Literature DB >> 26211410

Image reconstruction using a gradient impulse response model for trajectory prediction.

S Johanna Vannesjo1, Nadine N Graedel1, Lars Kasper1,2, Simon Gross1, Julia Busch1, Maximilian Haeberlin1, Christoph Barmet1,3, Klaas P Pruessmann1.   

Abstract

PURPOSE: Gradient imperfections remain a challenge in MRI, especially for sequences relying on long imaging readouts. This work aims to explore image reconstruction based on k-space trajectories predicted by an impulse response model of the gradient system. THEORY AND METHODS: Gradient characterization was performed twice with 3 years interval on a commercial 3 Tesla (T) system. The measured gradient impulse response functions were used to predict actual k-space trajectories for single-shot echo-planar imaging (EPI), spiral and variable-speed EPI sequences. Image reconstruction based on the predicted trajectories was performed for phantom and in vivo data. Resulting images were compared with reconstructions based on concurrent field monitoring, separate trajectory measurements, and nominal trajectories.
RESULTS: Image reconstruction using model-based trajectories yielded high-quality images, comparable to using separate trajectory measurements. Compared with using nominal trajectories, it strongly reduced ghosting, blurring, and geometric distortion. Equivalent image quality was obtained with the recent characterization and that performed 3 years prior.
CONCLUSION: Model-based trajectory prediction enables high-quality image reconstruction for technically challenging sequences such as single-shot EPI and spiral imaging. It thus holds great promise for fast structural imaging and advanced neuroimaging techniques, including functional MRI, diffusion tensor imaging, and arterial spin labeling. The method can be based on a one-time system characterization as demonstrated by successful use of 3-year-old calibration data. Magn Reson Med 76:45-58, 2016.
© 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

Keywords:  EPI; GIRF; linear time-invariant (LTI); magnetic field monitoring; single-shot imaging; spiral imaging

Mesh:

Year:  2015        PMID: 26211410     DOI: 10.1002/mrm.25841

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  8 in total

1.  Characterization of hardware-related spatial distortions for IR-PETRA pulse sequence using a brain specific phantom.

Authors:  Sima Ahmadian; Iraj Jabbari; Seyed Mehdi Bagherimofidi; Hamidreza Saligheh Rad
Journal:  MAGMA       Date:  2020-07-06       Impact factor: 2.310

Review 2.  Prospective motion correction in functional MRI.

Authors:  Maxim Zaitsev; Burak Akin; Pierre LeVan; Benjamin R Knowles
Journal:  Neuroimage       Date:  2016-11-11       Impact factor: 6.556

3.  Ultimate MRI.

Authors:  Lawrence L Wald
Journal:  J Magn Reson       Date:  2019-07-09       Impact factor: 2.229

4.  Self-gated 3D stack-of-spirals UTE pulmonary imaging at 0.55T.

Authors:  Ahsan Javed; Rajiv Ramasawmy; Kendall O'Brien; Christine Mancini; Pan Su; Waqas Majeed; Thomas Benkert; Himanshu Bhat; Anthony F Suffredini; Ashkan Malayeri; Adrienne E Campbell-Washburn
Journal:  Magn Reson Med       Date:  2021-11-16       Impact factor: 3.737

5.  Field camera versus phantom-based measurement of the gradient system transfer function (GSTF) with dwell time compensation.

Authors:  M Stich; J A J Richter; T Wech; T A Bley; R Ringler; H Köstler; A E Campbell-Washburn
Journal:  Magn Reson Imaging       Date:  2020-06-10       Impact factor: 2.546

Review 6.  Diffusion Imaging in the Post HCP Era.

Authors:  Steen Moeller; Pramod Pisharady Kumar; Jesper Andersson; Mehmet Akcakaya; Noam Harel; Ruoyun Emily Ma; Xiaoping Wu; Essa Yacoub; Christophe Lenglet; Kamil Ugurbil
Journal:  J Magn Reson Imaging       Date:  2020-06-20       Impact factor: 5.119

7.  Thermal variation in gradient response: measurement and modeling.

Authors:  Jennifer Nussbaum; Benjamin E Dietrich; Bertram J Wilm; Klaas P Pruessmann
Journal:  Magn Reson Med       Date:  2021-12-21       Impact factor: 3.737

8.  MaxGIRF: Image reconstruction incorporating concomitant field and gradient impulse response function effects.

Authors:  Nam G Lee; Rajiv Ramasawmy; Yongwan Lim; Adrienne E Campbell-Washburn; Krishna S Nayak
Journal:  Magn Reson Med       Date:  2022-04-21       Impact factor: 3.737

  8 in total