| Literature DB >> 26229687 |
Dorothy S K Ng1, Preeti Gupta1, Yih Chung Tham1, Chye Fong Peck2, Tien Yin Wong3, Mohammad Kamran Ikram4, Carol Y Cheung3.
Abstract
Purpose. To assess the repeatability of spectral-domain optical coherence tomography to measure macular and perimacular ganglion cell complex thicknesses and compare retinal ganglion cell parameters between algorithms. Methods. Ninety-two nonglaucomatous eyes from 92 participants underwent macular and perimacular ganglion cell complex thickness measurement using OCT-HS100 Glaucoma 3D algorithm and these measurements were repeated for 34 subjects. All subjects also had macular ganglion cell-inner plexiform layer thickness measured by Cirrus HD-OCT Ganglion Cell Analysis algorithm. Intraclass correlation coefficient and Pearson's correlation analyses were performed. Results. Subfields of both macular and perimacular ganglion cell complex thicknesses had high intraclass correlation coefficient values between 0.979 (95% confidence interval [CI]: 0.958-0.989) and 0.981 (95% CI: 0.963, 0.991) and between 0.70 (95% CI: 0.481-0.838) and 0.987 (95% CI: 0.956-0.989), respectively. The overall average ganglion cell complex and macular average ganglion cell-inner plexiform layer thicknesses were strongly correlated (r = 0.83, P < 0.001). Conclusions. The assessment of macular and perimacular retinal ganglion cell parameters by OCT-HS100 Glaucoma 3D algorithm is highly repeatable, and strongly correlates to retinal ganglion cell parameters assessed by Ganglion Cell Analysis algorithm. A comprehensive evaluation of retinal ganglion cells may be possible with OCT-HS100.Entities:
Year: 2015 PMID: 26229687 PMCID: PMC4502302 DOI: 10.1155/2015/605940
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Measurement grids of OCT-HS100 Glaucoma 3D algorithm. The grids are centered on the fovea and rotated in alignment with the centre of the optic disc. Both foveal centre and optic disc centre are aligned on the same axis to account for the influence of anatomic variation in the nerve fiber layer bundles.
Figure 2Ganglion cell complex (GCC) thickness map of the right eye in OCT-HS100 Glaucoma 3D algorithm. The area bounded by the inner (1.5 mm) and middle (5 mm) circles forms the inner ring, while the area bounded by the middle (5 mm) and outer (10 mm) circles forms the outer ring. Thicknesses are averaged for the overall measurement, superior and inferior hemifields over the 10 mm diameter circle, and inner (macular GCC) and outer (perimacular GCC) subfields. The section of the nasal outer ring that encroached onto the optic disc and the central 1.5 mm in diameter area are excluded.
Repeatability of ganglion cell complex parameters obtained by OCT-HS100 (n = 34).
| GCC thickness | Mean ± SD ( | ICC (95% CI) |
|---|---|---|
| Overall | 93.96 ± 7.28 | 0.992 (0.983, 0.996) |
| Superior hemifield | 90.13 ± 7.30 | 0.984 (0.969, 0.992) |
| Inferior hemifield | 97.82 ± 7.52 | 0.982 (0.964, 0.991) |
| Macula | ||
| Inner superonasal subfield | 117.72 ± 8.22 | 0.981 (0.963, 0.991) |
| Inner inferonasal subfield | 116.87 ± 8.42 | 0.981 (0.963, 0.991) |
| Inner inferotemporal subfield | 105.57 ± 7.79 | 0.979 (0.958, 0.989) |
| Inner superotemporal subfield | 99.60 ± 7.34 | 0.980 (0.960, 0.990) |
| Perimacula | ||
| Outer superonasal subfield | 107.40 ± 10.65 | 0.987 (0.956, 0.989) |
| Outer inferonasal subfield | 118.44 ± 10.59 | 0.984 (0.968, 0.992) |
| Outer inferotemporal subfield | 79.35 ± 7.32 | 0.868 (0.753, 0.931) |
| Outer superotemporal subfield | 70.54 ± 6.51 | 0.700 (0.481, 0.838) |
CI: confidence interval; GCC: ganglion cell complex; ICC: intraclass correlation coefficient; SD: standard deviation.
Pearson's correlation coefficient (r) between GCC thickness by OCT-HS100 and GC-IPL thickness by Cirrus HD-OCT (n = 92).
| Macular GC-IPL thickness by Cirrus HD-OCT† | GCC thickness by OCT-HS100 | |||
|---|---|---|---|---|
| Overall average‡ | Macular average§ | Superior hemifield‡ | Inferior hemifield‡ | |
| Average | 0.83 | 0.90 | ||
| Superior hemifield | 0.78 | |||
| Inferior hemifield | 0.82 | |||
GCC: ganglion cell complex; GC-IPL: ganglion cell-inner plexiform layer.
All r values were statistically significant (P < 0.001).
†Macular GC-IPL thickness was measured over 4 mm vertical diameter and 4.8 mm horizontal diameter.
‡Overall average and superior and inferior hemifields GCC thicknesses were computed over 10 mm in diameter centered on the fovea.
§Macular average GCC thickness was measured over 5 mm in diameter.