The presence of bacteria in urine can be used to monitor the onset or prognosis of urinary tract infection (UTI) and some sexually-transmitted diseases (STDs), such as gonorrhea. Typically, bacteria's presence in urine is confirmed by culturing samples overnight on agar plates, followed by a microscopic examination. Additionally, the presence of Escherichia coli in a urine sample can be indirectly confirmed through assaying for nitrite (generated by reducing nitrate in urine), however this is not sufficiently specific and sensitive. Species/strains identification of bacteria in a urine sample provides insight to appropriate antibiotic treatment options. In this work, a microfluidic paper analytical device (µPAD) was designed and fabricated for evaluating UTI (E. coli) and STD (Neisseria gonorrhoeae) from human urine samples. Anti-E. coli or anti-N. gonorrhoeae antibodies were conjugated to submicron particles then pre-loaded and dried in the center of each paper microfluidic channel. Human urine samples (undiluted) spiked with E. coli or N. gonorrhoeae were incubated for 5 min with 1% Tween 80. The bacteria-spiked urine samples were then introduced to the inlet of paper microfluidic channel, which flowed through the channel by capillary force. Data confirms that proteins were not filtered by μPAD, which is essential for this assay. Urobilin, the component responsible for the yellow appearance of urine and green fluorescence emission, was filtered by μPAD, resulting in significantly minimized false-positive signals. This filtration was simultaneously made during the μPAD assay and no pretreatment/purification step was necessary. Antibody-conjugated particles were immunoagglutinated at the center of the paper channel. The extent of immunoagglutination was quantified by angle-specific Mie scatter under ambient lighting conditions, utilizing a smartphone camera as a detector. The total μPAD assay time was less than 30s. The detection limit was 10 CFU/mL for both E. coli and N. gonorrhoeae, while commercially available gonorrhea rapid kit showed a detection limit of 10(6) CFU/mL. A commercially available nitrite assay test strip also had a detection limit of 10(6) CFU/mL, but this method is not antibody-based and thus not sufficiently specific. By optimizing the particle concentration, we were also able to extend the linear range of the assay up to 10(7) CFU/mL. The proposed prototype will serve as a low-cost, point-of-care, sensitive urinalysis biosensor to monitor UTI and gonorrhea from human urine.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Escherichia coli; Neisseria gonorrhoeae; Paper microfluidics; Point-of-care; Smartphone; Urinalysis; Urobilin

Mesh:

Substances:

Year:  2015        PMID: 26190472     DOI: 10.1016/j.bios.2015.07.014

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  22 in total

1.  In situ, dual-mode monitoring of organ-on-a-chip with smartphone-based fluorescence microscope.

Authors:  Soohee Cho; Argel Islas-Robles; Ariana M Nicolini; Terrence J Monks; Jeong-Yeol Yoon
Journal:  Biosens Bioelectron       Date:  2016-07-07       Impact factor: 10.618

2.  Angular photodiode array-based device to detect bacterial pathogens in a wound model.

Authors:  Robin E Sweeney; Jeong-Yeol Yoon
Journal:  IEEE Sens J       Date:  2017-09-14       Impact factor: 3.301

Review 3.  Biological characteristics and biomarkers of novel SARS-CoV-2 facilitated rapid development and implementation of diagnostic tools and surveillance measures.

Authors:  Gajanan Sampatrao Ghodake; Surendra Krushna Shinde; Avinash Ashok Kadam; Rijuta Ganesh Saratale; Ganesh Dattatraya Saratale; Asad Syed; Abdallah M Elgorban; Najat Marraiki; Dae-Young Kim
Journal:  Biosens Bioelectron       Date:  2021-01-04       Impact factor: 10.618

4.  Label-free Mie Scattering Identification of Tumor Tissue Using an Angular Photodiode Array.

Authors:  Matthew V Bills; Jeong-Yeol Yoon
Journal:  IEEE Sens Lett       Date:  2020-06-11

5.  Smartphone Detection of UV LED-Enhanced Particle Immunoassay on Paper Microfluidics.

Authors:  Tu San Park; Soohee Cho; Tigran G Nahapetian; Jeong-Yeol Yoon
Journal:  SLAS Technol       Date:  2016-07-10       Impact factor: 3.047

Review 6.  Point-of-Care Diagnostics: Recent Developments in a Connected Age.

Authors:  Samiksha Nayak; Nicole R Blumenfeld; Tassaneewan Laksanasopin; Samuel K Sia
Journal:  Anal Chem       Date:  2016-12-13       Impact factor: 6.986

7.  Human sensor-inspired supervised machine learning of smartphone-based paper microfluidic analysis for bacterial species classification.

Authors:  Sangsik Kim; Min Hee Lee; Theanchai Wiwasuku; Alexander S Day; Sujittra Youngme; Dong Soo Hwang; Jeong-Yeol Yoon
Journal:  Biosens Bioelectron       Date:  2021-05-14       Impact factor: 12.545

8.  Smartphone-Based Point-of-Care Urinalysis Under Variable Illumination.

Authors:  Moonsoo Ra; Mannan Saeed Muhammad; Chiawei Lim; Sehui Han; Chansung Jung; Whoi-Yul Kim
Journal:  IEEE J Transl Eng Health Med       Date:  2017-12-15       Impact factor: 3.316

Review 9.  A Systematic Review of Point of Care Testing for Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.

Authors:  Sasha Herbst de Cortina; Claire C Bristow; Dvora Joseph Davey; Jeffrey D Klausner
Journal:  Infect Dis Obstet Gynecol       Date:  2016-05-26

Review 10.  Improving the Sensitivity and Functionality of Mobile Webcam-Based Fluorescence Detectors for Point-of-Care Diagnostics in Global Health.

Authors:  Reuven Rasooly; Hugh Alan Bruck; Joshua Balsam; Ben Prickril; Miguel Ossandon; Avraham Rasooly
Journal:  Diagnostics (Basel)       Date:  2016-05-17