Abstract

BACKGROUND: Flow diversion is an established technique for treatment of cerebral aneurysms. The Pipeline embolization device (PED) is the only FDA-approved flow diverting stent in the USA. A second-generation device, PED Flex, has recently been released with modifications to the delivery system. Published reports of experience with this new device are limited.
OBJECTIVE: To describe the initial outcomes from the first North American series using the PED Flex-a single-center experience of 44 cases.
METHODS: All patients consecutively treated with the PED Flex embolization device from February 2015 through April 2015 were included in the study. Data were collected for patient demographics, aneurysm characteristics, technical procedural details, and early outcomes.
RESULTS: PED Flex treatment was attempted in 42 patients (mean 56.6±2.0 years) with 44 aneurysms (mean size 6.5±0.6 mm), 41/44 (93%) of which were anterior circulation and 3/44 (7%) were posterior circulation. PED Flex was successfully implanted in 43/44 cases (98%). A single device was used in 41/43 cases (95%), with a mean of 1.07±0.05 devices implanted per case. Resheathing was performed in 4/44 cases (9%). Mean postprocedure hospital length of stay was 1.3±0.2 days. One significant neurological complication (2.3%) occurred, which was a stroke in a patient non-compliant with the prescribed antiplatelet regimen.
CONCLUSIONS: Pipeline Flex is a second-generation flow diverter with enhanced features compared with the first-generation PED. These modifications allow for more reliable deployment with continued improvements in procedural safety. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

Entities:  

Keywords:  Aneurysm; Device; Flow Diverter; Intervention; Stent

Mesh:

Year:  2015        PMID: 26136500     DOI: 10.1136/neurintsurg-2015-011894

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  24 in total

1.  Utilization of a Novel, Multi-Durometer Intracranial Distal Access Catheter: Nuances and Experience in 110 Consecutive Cases of Aneurysm Flow Diversion.

Authors:  Geoffrey P Colby; Li-Mei Lin; Risheng Xu; Narlin Beaty; Matthew T Bender; Bowen Jiang; Judy Huang; Rafael J Tamargo; Alexander L Coon
Journal:  Interv Neurol       Date:  2017-02-03

2.  Flow Diversion for Intracranial Aneurysm Management: A New Standard of Care.

Authors:  Fawaz Al-Mufti; Krishna Amuluru; Chirag D Gandhi; Charles J Prestigiacomo
Journal:  Neurotherapeutics       Date:  2016-07       Impact factor: 7.620

3.  First US experience with Pipeline Flex with Shield Technology using aspirin as antiplatelet monotherapy.

Authors:  Ricardo A Hanel; Pedro Aguilar-Salinas; Leonardo Bc Brasiliense; Eric Sauvageau
Journal:  BMJ Case Rep       Date:  2017-05-04

4.  Treatment of Intracranial Aneurysms With Pipeline Embolization Device: Newer Applications and Technical Advances.

Authors:  Santosh B Murthy; Jharna Shah; Halinder S Mangat; Philip Stieg
Journal:  Curr Treat Options Neurol       Date:  2016-04       Impact factor: 3.598

Review 5.  On Flow Diversion: The Changing Landscape of Intracerebral Aneurysm Management.

Authors:  A A Dmytriw; K Phan; J M Moore; V M Pereira; T Krings; A J Thomas
Journal:  AJNR Am J Neuroradiol       Date:  2019-03-20       Impact factor: 3.825

6.  Treatment of Intra- and Extracranial Aneurysms Using the Flow-Redirection Endoluminal Device: Multicenter Experience and Follow-Up Results.

Authors:  F Drescher; W Weber; A Berlis; S Rohde; A Carolus; S Fischer
Journal:  AJNR Am J Neuroradiol       Date:  2016-11-03       Impact factor: 3.825

7.  Outcome Study of the Pipeline Embolization Device with Shield Technology in Unruptured Aneurysms (PEDSU).

Authors:  D Atasoy; N Kandasamy; J Hart; J Lynch; S-H Yang; D Walsh; C Tolias; T C Booth
Journal:  AJNR Am J Neuroradiol       Date:  2019-11-14       Impact factor: 3.825

Review 8.  Predictive modeling and in vivo assessment of cerebral blood flow in the management of complex cerebral aneurysms.

Authors:  Brian P Walcott; Clemens Reinshagen; Christopher J Stapleton; Omar Choudhri; Vitaliy Rayz; David Saloner; Michael T Lawton
Journal:  J Cereb Blood Flow Metab       Date:  2016-03-23       Impact factor: 6.200

9.  Clinical and angiographic outcomes in patients with intracranial aneurysms treated with the pipeline embolization device: intra-procedural technical difficulties, major morbidity, and neurological mortality decrease significantly with increased operator experience in device deployment and patient management.

Authors:  Josser E Delgado Almandoz; Yasha Kayan; Andrea Tenreiro; Adam N Wallace; Jill M Scholz; Jennifer L Fease; Anna M Milner; Maximilian Mulder; Kyle M Uittenbogaard; Oscar Tenreiro-Picón
Journal:  Neuroradiology       Date:  2017-10-06       Impact factor: 2.804

10.  Unruptured giant intracavernous aneurysms untolerate internal carotid artery occlusion test: Untreated and treated with flow-diversion.

Authors:  Xianli Lv; Jianjun Yu; Ting Liao; Jin Wang; Guihuai Wang
Journal:  Neuroradiol J       Date:  2020-01-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.