Masaki Watabe1, Satya N V Arjunan1, Seiya Fukushima2, Kazunari Iwamoto1, Jun Kozuka3, Satomi Matsuoka3, Yuki Shindo4, Masahiro Ueda2, Koichi Takahashi5. 1. Laboratory for Biochemical Simulation, Quantitative Biology Center, RIKEN, Suita, Osaka, Japan. 2. Laboratory for Cell Signaling Dynamics, Quantitative Biology Center, RIKEN, Suita, Osaka, Japan; Graduate School of Frontier Biosciences, Osaka University, Suita, Osaka, Japan. 3. Laboratory for Cell Signaling Dynamics, Quantitative Biology Center, RIKEN, Suita, Osaka, Japan. 4. Laboratory for Biochemical Simulation, Quantitative Biology Center, RIKEN, Suita, Osaka, Japan; Graduate School of Frontier Biosciences, Osaka University, Suita, Osaka, Japan. 5. Laboratory for Biochemical Simulation, Quantitative Biology Center, RIKEN, Suita, Osaka, Japan; Institute for Advanced Biosciences, Keio University, Tsuruoka, Yamagata, Japan.
Abstract
Using bioimaging technology, biologists have attempted to identify and document analytical interpretations that underlie biological phenomena in biological cells. Theoretical biology aims at distilling those interpretations into knowledge in the mathematical form of biochemical reaction networks and understanding how higher level functions emerge from the combined action of biomolecules. However, there still remain formidable challenges in bridging the gap between bioimaging and mathematical modeling. Generally, measurements using fluorescence microscopy systems are influenced by systematic effects that arise from stochastic nature of biological cells, the imaging apparatus, and optical physics. Such systematic effects are always present in all bioimaging systems and hinder quantitative comparison between the cell model and bioimages. Computational tools for such a comparison are still unavailable. Thus, in this work, we present a computational framework for handling the parameters of the cell models and the optical physics governing bioimaging systems. Simulation using this framework can generate digital images of cell simulation results after accounting for the systematic effects. We then demonstrate that such a framework enables comparison at the level of photon-counting units.
Using bioimaging technology, biologists have attempted to identify and document analytical interpretations that underlie biological phenomena in biological cells. Theoretical biology aims at distilling those interpretations into knowledge in the mathematical form of biochemical reaction networks and understanding how higher level functions emerge from the combined action of biomolecules. However, there still remain formidable challenges in bridging the gap between bioimaging and mathematical modeling. Generally, measurements using fluorescence microscopy systems are influenced by systematic effects that arise from stochastic nature of biological cells, the imaging apparatus, and optical physics. Such systematic effects are always present in all bioimaging systems and hinder quantitative comparison between the cell model and bioimages. Computational tools for such a comparison are still unavailable. Thus, in this work, we present a computational framework for handling the parameters of the cell models and the optical physics governing bioimaging systems. Simulation using this framework can generate digital images of cell simulation results after accounting for the systematic effects. We then demonstrate that such a framework enables comparison at the level of photon-counting units.
All scientific measurements are subject to some uncertainties. Experimental accuracy
and precision must be always estimated to establish the validity of our results
[1, 2]. It is also true for measurements using
bioimaging techniques such as fluorescence microscopy. The measurements are
generally influenced by systematic effects that arise from the stochastic nature of
biological cells, the imaging apparatus, and optical physics. Such systematic
effects are always present in all bioimaging systems and hinder the validation of
the mathematical models of biological cells. For example, the local precision of
reconstructed images obtained by precise localization microscopy, such as stochastic
optical reconstruction microscopy (STORM), and photoactivated localization
microscopy (PALM) is particularly limited by the systematic effects that are
governed by camera specifications and its operating conditions [3-5]. The limitation constrains the validation of
the mathematical models of the biological dynamics.Theory of model validation is often applied to obtain valid mathematical models in
physics and engineering fields [6-8]. It
can be also applied to biological science, because it offers a formal representation
of the progressive build-up of trust in the mathematical model of interest. In a
standard exercise of model validation, one performs an experiment and in parallel,
runs a simulation of the model. Then, using metrics controlled by the parameters
embedded in the model and the experimental configuration, the output of the model
simulation is iteratively compared and analyzed with the actual experimental output.
There are three important parts in the iterative process. (1) The experimental
outputs are generally influenced by the systematic effects that arise from various
sources in the bioimaging process. The outputs of the model simulation are usually
not presented in the most efficient way for comparison with the experimental
outputs. Simulations of the experimental techniques and their operating conditions
are essential for proper comparison and analysis. (2) The predictive capability of
the model is to go beyond the well-known parameter domain and into a new parameter
domain of unknown conditions and outcome. Calibration and validation are one of the
important processes of parameter adjustment in each domain. Calibration is defined
as the process of improving the agreement of a set of simulated outputs with a set
of actual outputs obtained under well-controlled experimental systems. Validation is
defined as the process of quantifying our confidence in the predictive capability
for a given application. (3) Analyses of parameter sensitivity and limitation are
also important to reduce the size of the parameter domain.In this article, we focus on the first (comparison) issue/part. In order to properly
compare spatial models of biological cell with actual cell images, we propose a
computational framework for managing parameter dependences by defining a uniform
interface and common organizational principles governing the systematic effects.
Such a framework allows us to efficiently handle the parameters defined in a spatial
cell model and the physical principles governing the bioimaging techniques and their
operating conditions. Using this framework, we program bioimaging simulation modules
to generate digital images of the cell simulation results after accounting for the
systematic effects. The intensity of the simulated images corresponds to the number
of photons detected in a light-sensitive device. Thus, the framework streamlines the
comparison at the level of photon-counting units. In particular, we implement the
simulation modules for relatively simple microscopy systems: total internal
reflection fluorescence microscopy (TIRFM) and laser-scanning confocal microscopy
(LSCM). We then evaluate the performance of the simulation modules by comparing a
simulated image with an actual image for simple particle models of fluorescent
molecules. Thus, these images are comparable at the level of photon-counting units.
Each simulated image is visually similar to the corresponding real one. In addition,
using the LSCM simulation module, we compared a more complex cell model with real
cell images obtained by the actual LSCM system. We construct the following spatial
cell models for the comparison: (i) the ERK nuclear translocation model for the
epidermal growth factor (EGF) signaling pathway, and (ii) the self-organizing wave
model of phosphatase and a tensinin homolog (PTEN) for the chemotactic pathway of
Dictyostelium discoideum. Using a test version of the TIRFM
simulation module, we compared the oscillation model of the Min proteins of
Escherichia coli with actual cell images [9].
Methods
Computational framework
To render the simulated output of a spatial cell model well suited for comparison
at the level of photon-counting units, we propose a computational framework for
simulating the passage of photons through fluorescent molecules and the optical
system. Simulations using this framework can generate simulated digital images
after accounting for the systematic effects that are governed by the parameters
embedded in spatial cell model and optics system. An overview of the
computational framework is schematically shown in Fig 1. The simulation of the optical system
is composed of three components: (1) an illumination system, (2) molecular
fluorescence, and (3) an image-forming system. The illumination system transfers
photon flux from a light source to the spatial cell model to create a prescribed
photon distribution and maximize the photon flux delivered to the cell model.
Fluorophores defined in the cell model absorb photons from the distribution and
are quantum mechanically excited to higher energy states. Molecular fluorescence
is the result of physical and chemical processes in which the fluorophores emit
photons from the excited states [10, 11].
Finally, the image-forming system relays a nearly exact image of the cell model
to a light-sensitive detector.
Fig 1
Schematic overview of the computational framework.
Direction of photon propagation is presented by thick blue arrows.
Schematic overview of the computational framework.
Direction of photon propagation is presented by thick blue arrows.
Simulation of cell model
In particular, the bioimaging simulation system requires the space-time
trajectory of each simulated molecule of interest to generate realistic
digital images. However, many cell simulation systems have been designed to
model and simulate both deterministic and stochastic biochemical processes,
assuming that simulated molecules are dimensionless and homogeneously
distributed in a compartment [12]. Here, we use spatial simulation
methods that can provide accurate space-time trajectories of molecules
[9, 13–17]. For a given cell
system, simulations using these methods include a statistical model of
biological fluctuation that arises from stochastic changes in the cellular
compartment geometry, number of molecules, type of molecule, molecular
state, and translational and rotational diffusion.
Simulation of optical system
Simulations of an optical system particularly require the computation of the
photon counting, propagation, and distribution. The optics simulations are
based on geometric optics (or wave optics) and the Monte Carlo method. Each
optics simulation includes a statistical model of the systematic effects
that are influenced by the parameters defined in optical devices such as the
light source, objective lens, special filter, and detector. The classical
theory of geometric optics is applied to simulate the photon propagation and
distribution through the illumination and image-forming systems, including
optical aberrations. Geometric optics approximates the photon propagation as
a ray (paraxial approximation), and provides the procedures to compute the
numerical or analytical forms of the photon distributions for a given photon
wavelength. It is an excellent approximation when the photon wavelength is
very small compared with the size of the structure with which the photon
interacts. However, it introduces normalization constant as an input
parameter, and is formalized without counting the number of photons
propagating through the optical system. The Monte Carlo method is applied to
the simulation of the stochastic process of counting photons for a given
probability density function. The details for each optics simulation are
described below.Illumination system [18, 19]: The bioimaging system
requires intense, near-monochromatic, illumination by a widely
spreading light source, such as lasers. Incident photons from
such a light source can illuminate a specimen. The surviving
photons after passing the excitation filters interact with the
fluorophores in the cell model, and excite the fluorophores to
electrically excited states. The optics simulations of the
focusing of the incident photons through the objective lens
include a statistical model of the systematic effects due to the
numerical aperture (NA), magnification, working distance, degree
of aberration, correction refracting surface radius, thickness,
refractive index, and details of each lens element.Molecular fluorescence: The incident photons propagating through
the illumination system are absorbed by the fluorophores in the
cell model. Fluorescence is the result of physical and chemical
processes in which the fluorophores emit photons from
electronically excited states [10, 11]. The Monte Carlo
simulation of the overall fluorescence process includes a
statistical model of the systematic effects that are influenced
by the absorption and emission spectra, quantum yield, lifetime,
quenching, photobleaching and blinking, anisotropy, energy
transfer, solvent effect, diffusion, complex formation, and a
host of environmental variables.Image-forming system [18, 19]: In an optical system
that employs incoherent illumination of the cell model, the
image-forming process can be considered as a linear system
[20].
Impulse response of the image-forming system to a point-like
fluorophore is described by the point spread function (PSF) of
the wavelength and position. When all fluorophores in the cell
model are imaged simultaneously, the distribution of emitted
photons of longer wavelengths that passed through the use of the
objective lens and special filters, is computed as the sum of
the PSFs of all fluorophores. The optics simulations of PSF
formation and convolution include a statistical model of the
systematic effects that are ruled by the parameters embedded in
the objective lens, the special filters, and each details of
lens elements.The emitted photons are finally detected by light-sensitive
devices, and digitized as an image at detection time. The
properties of the final image depend on the detector
specifications and conditions during the readout process that
converts an incident photon signal into a digital signal. The
Monte Carlo simulation for the detection process includes a
statistical model of the systematic effects that arise from
signal and background shot noises, and detector specifications
and conditions, such as pixel size, quantum efficiency (QE),
readout noise, dark current, excess noise factor, gain, offset,
exposure time, and binning.
Implementation
We provide a standard computational framework to simulate various different types
of bioimaing systems. In particular, we implemented the simulation modules for
relatively simple microscopy systems: TIRFM and LSCM. Optical configurations are
shown in Fig 2. The modules
are designed to generate digital images of the cell simulation results after
accounting for the systematic effects that are governed by the parameters
defined in the TIRFM and LSCM systems. A cell simulation method with Spatiocyte
is used to construct the spatial cell models [9]. For a given cell system, Spatiocyte can
provide a statistical model of biological fluctuation that arises from
stochastic changes in the cellular compartment geometry, number of molecules,
type of molecule, molecular state, and translational diffusion. The method can
be used to model complex reaction-diffusion mediated cellular processes
occurring on the surface and in the volume compartments of the cell at a
single-molecule resolution. To represent cell compartments and rapidly resolve
molecular collisions, the method discretizes space into a hexagonal
closed-packed lattice. Each molecule randomly walks from voxel to voxel.
Molecular collisions occur between walks. Immobile lipid molecules represent
surface compartments, such as cellular and nuclear membranes. Implementation
details are described in ref. [21]. In addition, other simulation methods such as Green Function
Reaction Dynamics (GFRD) [13, 14] will
be applied in the future implementation.
Fig 2
Optical configurations.
(A) TIRFM simulation module. (B) LSCM simulation modules. Grey arrows
represent direction of photon propagation.
Optical configurations.
(A) TIRFM simulation module. (B) LSCM simulation modules. Grey arrows
represent direction of photon propagation.The three dimensional point spreading function (3D-PSF) model plays a key role in
the bioimaing simulations [22]. Each point-like source of a fluorophore gives rise to a 3D-PSF
pattern in the image-forming systems. The normalization constant of the PSF is
usually considered as an user input parameter. However, the bioimaging
simulations requires the counting of the number of photons emitted from a single
fluorophore, and spatial PSF integration to be unity within infinite volume
region (). The PSF decays in an oscillatory manner
at tails along the radial and axial axes. Such damping characteristics hinders
the estimation of an exact or approximate form of the PSF normalization
constant. A wrong estimation can easily lead to the miscounting of the number of
photons, and provide a wrong intensity of the final images. Such problematic
normalization has not been well discussed in the literature. In addition,
optical aberrations can lead to a non-uniform distribution of the 3D-PSF. The
aberrations are deviations in an image that occur when photons from one point of
an object does not converge into a single point after propagating through an
optical system. They can be caused by artifacts that arise from the interaction
of photons with glass lenses. Using first order paraxial approximation, makers
of optical instruments typically correct the optical systems to compensate for
the optical aberrations.Assuming the first order paraxial approximation, and the spatial PSF integration
to be unity within a limited volume region (), we implement the TIRFM and LSCM
simulation modules. Step-by-step instructions are provided below. More details
are discussed in the supporting information (see S1 Text).
Simulation studies to estimate the errors that arise from the PSF normalization
and the optical aberrations are required for the future implementation.The TIRFM simulation module enables selective visualization of the
basal surface regions of the cell model. Incident beam photons of
the excitation wavelength (λ) can uniformly
illuminate the specimen. Evanescent field is generated along z-axes
as perpendicular to the total internal reflection surface, and
capable of exciting the fluorophores near the surface. The incident
photon flux density at the level of photon-counting unit is defined
by where
ϕ and are the incident beam flux
density (W/cm2) and single photon energy, respectively.
h and c are Planck constant
and a speed of light. A
is
the amplitude of the incident photon flux density.Because of the desperate timescales of the quantum transitions, we
simply assume that the fluorescence molecules subsequently emit
single photon of longer wavelength while they absorb one million
photons of excitation wavelength, and the cross-section of
photon-molecule interaction is roughly 10−14
cm2 [23]. No other physical processes is simulated. The
expected number of photons emitted from a single fluorophore is
defined by where
∣A
∣2,
σ, and δT are
the transmitted beam flux density, the cross-section, and detection
time. The detector is located in a specific direction. We expect to
observe the number of photons devided by an unit surface area of a
sphere (4π). The amplitude of the
transmitted beam flux density depends on the index refraction, and
the incident beam angle, amplitude, and polarization.When all the fluorophores in the cell model are imaged
simultaneously, the distribution of the emitted photon of longer
wavelengths that passed through the use of objective lens and
special filters is computed as the sum of the PSFs of all the
fluorophores. In particular, we use the Born-Wolf PSF model [22]. For an
optimal wavelength (λ′) of a
fluorophore, we estimated that 55% of the emitted photons that
passed through the Dichroic mirror and emission filter survive
(). The expected image plane at
the focal point (z =
z
0) is given by the convolution of the
PSF and written in the form of where N and
M are the total number of fluorophores, and
optical magnification, respectively. is the position of the
k-th fluorophore. is the position in an image
plane. The PSF is normalized within a ±1.0
μm range of radial and axial axes. In
addition, polarization of the evanescent field is non-isotropic,
which means that dipoles of different orientations are excited with
different probabilities per unit time. In order to accurately
simulate image-formation process, the polarized form of the PSF is
required for the future implementations.The emitted photons are finally detected by CMOS or EMCCD cameras,
and digitized as an image at a detection time. The readout process
can convert expected incident photon signals to digital signals
relies on camera specifications and camera operating conditions to
carry out the properties for final images. The observed image of the
cell model can be obtained using the Monte Carlo method in the
presence of systematic sources, including statistical fluctuations
in photon counting (photon shot noise), and camera specification and
camera operating conditions. Finally, photoelectron signals can be
linearly converted to digital signals. Unit conversions are given byThe LSCM simulation module can visualize focal regions of the cell
model. In general, laser beam propagation of excitation wavelength
can be approximated by assuming that the laser beam has an ideal
Gaussian beam profile. The incident beam flux of excitation
wavelength (λ) and continuously illuminates
specimen, and is focused into a confocal volume at a given scan time
and beam position. Incident photon flux is defined by where Φ and
are the incident beam flux (W)
and single photon energy. h and c
are Planck constant and speed of light, respectively.We also assume that the linear conversion of photon emission is by
10−6, and the cross-section of photon-molecule
interaction is roughly 10−14 cm2 [23]. No other
physical processes are simulated. For a given position and time, the
expected number of photons emitted from a single fluorophore is
defined by where
I(r,z),
σ, and δT are
the transmitted beam flux density, cross-section, and scan time per
pixel, respectively. The detector is located in a specific
direction. We expect to observe the number of photons divided by an
unit surface area of a sphere (4π). The
transmitted beam flux density depends on the incident photon flux,
and the beam waist radius at the focal plane where the wavefront is
assumed to be flat.When all the fluorophores in the cell model are imaged
simultaneously, the distribution of the emitted photon of longer
wavelengths that passed through the use of objective lens and
pinhole is computed as the sum of the PSFs of all the fluorophore.
In particular, we use the Born-Wolf PSF model [22]. As an incident beam is
scanned across the cell model in horizontal and vertical axes, a
digital image is generated at a time. For a given scan time and beam
central position, the expected image plane at the focal point
(z = z
0) is given
by the integration of the image plane obtained from the PSF
convolution. It is written in the form of
where N,
R and M are the total number
of fluorophores, pinhole radius, and optical magnification,
respectively. is the position of the
k-th fluorophore. is the position of beam center.
is position in the image plane.
The PSF is normalized within a ±1.0
μm range of radial and axial axes.The emitted photons are finally detected by a photomultipliers tube
(PMT), and digitized as an image at a given scan time. The observed
image of the cell model can be obtained using the Monte Carlo method
in the presence of systematic sources, including statistical
fluctuations in photon counting (photon shot noise), and PMT
specifications and PMT operating conditions. Finally, photoelectron
signals can be linearly converted to digital signals. Unit
conversions are given by
Results
Comparison of in vitro images
We evaluated the performance of our simulation modules by comparing the simulated
images with the actual photographed ones for simple particle models of
fluorescent molecules. We simulated imaging of the focal region of those simple
models for the optical system with the detector specifications and detector
operating (see S2 Text). Details of the in vitro comparison are
described in the supporting information (see S2 Text).
The results are shown in Figs 3, 4 and 5. The intensity of the
simulated images corresponds to the number of photons detected in the digital
cameras or the PMT. Each simulated image is visually similar to the
corresponding real ones. Thus, the simulated images were compared with images
obtained using actual microscopy systems at the level of photon-counting units.
However, differences still remain in the resulting images owing to calibration.
Calibration is the process of improving the agreement of the code calculation
with a chosen set of benchmarks through the adjustment of the parameters
implemented in the simulation modules [6-8]. Such a calibration process is required
in all experiments to improve the agreement of the simulated outputs with the
in vitro data sets. Even though the results of a simple
calibration were used, the first version of our simulation modules was capable
of generating images that closely reproduce images obtained with an actual
microscopy system. A more elaborate set of calibration is required in the
future. More details are described below.
Fig 3
Simple models (A) 100 stationary HaloTag-TMR molecules are
distributed on a glass surface. (B) 19,656 HaloTag-TMR molecules are
distributed in a 30 × 30 × 6
μm3 box of aqueous solution (= 5
nM), and rapidly diffuse at 100
μm2/sec.
Fig 4
Using HaloTag-TMR molecules distributed on a glass surface to
evaluate the performance of TIRFM simulation module.
(A) Expected images of the simple particle model at various beam flux
densities (20,30,40 and 50 W/cm2). The expected images are
obtained by averaging 100 images over 3 sec exposure period. Intensity
histograms are also shown below each expected images and presented with
black-colored bars. Each histograms are logarithmically scaled and
presented with grey-colored bars. (B) Simulated digital images of the
simple particle model are shown at various beam flux densities (20,30,40
and 50 W/cm2). Size of each images is 152 × 156 pixel.
Orange scalebar represents 3.15 μm. Intensity
histograms are also shown below each simulated images. (C) Real captured
images obtained from in vitro experiment are shown at
various beam flux densities (20,30,40 and 50 W/cm2). The
maximum value of the grayscale is adjusted to improve visualization of
each image. Intensity histograms are also shown below each actual
images.
Fig 5
Using HaloTag-TMR molecules to evaluate the performance of LSCM
simulation modules.
(A) Expected images of the simple particle model at various beam flux
(5,10,30,50, and 100 μW). Each expected images
are generated by averaging 30 images over 30 sec exposure period.
Intensity histograms are also shown below each expected images and
presented with black-colored bars. Each histograms are logarithmically
scaled and presented with grey-colored bars. (B) Simulated digital
images of the simple particle model are shown for various beam flux
(5,10,30,50, and 100 μW). Size of each images is
100 × 100 pixel. Orange scalebar represents 5.18 μm.
Intensity histograms are also shown below each simulated images. (C)
Real captured images obtained from in vitro experiment
are shown for various beam flux (5,10,30,50, and 100
μW). Size of each images is 100 × 100
pixel. The maximum value of the grayscale is adjusted to improve
visualization of each image. Intensity histograms are also shown below
each actual images.
Using HaloTag-TMR molecules distributed on a glass surface to
evaluate the performance of TIRFM simulation module.
(A) Expected images of the simple particle model at various beam flux
densities (20,30,40 and 50 W/cm2). The expected images are
obtained by averaging 100 images over 3 sec exposure period. Intensity
histograms are also shown below each expected images and presented with
black-colored bars. Each histograms are logarithmically scaled and
presented with grey-colored bars. (B) Simulated digital images of the
simple particle model are shown at various beam flux densities (20,30,40
and 50 W/cm2). Size of each images is 152 × 156 pixel.
Orange scalebar represents 3.15 μm. Intensity
histograms are also shown below each simulated images. (C) Real captured
images obtained from in vitro experiment are shown at
various beam flux densities (20,30,40 and 50 W/cm2). The
maximum value of the grayscale is adjusted to improve visualization of
each image. Intensity histograms are also shown below each actual
images.
Using HaloTag-TMR molecules to evaluate the performance of LSCM
simulation modules.
(A) Expected images of the simple particle model at various beam flux
(5,10,30,50, and 100 μW). Each expected images
are generated by averaging 30 images over 30 sec exposure period.
Intensity histograms are also shown below each expected images and
presented with black-colored bars. Each histograms are logarithmically
scaled and presented with grey-colored bars. (B) Simulated digital
images of the simple particle model are shown for various beam flux
(5,10,30,50, and 100 μW). Size of each images is
100 × 100 pixel. Orange scalebar represents 5.18 μm.
Intensity histograms are also shown below each simulated images. (C)
Real captured images obtained from in vitro experiment
are shown for various beam flux (5,10,30,50, and 100
μW). Size of each images is 100 × 100
pixel. The maximum value of the grayscale is adjusted to improve
visualization of each image. Intensity histograms are also shown below
each actual images.To test the performance of the TIRFM simulation module, we
constructed a simple particle model of 100 stationary
HaloTag-with-tetramethylrhodamine (HaloTag-TMR) molecules
distributed on a glass surface, as shown in Fig 3A. We simulated imaging of
the basal region of the simple model for the TIRFM specifications
and TIRFM operating conditions (see S2
Text). Fig
4A shows the expected optical distribution used for the
simulation, which was generated by averaging 100 images over a 3 sec
exposure period. Intensity histograms of each expected images are
also shown in Fig
4A. Fig 4B and 4C show the simulated images and the real captured ones
at various beam flux densities. The intensity of the simulated
images corresponded to the number of photons detected in the EMCCD
camera. Increasing the beam flux density results in a relatively
brighter image. Each simulated image is visually similar to the
corresponding real one. Thus, the simulated images were compared
with the images obtained using the actual TIRFM systems at the level
of photon-counting units. However, differences still remain in the
resulting images owning to calibration. A more elaborate set of
calibrations is required in the future.To test the performance of the LSCM simulation module, we constructed
a simple particle model of 19,656 HaloTag-TMR molecules diffused in
an aqueous solution as shown in Fig 3B. We simulated imaging of
the middle region of the simple model for the LSCM specifications
and LSCM operating conditions (see S2
Text). Fig
4A shows the expected optical distribution used for the
simulation, which was obtained by averaging 30 images over a 30 sec
exposure period. Intensity histograms of each expected images are
also shown in Fig
4A. Fig 5B and 5C show the simulated images and the real captured ones
at various beam fluxes. The intensity of the simulated images
corresponds to the number of photon pulses detected in the PMT.
Increasing the beam flux results in relatively brighter image. Each
simulated image is visually similar to the corresponding real ones.
Thus, the simulated images were compared with the images obtained
using the actual LSCM systems at the level of photon-counting units.
However, differences still remain in the resulting images owning to
calibration. A more elaborate set of calibrations is required in the
future.
Comparison of in vivo images
Using the LSCM simulation module, we compared a more complex cell model with real
cell images obtained using the actual LSCM system. We constructed the following
spatial cell models: (i) the ERK nuclear translocation model for the EGF
signaling pathway, and (ii) the self-organizing wave model of PTEN for the
chemotactic pathway of D. discoideum. We developed these cell
models, which are not available in the literature. We assumed that the
parameters of each cell model and the LSCM system are well evaluated with
in vitro data sets. We then simulated imaging of the focal
region of those cell models for the LSCM specifications and LSCM operating
conditions (see S3 Text). Details of the in vivo comparison are
described in the supporting information (see S3 Text).
The results are shown in Figs 6 and 7. The
intensity of the simulated images corresponds to the number of photon pulses
detected in the PMT. Thus, the simulated cell images were compared with the
images obtained by the actual microscopy systems at the level of photon-counting
units. Significant new insight on the cell models will be published in the
future.
Fig 6
ERK nuclear translocation model of EGF signaling pathway.
(A) Reaction network. (B) Geometry of PC-12 cell model. A hemispherical
cell measuring 20 μm in diameter and 7
μm in height is assumed. (C) Time-lapse
images of the ERK nuclear translocation model observed using the LSCM
simulation module. Size of each images is 90 × 90 pixel. Orange
scalebar represents 4.66 μm. (D) The time-lapse
images obtained from the experiment. The grayscale of each images is
adjusted in the range of 0 to 225.
Fig 7
Self-organizing wave model of PTEN for the chemotactic pathway of
D. discoideum.
(A) Reaction network. (B) Geometry of D. discoiduem cell
model. A hemispherical cell measuring 25 μm in
diameter and 5 μm in height is assumed. (C)
Time-lapse image of the self-organizing wave model observed using the
LSCM simulation module. Size of each images is 52 × 51 pixel.
Orange scalebar represents 5.39 μm. (D)
Time-lapse images obtained from the experiment. Red and green indicate
PTEN-TMR and PH-EGFP, respectively. The colorscale of each images is
adjusted in the range of 0 to 225.
We constructed the cell model of ERK nuclear translocation for the
EGF signaling pathway. We assumed the PC12 cell model that
represents the ERK molecules tagged with the enhanced green
fluorescent protein (ERK-mEGFP). Fig 6A and 6B show the main reaction network
and the geometry of the model, respectively. The cell was placed on
the glass surface, and was nearly hemispherical. The size of the
hemispherical cell was estimated by experimentalists. A cell
measuring 20 μm in diameter and 7
μm in height was assumed. The model
consisted of 75 chemical species, 143 reactions, and 85 kinetic
parameters. A maximum of 100,000 ERK molecules were distributed in
the cell cytoplasm and rapidly diffuse at 1.00
μm2/sec. The input of the EGF
ligand could drive the transport of 30% of the ERK molecules into
the nucleus and back to the initial condition in 10 min. We
simulated imaging of the middle regions of the cell model for the
LSCM specifications and LSCM operating conditions (see S3
Text). Fig 6C and 6D show the simulated cell images and the cell images
obtained using the actual LSCM system. The intensity of the
simulated images corresponds to the number of photon pulses detected
in the PMT. Therefore, the simulated images were compared with
images obtained using the actual LSCM system at the level of
photon-counting units (see S1 Video). Each simulated image
was visually similar to the corresponding real one, but differences
still remain in the resulting images owning to calibrations. A more
elaborate set of calibration is required in the future.We also constructed a self-organizing wave model of PTEN for the
chemotactic pathway of D. discoideum to validate
the performance of two-color imaging for the LSCM simulation module.
We assumed a D. discoideum cell model that
expresses the fluorescently labeled pleckstrin homology domain of
Akt/PKB (PH) and PTEN, where PH and PTEN are indicators for
phosphorylates phosphatidylinositol 3,4,5-trisphosphate (PIP3)
metabolism. PH can bind to PIP3 at the membrane, whereas PTEN
catalyzes the degradation of PIP3 and has a binding motif for
phosphatidylinositol 4,5-biphosphate (PIP2). PH was tagged with EGFP
(PH-EGFP), whereas PTEN was tagged with HaloTag with TMR (PTEN-TMR).
A maximum of 10,000 molecules of PTEN-TMR and PH-EGFP were
homogeneously distributed in the cell cytoplasm. On the membrane,
PI3K catalyzed PIP2 phosphorylation to PIP3, whereas PTEN
dephosphorylated PIP3 into PIP2. Cytosolic PTEN was recruited to the
membrane regions containing PIP2. Nonetheless, PIP3 could dislodge
PTEN from PIP2 into the cytosol when they came in contact with each
other. This last reaction acted as a positive feedback for PIP3
accumulation. Fig 7A and 7B show the main reaction network and the geometry of
the model, respectively. A cell was placed on the glass surface, and
was nearly hemispherical. The size of the hemispherical cell was
estimated by experimentalists. The cell measuring 25
μm in diameter and 5
μm in height was assumed. The model
involved 8 chemical species, 12 reactions, and 12 kinetic
parameters. Lattice-based particle simulation of the cell model
enabled of the reproduction of the local oscillatory dynamics of
PTEN-TMR and PH-EGFP. We simulated imaging of the middle region of
the cell model for the LSCM specifications and LSCM operating
conditions (see S3 Text). Fig 7C and 7D show the
simulated cell images and the cell images obtained by the actual
LSCM system. The intensity of the simulated images corresponds to
the number of photon pulses detected in the PMT. Therefore, the
simulated images were compared with the images obtained using the
actual LSCM system at the level of photon-counting units (see S2
Video). Each simulated image was visually similar to the
corresponding real one, but intensity differences still remained in
the resulting images. The number of PTEN-TMR and PH-EGFP in the wave
model are approximately 4,000 for each, but we expect more
(∼30,000) in the observed images. A more elaborate set of
calibration is required in the future.
ERK nuclear translocation model of EGF signaling pathway.
(A) Reaction network. (B) Geometry of PC-12 cell model. A hemispherical
cell measuring 20 μm in diameter and 7
μm in height is assumed. (C) Time-lapse
images of the ERK nuclear translocation model observed using the LSCM
simulation module. Size of each images is 90 × 90 pixel. Orange
scalebar represents 4.66 μm. (D) The time-lapse
images obtained from the experiment. The grayscale of each images is
adjusted in the range of 0 to 225.
Self-organizing wave model of PTEN for the chemotactic pathway of
D. discoideum.
(A) Reaction network. (B) Geometry of D. discoiduem cell
model. A hemispherical cell measuring 25 μm in
diameter and 5 μm in height is assumed. (C)
Time-lapse image of the self-organizing wave model observed using the
LSCM simulation module. Size of each images is 52 × 51 pixel.
Orange scalebar represents 5.39 μm. (D)
Time-lapse images obtained from the experiment. Red and green indicate
PTEN-TMR and PH-EGFP, respectively. The colorscale of each images is
adjusted in the range of 0 to 225.
Discussion
Measurements using bioimaging techniques are generally influenced by systematic
effects that arise from the stochastic nature of biological cells, the
photon-molecule interaction, and the optical configuration. Such systematic effects
are always present in all bioimaging systems and hinder the comparison between the
cell model and the real cell image. Combining optics and cell simulation
technologies, we proposed a computational framework for handling the parameters
embedded in the cell model and the optical principles governing the bioimaging
systems. The simulation using this framework generated digital images from cell
simulation results after accounting for the systematic effects. In particular, we
demonstrated that the simulated digital images are visually similar to the images
obtained using actual TIRFM and LSCM systems. Each pixel intensity corresponded to
the number of photon pulses detected in the camera or the PMT. Thus, the framework
streamlines the comparison at the level of photon-counting units. However, the image
comparison is insufficient to check the validity of the simulation modules.
Verification is the process of confirming the simulation modules are correctly
implemented with respect to conceptual description and analytical solutions [6-8]. During the verification process, the
simulation modules must be tested to find and estimate numerical errors in the
implementations. The simulation modules are designed to count the number of photons
that passed through the optical configurations. A wrong estimation of the numerical
errors that arise from the photon-counting principle can provide a wrong intensity
of the final images. For example, a wrong PSF normalization can miscount the number
of photons, and lead to wrong final images. Furthermore, the simulated images can be
also compared with a chosen set of experimental benchmarks defined in calibration
and validation parameter domains [6-8].
Systematic variance and covariance that arise from various different parameter
settings must be estimated to establish the validity of the simulation modules.
Analyses to quantify the systematic uncertainties are required for the future
implementation.One of the key challenges of transforming biology from a phenomenological science to
a predictive one is how to bridge the gap between a cell model and an actual
biological cell [24-28].
Over the last two decades, large-scale, accurate, and comprehensive simulations of
cell models have greatly improved our understanding of many cellular networks and
processes [29-31]. However, we are still far
away from having predictive cell models for actual applications in medicine and
biotechnology. In this work, we focused on the “comparison” part of
the model validation and demonstrated the single cell-to-cell image comparison at
the level of photon-counting units. For future implementation, it is important to
fully simulate optical systems and to demonstrate other important parts of the model
validation [6-8]. Within this framework, the
functionality and capability of the cell models will be more easily seen and
understood. Future tasks required for the model validation include studying
diversity in cell populations and obtaining the nominal and predicted probability
distributions of the cell model. The behavior of individual cells depends on the
internal variables and the environmental conditions. The nominal and predicted
probability distributions of those variables are characterized by their statistical
quantities. A likelihood that quantifies the discrepancy between the predicted
distribution and the observed one can be evaluated by using a statistical test of
significance. If the result of the statistical test satisfies a certain confidence
level, then the cell model is either rejected or accepted with respect to real cell
images. Consequently, such model fitting will support discoveries in biological
science.Bioimaging simulation using the computational framework presented here is not meant
to replace biological experiments. It provides a realistic estimate of the output
that would be obtained in specific biological applications. Biologists often use
commercial bioimaging systems for their own biological interests. Optical properties
of biological molecules and/or phenomena uniquely change, according to the
experimenter’s skills and experiences in handling biological samples and
optical equipments. The commercial systems are designed for general usage, and are
not optimized to measure the optical properties of all biological samples. Although
some biologists assemble specialized optical imaging systems for a particular
application, it is still difficult for them to adjust systems parameters without
expected outputs. Such an approach is quite inefficient since it depends on the
experimenter’s skills and experiences. A more systematic approach is required
to reduce or eliminate unintended experimenter’s bias. In order to
objectively handle biological and physical principles in an organized manner, it is
important to develop an object-oriented simulation toolkit of biological imaging.
The simulation toolkit is constructed on the basis of a set of numerous biological
and physical processes to handle diverse interactions of photons with molecules over
a wide energy range. The toolkit provides a complete set of software components for
all area of bioimaging simulations: optical configuration, spatial cell models, run,
parameter management, visualization and user interface. Such a multi-disciplinary
nature of the toolkit allows a user to easily design, customize and extend
bioimaging and/or experimental systems well optimized for specific biological
applications. For example, the computational framework can also be applied to
simulate other bioimaging techniques including fluorescence recovery after
photobleaching (FRAP), fluorescence correlation spectroscopy (FCS), Forster
resonance energy transfer (FRET) and localization microscopy. All simulation modules
can be objectively handled in a uniform software platform.However, there are two problems in constructing such a software platform. (1)
Computational speed is not well optimized for the TIRFM and LSCM simulation modules.
The speed of generating a simulated image is proportional to the number of
fluorophores embbeded in a cell models. Bioimaging simulation of a cell model
containing 100,000 fluorophores, requires about one day to obtain the final image.
Optimization is required in the near future. (2) The optical properties of many
commercial materials are not publicly available. In particular, information on the
objective lens used is important for predicting an exact PSFs in a wide field. A
question is how we can overcome such nonscience-related problems (probably, it is a
matter of business model). In conventional approaches to biological research,
biologists and optical physicists work independently, and do not interact much
technologically. In order to properly design and customize the bioimaging and
experimental systems well optimized for the specific biological applications,
collaborative work with optical physicists and engineers will be required for the
future biological research. Clearly, the bioimaging simulation toolkit allows us to
better communicate with optical physicists and engineers, and to perform the
simulation studies of bioimaging systems and their operating conditions. Optical
materials are well designed by optical physicists and engineers, and their
performance is generally validated by simulation studies of physical principles and
their boundary conditions. Simulation studies are essential for the objective
examination of the response of the optical equipments. However, such simulation
studies have not been well performed for biological samples. Without the results of
simulation studies for biological samples, the collaboration could easily fail.
Then, information on the optical materials could not be shared. Using whatever form
of PSF as realistically as possible, it is important to estimate experimental
accuracy and precision for valuable discussion. We believe that the simulation
toolkit can bridge the gap between biology and optics.
Implementation details.
(PDF)Click here for additional data file.
Details of in vitro image comparison.
(PDF)Click here for additional data file.
Details of in vivo image comparison.
(PDF)Click here for additional data file.
Parameterization for ERK nuclear translocation model.
(PDF)Click here for additional data file.
Parameterization for self-organizing wave model of PTEN.
(PDF)Click here for additional data file.
In vivo image comparison movie for ERK nuclear
translocation model.
(MOV)Click here for additional data file.
In vivo image comparison movie for self-organizing wave
model of PTEN.
Authors: Jonathan R Karr; Jayodita C Sanghvi; Derek N Macklin; Miriam V Gutschow; Jared M Jacobs; Benjamin Bolival; Nacyra Assad-Garcia; John I Glass; Markus W Covert Journal: Cell Date: 2012-07-20 Impact factor: 41.582
Authors: Marija Cvijovic; Joachim Almquist; Jonas Hagmar; Stefan Hohmann; Hans-Michael Kaltenbach; Edda Klipp; Marcus Krantz; Pedro Mendes; Sven Nelander; Jens Nielsen; Andrea Pagnani; Natasa Przulj; Andreas Raue; Jörg Stelling; Szymon Stoma; Frank Tobin; Judith A H Wodke; Riccardo Zecchina; Mats Jirstrand Journal: Mol Genet Genomics Date: 2014-04-13 Impact factor: 3.291
Authors: Fang Huang; Tobias M P Hartwich; Felix E Rivera-Molina; Yu Lin; Whitney C Duim; Jane J Long; Pradeep D Uchil; Jordan R Myers; Michelle A Baird; Walther Mothes; Michael W Davidson; Derek Toomre; Joerg Bewersdorf Journal: Nat Methods Date: 2013-05-26 Impact factor: 28.547
Authors: Jayodita C Sanghvi; Sergi Regot; Silvia Carrasco; Jonathan R Karr; Miriam V Gutschow; Benjamin Bolival; Markus W Covert Journal: Nat Methods Date: 2013-11-03 Impact factor: 28.547
Authors: John Graf; Sanghee Cho; Elizabeth McDonough; Alex Corwin; Anup Sood; Andreas Lindner; Manuela Salvucci; Xanthi Stachtea; Sandra Van Schaeybroeck; Philip D Dunne; Pierre Laurent-Puig; Daniel Longley; Jochen H M Prehn; Fiona Ginty Journal: Bioinformatics Date: 2021-10-02 Impact factor: 6.937