| Literature DB >> 26207113 |
Irene Humer1, Christian Kirisits2, Daniel Berger1, Petra Trnková1, Richard Pötter2, Nicole Nesvacil2.
Abstract
PURPOSE: The path of subsequent dwell positions of an afterloader source being moved through a ring applicator for cervix cancer brachytherapy deviates from an ideal circle and the position of marker wires. This can lead to deviations of several millimetres between real and assumed dwell positions for treatment planning with simplified source path models. The aim of this study was to test video- and autoradiography-based methods for source path determination, and to study the influence of dwell position accuracy on dose-volume histogram (DVH)-parameters.Entities:
Keywords: applicator geometry; brachytherapy; gynaecological; quality assurance; source path reconstruction; treatment planning
Year: 2015 PMID: 26207113 PMCID: PMC4499519 DOI: 10.5114/jcb.2015.52622
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Template for the R34 ring applicator, showing the inner and outer diameter (dashed lines), positions of the holes for interstitial needles (grey circles), dwell positions observed in video analysis (blue circles), positions predicted by the corrected source path model (red crosses), the nominal path (solid line), and most frequently active dwell positions (black circles) of a standard loading pattern
Results of the iterative source path correction
| Applicator | R34 | R30 | R26 |
|---|---|---|---|
|
| 17 | 15 | 13 |
|
| 3.2 ± 0.9 | 4.5 ± 1.3 | 3.6 ± 1.3 |
|
| 4.6 | 6.1 | 5.7 |
| Fixed dwell position | 9 | 7 | 6 |
|
| 17.50 | 15.85 | 13.95 |
|
| 0.6 ± 0.3 | 0.7 ± 0.4 | 0.8 ± 0.5 |
|
| 1.3 | 1.3 | 2.0 |
|
| 0.1 ± 0.7 | 0.1 ± 0.7 | –0.1 ± 0.8 |
Fig. 2Shifts d projected along the source path between dwell positions derived from video analysis and those corresponding to the corrected source path model. Filled bars indicate the 8 most relevant positions for the R34 standard loading pattern. Dwell position no. 9, which was fixed during the iteration process, is marked with a cross
Results from autoradiography measurements of the corrected circular source path
| Applicator | R34 | R30 | R26 |
|---|---|---|---|
|
| 0.8 ± 1.3 | 0.1 ± 1.5 | 0.1 ± 1.5 |
Ranges for average deviations (mean ± 1σ) for D2cm of organs at risk and D90 of targets caused by rotation of the loading pattern by one or two dwell positions in clockwise (+) or counter clockwise (–) direction
| Rotation | –5.0 mm | –2.5 mm | +2.5 mm | +5.0 mm |
|---|---|---|---|---|
| D2cm3 bladder | (+0.65 ± 5.25)% | (+0.15 ± 2.65)% | (+1.30 ± 1.90)% | (+3.15 ± 3.55)% |
| D2cm3 rectum | (–1.80 ± 6.00)% | (–1.65 ± 2.95)% | (+2.55 ± 2.65)% | (+6.65 ± 5.05)% |
| D2cm3 sigmoid | (–0.05 ± 3.55)% | (–0.35 ± 1.95)% | (–0.20 ± 2.30)% | (+0.05 ± 4.15)% |
| D90 CTVHR | (+0.35 ± 2.65)% | (+0.35 ± 1.55)% | (–0.45 ± 1.25)% | (–1.20 ± 2.90)% |
| D90 GTV | (+0.30 ± 3.20)% | (+0.25 ± 1.65)% | (–0.40 ± 1.70)% | (–1.10 ± 3.50)% |