Literature DB >> 26228297

Effect of Tri-Staple Technology and Slow Firing on Secure Stapling Using an Endoscopic Linear Stapler.

Suguru Hasegawa1, Shinya Nakayama, Koya Hida, Kenji Kawada, Yoshiharu Sakai.   

Abstract

AIM: To evaluate the effect of Tri-staple technology and slow firing using the Endo-GIA endoscopic linear stapler.
METHODS: The cardiac and pyloric portions of porcine stomachs were divided using the endoscopic linear stapler with different reload types. A total of 8 min of waiting time was employed during firing in the slow-firing group and no waiting time was employed in the normal-firing group. The shape of the staples was then evaluated. The length of the staple line and serosal laceration was also determined.
RESULTS: There was a moderate negative correlation between tissue thickness and secure staple formation. Tri-staple reloads (purple, black) offered more secure staple formation compared with Universal green reload. Although slow firing enhanced secure staple formation, its effect was greater when using green reload, compared with Tri-staple reloads. Significantly shorter staple line length and longer serosal laceration was observed in the thick tissue. Although the cartridge type did not influence lengths of the staple line or serosal laceration, both were better in the slow-firing group.
CONCLUSIONS: Tri-staple reloads offered more secure staple formation compared with the Universal reload. Although slow firing improved staple line shortening and serosal laceration, its effect on secure stapling was relatively small when using Tri-staple reloads.
© 2015 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2015        PMID: 26228297     DOI: 10.1159/000437216

Source DB:  PubMed          Journal:  Dig Surg        ISSN: 0253-4886            Impact factor:   2.588


  7 in total

1.  Evaluation of intestinal perfusion by ICG fluorescence imaging in laparoscopic colorectal surgery with DST anastomosis.

Authors:  Kenji Kawada; Suguru Hasegawa; Toshiaki Wada; Ryo Takahashi; Shigeo Hisamori; Koya Hida; Yoshiharu Sakai
Journal:  Surg Endosc       Date:  2016-06-28       Impact factor: 4.584

Review 2.  Preoperative, intraoperative and postoperative risk factors for anastomotic leakage after laparoscopic low anterior resection with double stapling technique anastomosis.

Authors:  Kenji Kawada; Yoshiharu Sakai
Journal:  World J Gastroenterol       Date:  2016-07-07       Impact factor: 5.742

3.  ICG fluorescence imaging for quantitative evaluation of colonic perfusion in laparoscopic colorectal surgery.

Authors:  Toshiaki Wada; Kenji Kawada; Ryo Takahashi; Mami Yoshitomi; Koya Hida; Suguru Hasegawa; Yoshiharu Sakai
Journal:  Surg Endosc       Date:  2017-03-09       Impact factor: 4.584

4.  Roux-en-Y Gastric Bypass: Does the Direction of Staples Matter?

Authors:  Mohamed Sahloul; Ahmed Karim; Kuljit S Chohan; Abd A Tahrani; Kamal Mahawar; Martin Richardson; Christian Ludwig; Markos Daskalakis; Rishi Singhal
Journal:  Obes Surg       Date:  2018-09       Impact factor: 4.129

5.  Tubeless single-port thoracoscopic sublobar resection: indication and safety.

Authors:  Chao-Yu Liu; Po-Kuei Hsu; Hung-Che Chien; Chih-Cheng Hsieh; Chien-Kun Ting; Mei-Yung Tsou
Journal:  J Thorac Dis       Date:  2018-06       Impact factor: 2.895

6.  Micro-CT imaging as a method for comparing perfusion in graduated-height and single-height surgical staple lines.

Authors:  Matthew Eschbach; Gregory M Sindberg; Marisha L Godek; Matthew Nagelschmidt; Nicholas Paquette; Michael Wegener; James Alberino; Jane Mayotte; Amit Vasanji; Andrew M Miesse
Journal:  Med Devices (Auckl)       Date:  2018-08-15

7.  Robot-assisted low anterior resection after aluminum potassium sulfate and tannic acid sclerosing therapy for internal hemorrhoids.

Authors:  Yoshiro Itatani; Tomoaki Okada; Kenji Kawada; Koya Hida; Nobu Oshima; Susumu Inamoto; Rei Mizuno; Yoshihisa Okuchi; Yoshiharu Sakai
Journal:  Surg Case Rep       Date:  2019-10-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.