Vector Analysis, Rotational Stability, and Visual Outcome After Implantation of a New Aspheric Toric IOL.

Alexander Bachernegg, Theresa Rückl, Clemens Strohmaier, Gerlinde Jell, Günther Grabner, Alois K Dexl.   

Abstract

PURPOSE: To evaluate vector analysis, rotational stability, and refractive and visual outcome of a new toric intraocular lens (IOL) for correction of preexisting corneal astigmatism during routine cataract surgery.
METHODS: In this prospective, interventional case series, 30 toric, aspheric Bi-Flex T toric IOLs (Medicontur Medical Engineering Ltd., Inc., Zsámbék, Hungary) were implanted in 20 consecutive patients with topographic corneal astigmatism between 1.50 and 4.00 diopters (D) and evaluated within the first year after implantation. Appropriate IOL-toric alignment was facilitated by combined imaging/eye tracking technology. Postoperative evaluation included refraction and uncorrected and corrected distance visual acuities (UDVA, CDVA). For each visit, photodocumentation in retroillumination was performed to evaluate toric alignment and potential toric IOL rotation. Vector analysis of refractive astigmatism was performed using the Alpins method.
RESULTS: At 12 months postoperatively, a reduction of the refractive astigmatism from 1.93±0.90 D (range: 0.50 to 4.00 D) to 0.28±0.61 D (range: 0.00 to 1.50 D) could be found, with patients achieving a mean UDVA of 0.06±0.16 logMAR (range: -0.18 to 0.40 logMAR; Snellen 20/20). Intraoperative to 12-month postoperative comparison of IOL axis alignment showed low levels of rotation (0.2°±2.41°; range: +4° to -5°). Vector analysis showed target induced astigmatism of 0.60 D @180°, surgically induced astigmatism of 0.80 D @177°, correction index of 1.02±0.25, and a difference vector of 0.30 D @82°.
CONCLUSIONS: Implantation of the new Bi-Flex T IOL was a safe, stable, and effective method to correct preexisting regular corneal astigmatism during cataract surgery. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Year:  2015        PMID: 26274517     DOI: 10.3928/1081597X-20150727-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  8 in total

1.  Evaluation of Astigmatism-Correcting Efficiency and Rotational Stability after Cataract Surgery with a Double-Loop Haptic Toric Intraocular Lens: A 1-Year Follow-Up.

Authors:  Ladislav Viktor Nováček; Marie Němcová; Kateřina Tyx; Kristýna Lahodová; Leoš Rejmont; Pavel Rozsíval; Pavel Studený
Journal:  Biomed Hub       Date:  2021-02-17

2.  Evaluation of biometry and corneal astigmatism in cataract surgery patients from Central China.

Authors:  Ji-Guo Yu; Jie Zhong; Zhong-Ming Mei; Fang Zhao; Na Tao; Yi Xiang
Journal:  BMC Ophthalmol       Date:  2017-04-26       Impact factor: 2.209

3.  Evaluation of Decentration, Tilt and Angular Orientation of Toric Intraocular Lens.

Authors:  Martin Fus; Sarka Pitrova
Journal:  Clin Ophthalmol       Date:  2021-12-19

4.  Simultaneous Presbyopia and Astigmatism Correction with a Novel Trifocal Toric Intraocular Lens-A One-Year Follow-Up.

Authors:  Ladislav Viktor Nováček; Marie Němcová; Kristýna Sičová; Kateřina Tyx; Pavel Rozsíval; Jan Němčanský; Pavel Studený
Journal:  J Clin Med       Date:  2022-07-19       Impact factor: 4.964

5.  Clinical Outcomes after Uncomplicated Cataract Surgery with Implantation of the Tecnis Toric Intraocular Lens.

Authors:  Wojciech Lubiński; Beata Kaźmierczak; Jolanta Gronkowska-Serafin; Karolina Podborączyńska-Jodko
Journal:  J Ophthalmol       Date:  2016-02-28       Impact factor: 1.909

6.  Comparison of Two Toric IOL Calculation Methods.

Authors:  C Kern; K Kortüm; M Müller; A Kampik; S Priglinger; W J Mayer
Journal:  J Ophthalmol       Date:  2018-01-10       Impact factor: 1.909

7.  Comparison of two techniques for toric intraocular lens implantation: hydroimplantation versus ophthalmic viscosurgical devices.

Authors:  Yueqin Chen; Qian Cao; Chunyan Xue; Zhenping Huang
Journal:  BMC Ophthalmol       Date:  2018-04-24       Impact factor: 2.209

8.  Comparison of the clinical efficacy of AcrySof® IQ and TECNIS® toric intraocular lenses: A real-world study.

Authors:  Juan-Juan Yang; Ya-Zhou Qin; Li Qin; Jing-Ming Li
Journal:  Exp Ther Med       Date:  2020-08-28       Impact factor: 2.447

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.