Literature DB >> 26198448

Hemodynamics of a functional centrifugal-flow total artificial heart with functional atrial contraction in goats.

Takuya Shiga1, Yasuyuki Shiraishi2, Kyosuke Sano3, Yasunori Taira3, Yusuke Tsuboko3, Akihiro Yamada3, Hidekazu Miura2, Shintaro Katahira4, Masatoshi Akiyama4, Yoshikatsu Saiki4, Tomoyuki Yambe2.   

Abstract

Implantation of a total artificial heart (TAH) is one of the therapeutic options for the treatment of patients with end-stage biventricular heart failure. There is no report on the hemodynamics of the functional centrifugal-flow TAH with functional atrial contraction (fCFTAH). We evaluated the effects of pulsatile flow by atrial contraction in acute animal models. The goats received fCFTAH that we created from two centrifugal-flow ventricular assist devices. Some hemodynamic parameters maintained acceptable levels: heart rate 115.5 ± 26.3 bpm, aortic pressure 83.5 ± 10.1 mmHg, left atrial pressure 18.0 ± 5.9 mmHg, pulmonary pressure 28.5 ± 9.7 mmHg, right atrial pressure 13.6 ± 5.2 mmHg, pump flow 4.0 ± 1.1 L/min (left) 3.9 ± 1.1 L/min (right), and cardiac index 2.13 ± 0.14 L/min/m(2). fCFTAH with atrial contraction was able to maintain the TAH circulation by forming a pulsatile flow in acute animal experiments. Taking the left and right flow rate balance using the low internal pressure loss of the VAD pumps may be easier than by other pumps having considerable internal pressure loss. We showed that the remnant atrial contraction effected the flow rate change of the centrifugal pump, and the atrial contraction waves reflected the heart rate. These results indicate that remnant atria had the possibility to preserve autonomic function in fCFTAH. We may control fCFTAH by reflecting the autonomic function, which is estimated with the flow rate change of the centrifugal pump.

Entities:  

Keywords:  Atrial contraction; Autonomic regulation; Centrifugal pump; Continuous flow; Total artificial heart

Mesh:

Year:  2015        PMID: 26198448     DOI: 10.1007/s10047-015-0852-y

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  16 in total

1.  Heart disease and stroke statistics--2011 update: a report from the American Heart Association.

Authors:  Véronique L Roger; Alan S Go; Donald M Lloyd-Jones; Robert J Adams; Jarett D Berry; Todd M Brown; Mercedes R Carnethon; Shifan Dai; Giovanni de Simone; Earl S Ford; Caroline S Fox; Heather J Fullerton; Cathleen Gillespie; Kurt J Greenlund; Susan M Hailpern; John A Heit; P Michael Ho; Virginia J Howard; Brett M Kissela; Steven J Kittner; Daniel T Lackland; Judith H Lichtman; Lynda D Lisabeth; Diane M Makuc; Gregory M Marcus; Ariane Marelli; David B Matchar; Mary M McDermott; James B Meigs; Claudia S Moy; Dariush Mozaffarian; Michael E Mussolino; Graham Nichol; Nina P Paynter; Wayne D Rosamond; Paul D Sorlie; Randall S Stafford; Tanya N Turan; Melanie B Turner; Nathan D Wong; Judith Wylie-Rosett
Journal:  Circulation       Date:  2010-12-15       Impact factor: 29.690

2.  Hemodynamic effects of pressure-volume relation in the atrial contraction model on the total artificial heart using centrifugal blood pumps.

Authors:  Takuya Shiga; Takehito Kuroda; Yusuke Tsuboko; Hidekazu Miura; Yasuyuki Shiraishi; Tomoyuki Yambe
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2013

3.  Sixth INTERMACS annual report: a 10,000-patient database.

Authors:  James K Kirklin; David C Naftel; Francis D Pagani; Robert L Kormos; Lynne W Stevenson; Elizabeth D Blume; Marissa A Miller; J T Baldwin; J Timothy Baldwin; James B Young
Journal:  J Heart Lung Transplant       Date:  2014-04-21       Impact factor: 10.247

4.  The BiVACOR rotary biventricular assist device: concept and in vitro investigation.

Authors:  Daniel Timms; John Fraser; Mark Hayne; John Dunning; Keith McNeil; Mark Pearcy
Journal:  Artif Organs       Date:  2008-10       Impact factor: 3.094

Review 5.  Contemporary use of ventricular assist devices.

Authors:  Cesare M Terracciano; Leslie W Miller; Magdi H Yacoub
Journal:  Annu Rev Med       Date:  2010       Impact factor: 13.739

6.  Advanced heart failure treated with continuous-flow left ventricular assist device.

Authors:  Mark S Slaughter; Joseph G Rogers; Carmelo A Milano; Stuart D Russell; John V Conte; David Feldman; Benjamin Sun; Antone J Tatooles; Reynolds M Delgado; James W Long; Thomas C Wozniak; Waqas Ghumman; David J Farrar; O Howard Frazier
Journal:  N Engl J Med       Date:  2009-11-17       Impact factor: 91.245

7.  Remnant atrial function in human and animal recipients of a total artificial heart.

Authors:  K D Murray; P F Binkley; K M Watson; E P Howanitz; P Ross; P D Myerowitz
Journal:  J Heart Lung Transplant       Date:  1992 Nov-Dec       Impact factor: 10.247

Review 8.  Axial and centrifugal continuous-flow rotary pumps: a translation from pump mechanics to clinical practice.

Authors:  Nader Moazami; Kiyotaka Fukamachi; Mariko Kobayashi; Nicholas G Smedira; Katherine J Hoercher; Alex Massiello; Sangjin Lee; David J Horvath; Randall C Starling
Journal:  J Heart Lung Transplant       Date:  2013-01       Impact factor: 10.247

9.  Total artificial heart freedom driver in a patient with end-stage biventricular heart failure.

Authors:  Kristin Friedline; Pamela Hassinger
Journal:  AANA J       Date:  2012-04

10.  Long-term mechanical circulatory support (destination therapy): on track to compete with heart transplantation?

Authors:  James K Kirklin; David C Naftel; Francis D Pagani; Robert L Kormos; Lynne Stevenson; Marissa Miller; James B Young
Journal:  J Thorac Cardiovasc Surg       Date:  2012-07-15       Impact factor: 5.209

View more
  1 in total

Review 1.  Journal of Artificial Organs 2016: the year in review : Journal of Artificial Organs Editorial Committee.

Authors:  Y Sawa; G Matsumiya; K Matsuda; E Tatsumi; T Abe; K Fukunaga; S Ichiba; A Kishida; K Kokubo; T Masuzawa; A Myoui; M Nishimura; T Nishimura; T Nishinaka; E Okamoto; S Tokunaga; T Tomo; T Tsukiya; Y Yagi; T Yamaoka
Journal:  J Artif Organs       Date:  2017-02-14       Impact factor: 1.731

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.