GMctdospp: Description and validation of a CT dose calculation system.

Ralph Schmidt1, Jörg Wulff1, Klemens Zink2.   

Abstract

PURPOSE: To develop a Monte Carlo (MC)-based computed tomography (CT) dose estimation method with a graphical user interface with options to define almost arbitrary simulation scenarios, to make calculations sufficiently fast for comfortable handling, and to make the software free of charge for general availability to the scientific community.
METHODS: A framework called GMctdospp was developed to calculate phantom and patient doses with the MC method based on the EGSnrc system. A CT scanner was modeled for testing and was adapted to half-value layer, beam-shaping filter, z-profile, and tube-current modulation (TCM). To validate the implemented variance reduction techniques, depth-dose and cross-profile calculations of a static beam were compared against DOSXYZnrc/EGSnrc. Measurements for beam energies of 80 and 120 kVp at several positions of a CT dose-index (CTDI) standard phantom were compared against calculations of the created CT model. Finally, the efficiency of the adapted code was benchmarked against EGSnrc defaults.
RESULTS: The CT scanner could be modeled accurately. The developed TCM scheme was confirmed by the dose measurement. A comparison of calculations to DOSXYZnrc showed no systematic differences. Measurements in a CTDI phantom could be reproduced within 2% average, with a maximal difference of about 6%. Efficiency improvements of about six orders of magnitude were observed for larger organ structures of a chest-examination protocol in a voxelized phantom. In these cases, simulations took 25 s to achieve a statistical uncertainty of ∼0.5%.
CONCLUSIONS: A fast dose-calculation system for phantoms and patients in a CT examination was developed, successfully validated, and benchmarked. Influences of scan protocols, protection method, and other issues can be easily examined with the developed framework.

Entities:  

Mesh:

Year:  2015        PMID: 26133624     DOI: 10.1118/1.4922391

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Determination and verification of the x-ray spectrum of a CT scanner.

Authors:  Ahmad Ibrahim Hassan; Martin Skalej; Helmut Schlattl; Christoph Hoeschen
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-07

2.  Comparison of Diagnostic Accuracy of Radiation Dose-Equivalent Radiography, Multidetector Computed Tomography and Cone Beam Computed Tomography for Fractures of Adult Cadaveric Wrists.

Authors:  Jakob Neubauer; Matthias Benndorf; Carolin Reidelbach; Tobias Krauß; Florian Lampert; Horst Zajonc; Elmar Kotter; Mathias Langer; Martin Fiebich; Sebastian M Goerke
Journal:  PLoS One       Date:  2016-10-27       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.