392.3 ng/mL) | 6.6 | 4.6 |
| ULOQ (791.8 ng/mL) | 4.3 | −1.3 | ULOQ (802.5 ng/mL) | 5.9 | 7.0 |
| IgA | IgA | ||||
| LLOQ (4.0 ng/mL) | 3.7 | −9.3 | LLOQ (4.5 ng/mL) | 11.1 | −4.9 |
| Low QC (8.5 ng/mL) | 2.7 | −5.9 | Low QC (9.0 ng/mL) | 7.0 | −3.8 |
| Middle QC (17.0 ng/mL) | 2.6 | −3.9 | Middle QC (17.7 ng/mL) | 5.8 | −5.5 |
| High QC (35.3 ng/mL) | 2.8 | −2.7 | High QC (36.3 ng/mL) | 4.9 | −3.3 |
| ULOQ (71.9 ng/mL) | 3.5 | −1.5 | ULOQ (73.0 ng/mL) | 4.4 | −2.6 |
CV coefficient of variation, LLOQ lower limit of quantitation, QC quality control, LQC low range of the dynamic range QC, MQC middle of the dynamic range QC, HQC high range of the dynamic range QC, QC quality control, ULOQ upper limit of quantitation
Specificity and selectivity of the Ig-DBS assay
| IgG | ||||||||
| Specificity | Selectivity | |||||||
| IgG calibrator captured by anti-IgA | IgG calibrator captured by anti-IgM | |||||||
| Matrix | IgG added at 3.1 ng/mL | 37.5 ng/mL | 18.75 ng/mL | 9.38 ng/mL | 37.5 ng/mL | 18.75 ng/mL | 9.38 ng/mL | |
| Mean | <LLOQ | 1.9 | 2.6a | 2.8a | 2.4a | 3.2a | 2.5a | 2.2a |
| SD | – | 0.2 | – | – | – | – | – | – |
| % CV | – | 8.9 | – | – | – | – | – | – |
| % accuracy | – | 62.3 | – | – | – | – | – | – |
| a<LLOQ of 4.7 ng/mL | ||||||||
| IgM | ||||||||
| Specificitya | Selectivity | |||||||
| IgM calibrator captured by anti-IgA | IgM calibrator captured by anti-IgG | |||||||
| Matrix | IgM added at 15.6 ng/mL | 375.0 ng/mL | 187.5 ng/mL | 93.8 ng/mL | 375.0 ng/mL | 187.5 ng/mL | 93.8 ng/mL | |
| Mean | <LLOQ | <LLOQ | <LLOQ | <LLOQ | <LLOQ | <LLOQ | <LLOQ | <LLOQ |
| bLLOQ = 46.9 ng/mL | ||||||||
| IgA | ||||||||
| Specificity | Selectivity | |||||||
| IgA calibrator captured by anti-IgG | IgA calibrator captured by anti-IgM | |||||||
| Matrix | IgA added at 3.1 ng/mL | 37.5 ng/mL | 18.75 ng/mL | 9.38 ng/mL | 37.5 ng/mL | 18.75 ng/mL | 9.38 ng/mL | |
| Mean | <LLOQ | 2.8 | 1.1a | 1.1a | 1.1a | 4.2a | 2.5a | 1.8a |
| SD | – | 0.1 | – | – | – | – | – | – |
| % CV | – | 3.7 | – | – | – | – | – | – |
| % accuracy | – | 88.6 | – | – | – | – | – | – |
| a<LLOQ of 4.7 ng/mL | ||||||||
CV coefficient of variation, LLOQ lower limit of quantitation, SD standard deviation
aThe assay was unable to quantify analyte added to matrix at 15.6 ng/mL as well as in matrix without added analyte, thereby demonstrating specificity
bResults in the assay for all concentrations that are below the LLOQ of 46.9 ng/mL for the assay indicate no selectivity issue
Fig. 2Correlations between standard nephelometry assay and Ig-DBS for IgG (a), IgM (b), and IgA (c) (scatterplot and regression line)
Days of stability of IgG, IgM, and IgA in DBSa
| Analyte | Room temperature | 2 to 8 °C | 36 to 38 °C | −20 to −40 °C |
|---|---|---|---|---|
| IgG | 14 | 14 | 4 | 14 |
| IgM | 14 | 14 | 4 | 14 |
| IgA | 14 | 14 | 3 | 10 |
aStorage in polyethylene bag
Maximum number of days evaluated at room temperature = 14; at 2°–8 °C = 14; at 36°–38 °C = 4; at −20° to −40 °C = 14