Literature DB >> 26236068

SmartCAM: An adaptive clinical SPECT camera.

Chris Dumas1, Adam Bernstein1, Alonzo Espinoza1, Donovan Morgan1, Kevin Lewis1, Matt Nipper1, Harrison H Barrett2, Matthew A Kupinski3, Lars R Furenlid2.   

Abstract

An adaptive pinhole aperture that fits a GE MaxiCam Single-Photon-Emission Computed Tomography (SPECT) system has been designed, built, and is undergoing testing. The purpose of an adaptive aperture is to allow the imaging system to make adjustments to the aperture while imaging data are being acquired. Our adaptive pinhole aperture can alter several imaging parameters, including field of view, resolution, sensitivity, and magnification. The dynamic nature of such an aperture allows for imaging of specific regions of interest based on initial measurements of the patient. Ideally, this mode of data collection will improve the understanding of a patient's condition, and will facilitate better diagnosis and treatment. The aperture was constructed using aluminum and a low melting point, high-stopping-power metal alloy called Cerrobend. The aperture utilizes a rotating disk for the selection of a pinhole configuration; as the aluminum disk rotates, different pinholes move into view of the camera face and allow the passage of gamma rays through that particular pinhole. By controlling the angular position of the disk, the optical characteristics of the aperture can be modified, allowing the system to acquire data from controlled regions of interest. First testing was performed with a small radioactive source to prove the functionality of the aperture.

Entities:  

Year:  2013        PMID: 26236068      PMCID: PMC4520707          DOI: 10.1117/12.2030090

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  3 in total

1.  A Task-Based Approach to Adaptive and Multimodality Imaging: Computation techniques are proposed for figures-of-merit to establish feasibility and optimize use of multiple imaging systems for disease diagnosis and treatment-monitoring.

Authors:  Eric Clarkson; Matthew A Kupinski; Harrison H Barrett; Lars Furenlid
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2008-03       Impact factor: 10.961

2.  Adaptive SPECT.

Authors:  Harrison H Barrett; Lars R Furenlid; Melanie Freed; Jacob Y Hesterman; Matthew A Kupinski; Eric Clarkson; Meredith K Whitaker
Journal:  IEEE Trans Med Imaging       Date:  2008-06       Impact factor: 10.048

3.  The Effects of Incorrect Modeling on Noise and Resolution Properties of ML-EM Images.

Authors:  D W Wilson; H H Barrett
Journal:  IEEE Trans Nucl Sci       Date:  2002-06       Impact factor: 1.679