OBJECTIVE: The purpose of this article is to study the added value of model-based iterative reconstruction (MBIR) on metal artifact reduction on CT compared with standard filtered back projection (FBP).
MATERIALS AND METHODS: Ex vivo imaging was performed on several metal implants. Datasets were reconstructed with standard FBP and MBIR algorithms. The sizes of the artifacts surrounding the metal implant were recorded and compared. In vivo imaging was performed on 62 patients with metal implants. Each dataset was reconstructed with FBP and MBIR algorithms. Objective image quality was assessed by measuring the size of the artifact generated by the metal implant. Subjective image quality was graded on a 3-point scale, taking into account the visibility of the bone-metal interface, as well as the visibility of the neighboring soft tissues.
RESULTS: Ex vivo analysis yielded a reduction of 82% in the size of the artifact when using the MBIR algorithm, compared with the FBP algorithm. The mean (SD) size of the artifacts was 1.4 ± 0.8 and 0.25 ± 0.06 cm(2) with FBP and MBIR, respectively. In vivo, the mean size of the artifacts decreased from 7.3 ± 1.5 cm(2) to 4.0 ± 0.9 cm(2) for FBP and MBIR, respectively (p = 0.012). The subjective image quality analysis showed an equal or better bone-metal interface of MBIR algorithm in 85% of cases. Visibility of the soft tissue surrounding the metal implant was determined to be equal or better in 97% of cases in which MBIR was used.
CONCLUSION: This study shows that the MBIR algorithm allows a clear reduction of metal artifacts on CT images and, hence, a better analysis of the soft tissue surrounding the metal implant compared with FBP.

Entities:  

Keywords:  CT; metal artifact reduction; model-based iterative reconstruction

Mesh:

Substances:

Year:  2015        PMID: 26204291     DOI: 10.2214/AJR.14.13334

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  13 in total

1.  Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results.

Authors:  Seema Toso; Meryle Laurent; Elise Dupuis Lozeron; Pauline Brindel; Marirosa Cristallo Lacalamita; Sylviane Hanquinet
Journal:  Pediatr Radiol       Date:  2018-07-28

2.  Submillisievert CT using model-based iterative reconstruction with lung-specific setting: An initial phantom study.

Authors:  Akinori Hata; Masahiro Yanagawa; Osamu Honda; Tomoko Gyobu; Ken Ueda; Noriyuki Tomiyama
Journal:  Eur Radiol       Date:  2016-03-17       Impact factor: 5.315

3.  The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol.

Authors:  Ji Young Jeon; Sheen-Woo Lee; Yu Mi Jeong; Han Joo Baek
Journal:  Eur Radiol       Date:  2018-07-09       Impact factor: 5.315

4.  Metal artefact reduction for patients with metallic dental fillings in helical neck computed tomography: comparison of adaptive iterative dose reduction 3D (AIDR 3D), forward-projected model-based iterative reconstruction solution (FIRST) and AIDR 3D with single-energy metal artefact reduction (SEMAR).

Authors:  Koichiro Yasaka; Kouhei Kamiya; Ryusuke Irie; Eriko Maeda; Jiro Sato; Kuni Ohtomo
Journal:  Dentomaxillofac Radiol       Date:  2016-06-22       Impact factor: 2.419

5.  Diagnostic Value of Model-Based Iterative Reconstruction Combined with a Metal Artifact Reduction Algorithm during CT of the Oral Cavity.

Authors:  Y Kubo; K Ito; M Sone; H Nagasawa; Y Onishi; N Umakoshi; T Hasegawa; T Akimoto; M Kusumoto
Journal:  AJNR Am J Neuroradiol       Date:  2020-09-24       Impact factor: 3.825

6.  Metal artefact reduction in the oral cavity using deep learning reconstruction algorithm in ultra-high-resolution computed tomography: a phantom study.

Authors:  Yuki Sakai; Erina Kitamoto; Kazutoshi Okamura; Masato Tatsumi; Takashi Shirasaka; Ryoji Mikayama; Masatoshi Kondo; Hiroshi Hamasaki; Toyoyuki Kato; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2021-04-29       Impact factor: 3.525

7.  Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction.

Authors:  R H H Wellenberg; M F Boomsma; J A C van Osch; A Vlassenbroek; J Milles; M A Edens; G J Streekstra; C H Slump; M Maas
Journal:  Skeletal Radiol       Date:  2017-02-15       Impact factor: 2.199

8.  Metallic artifact reduction by evaluation of the additional value of iterative reconstruction algorithms in hip prosthesis computed tomography imaging.

Authors:  Angeliki Neroladaki; Steve Philippe Martin; Ilias Bagetakos; Diomidis Botsikas; Marion Hamard; Xavier Montet; Sana Boudabbous
Journal:  Medicine (Baltimore)       Date:  2019-02       Impact factor: 1.817

9.  CT radiation dose reduction in patients with total hip arthroplasties using model-based iterative reconstruction and orthopaedic metal artefact reduction.

Authors:  Ruud H H Wellenberg; Jochen A C van Osch; Henk J Boelhouwers; Mireille A Edens; Geert J Streekstra; Harmen B Ettema; Martijn F Boomsma
Journal:  Skeletal Radiol       Date:  2019-04-24       Impact factor: 2.199

10.  Knowledge-based iterative model reconstruction: Comparative image quality with low tube voltage cerebral CT angiography.

Authors:  Xinrui Wang; Chengcheng Zhu; Jing Li; Andrew J Degnan; Tao Jiang; Jianping Lu
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.