| Literature DB >> 26215491 |
Mohammad Najafi1, Purang Abolmaesumi1, Robert Rohling2.
Abstract
Many common needle intervention procedures are performed with ultrasound guidance because it is a flexible, cost-effective and widely available intra-operative imaging modality. In a needle insertion procedure with ultrasound guidance, real-time calculation and visualization of the needle trajectory can help to guide the choice of puncture site and needle angle to reach the target depicted in the ultrasound image. We found that it is feasible to calculate the needle trajectory with a single camera mounted directly on the ultrasound transducer by using the needle markings. Higher accuracy is achieved compared with other similar transducer-mounted needle trackers. We used an inexpensive, real-time and easy-to-use tracking method based on an automatic feature extraction algorithm and a closed-form method for pose estimation of the needle. The overall accuracy was 0.94 ± 0.46 mm.Keywords: Camera; Closed-form; Guidance; Needle insertion; Needle tracking; Ultrasound
Mesh:
Year: 2015 PMID: 26215491 DOI: 10.1016/j.ultrasmedbio.2015.05.016
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998