Literature DB >> 26183218

Diffusion MRI with concurrent magnetic field monitoring.

Bertram J Wilm1, Zoltan Nagy2, Christoph Barmet1,3, S Johanna Vannesjo1, Lars Kasper1,4, Max Haeberlin1, Simon Gross1, Benjamin E Dietrich1, David O Brunner1, Thomas Schmid1, Klaas P Pruessmann1.   

Abstract

PURPOSE: Diffusion MRI is compromised by unknown field perturbation during image encoding. The purpose of this study was to address this problem using the recently described approach of concurrent magnetic field monitoring.
METHODS: Magnetic field dynamics were monitored during the echo planar imaging readout of a common diffusion-weighted MRI sequence using an integrated magnetic field camera setup. The image encoding including encoding changes over the duration of entire scans were quantified and analyzed. Field perturbations were corrected by accounting for them in generalized image reconstruction. The impact on image quality along with geometrical congruence among different diffusion-weighted images was assessed both qualitatively and quantitatively.
RESULTS: The most significant field perturbations were found to be related to higher-order eddy currents from diffusion-weighting gradients and B0 field drift as well as gradual changes of short-term eddy current behavior and mechanical oscillations during the scan. All artifacts relating to dynamic field perturbations were eliminated by incorporating the measured encoding in image reconstruction.
CONCLUSION: Concurrent field monitoring combined with generalized reconstruction enhances depiction fidelity in diffusion imaging. In addition to artifact reduction, it improves geometric congruence and thus facilitates image combination for quantitative diffusion analysis.
© 2015 Wiley Periodicals, Inc.

Keywords:  DTI; DWI; artifact correction; diffusion; distortion correction; field monitoring

Mesh:

Year:  2015        PMID: 26183218     DOI: 10.1002/mrm.25827

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  9 in total

1.  Simple noise reduction for diffusion weighted images.

Authors:  Yuto Konishi; Yuki Kanazawa; Takatoshi Usuda; Yuki Matsumoto; Hiroaki Hayashi; Tsuyoshi Matsuda; Junji Ueno; Masafumi Harada
Journal:  Radiol Phys Technol       Date:  2016-03-16

2.  Highly accelerated time-of-flight magnetic resonance angiography using spiral imaging improves conspicuity of intracranial arterial branches while reducing scan time.

Authors:  Tobias Greve; Nico Sollmann; Andreas Hock; Silke Hey; Velmurugan Gnanaprakasam; Marco Nijenhuis; Claus Zimmer; Jan S Kirschke
Journal:  Eur Radiol       Date:  2019-10-29       Impact factor: 5.315

Review 3.  Connectome 2.0: Developing the next-generation ultra-high gradient strength human MRI scanner for bridging studies of the micro-, meso- and macro-connectome.

Authors:  Susie Y Huang; Thomas Witzel; Boris Keil; Alina Scholz; Mathias Davids; Peter Dietz; Elmar Rummert; Rebecca Ramb; John E Kirsch; Anastasia Yendiki; Qiuyun Fan; Qiyuan Tian; Gabriel Ramos-Llordén; Hong-Hsi Lee; Aapo Nummenmaa; Berkin Bilgic; Kawin Setsompop; Fuyixue Wang; Alexandru V Avram; Michal Komlosh; Dan Benjamini; Kulam Najmudeen Magdoom; Sudhir Pathak; Walter Schneider; Dmitry S Novikov; Els Fieremans; Slimane Tounekti; Choukri Mekkaoui; Jean Augustinack; Daniel Berger; Alexander Shapson-Coe; Jeff Lichtman; Peter J Basser; Lawrence L Wald; Bruce R Rosen
Journal:  Neuroimage       Date:  2021-08-28       Impact factor: 7.400

Review 4.  Mapping the human connectome using diffusion MRI at 300 mT/m gradient strength: Methodological advances and scientific impact.

Authors:  Qiuyun Fan; Cornelius Eichner; Maryam Afzali; Lars Mueller; Chantal M W Tax; Mathias Davids; Mirsad Mahmutovic; Boris Keil; Berkin Bilgic; Kawin Setsompop; Hong-Hsi Lee; Qiyuan Tian; Chiara Maffei; Gabriel Ramos-Llordén; Aapo Nummenmaa; Thomas Witzel; Anastasia Yendiki; Yi-Qiao Song; Chu-Chung Huang; Ching-Po Lin; Nikolaus Weiskopf; Alfred Anwander; Derek K Jones; Bruce R Rosen; Lawrence L Wald; Susie Y Huang
Journal:  Neuroimage       Date:  2022-02-23       Impact factor: 7.400

Review 5.  What's new and what's next in diffusion MRI preprocessing.

Authors:  Chantal M W Tax; Matteo Bastiani; Jelle Veraart; Eleftherios Garyfallidis; M Okan Irfanoglu
Journal:  Neuroimage       Date:  2021-12-26       Impact factor: 7.400

6.  A field-monitoring-based approach for correcting eddy-current-induced artifacts of up to the 2nd spatial order in human-connectome-project-style multiband diffusion MRI experiment at 7T: A pilot study.

Authors:  Ruoyun Ma; Mehmet Akçakaya; Steen Moeller; Edward Auerbach; Kâmil Uğurbil; Pierre-François Van de Moortele
Journal:  Neuroimage       Date:  2020-04-16       Impact factor: 6.556

7.  Decoupled dynamic magnetic field measurements improves diffusion-weighted magnetic resonance images.

Authors:  Ying-Hua Chu; Yi-Cheng Hsu; Fa-Hsuan Lin
Journal:  Sci Rep       Date:  2017-09-14       Impact factor: 4.379

8.  A comprehensive approach for correcting voxel-wise b-value errors in diffusion MRI.

Authors:  Yoojin Lee; Adam O Kettinger; Bertram Jakob Wilm; Ralf Deichmann; Nikolaus Weiskopf; Christian Lambert; Klaas Paul Pruessmann; Zoltan Nagy
Journal:  Magn Reson Med       Date:  2019-12-16       Impact factor: 4.668

9.  Strong diffusion gradients allow the separation of intra- and extra-axonal gradient-echo signals in the human brain.

Authors:  Elena Kleban; Chantal M W Tax; Umesh S Rudrapatna; Derek K Jones; Richard Bowtell
Journal:  Neuroimage       Date:  2020-04-23       Impact factor: 7.400

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.