Literature DB >> 26118175

National Survey on Internal Quality Control for HbA(1c) Analytical Instruments in 331 Hospital Laboratories of China.

Rong Zeng, Wei Wang, Haijian Zhao, Yang Fei, Zhiguo Wang.   

Abstract

BACKGROUND: The narrow gap of HbA1 value of mass fraction between "normal" (< 6.0%) and "diabetes" (≥ 6.5%) necessitates tight control of inter-assay standardization, assay precision, and trueness. This survey was initiated to obtain knowledge of the current situation of internal quality control (IQC) practice for HbA(1c) in China and find out the most appropriate quality specifications.
METHODS: Data of IQC for HbA(1c) in 331 institutions participating in the national proficiency testing (PT) programs in China were evaluated using four levels of quality specifications, and the percentages of laboratories meeting the quality requirement were calculated to find out the most appropriate quality specifications for control materials of HbA(1c) in China.
RESULTS: The IQC data varied vastly among 331 clinical laboratories in China. The measurement of control materials covered a wide range from 4.52% to 12.24% (inter-quartile range) and there were significant differences among the CVs of different methods, including LPLC, CE-HPLC, AC-HPLC, immunoturbidimetry, and others. Among the four main methods, CE-HPLC and AC-HPLC achieved a better precision.
CONCLUSIONS: As we can see, the performance of laboratories for HbA(1c) has yet to be improved. Clinical laboratories in China should improve their performance with a stricter imprecision criteria.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26118175     DOI: 10.7754/clin.lab.2014.141123

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  4 in total

1.  National continuous surveys on internal quality control for HbA1c in 306 clinical laboratories of China from 2012 to 2016: Continual improvement.

Authors:  Tingting Li; Wei Wang; Haijian Zhao; Falin He; Kun Zhong; Shuai Yuan; Zhiguo Wang
Journal:  J Clin Lab Anal       Date:  2016-11-21       Impact factor: 2.352

2.  Status of internal quality control for thyroid hormones immunoassays from 2011 to 2016 in China.

Authors:  Shishi Zhang; Wei Wang; Haijian Zhao; Falin He; Kun Zhong; Shuai Yuan; Zhiguo Wang
Journal:  J Clin Lab Anal       Date:  2017-02-15       Impact factor: 2.352

3.  Internal quality control status for BNP and NT-proBNP in China from 2014 to 2017.

Authors:  Huizhen Sun; Wei Wang; Haijian Zhao; Chuanbao Zhang; Falin He; Kun Zhong; Shuai Yuan; Zhiguo Wang
Journal:  J Clin Lab Anal       Date:  2018-08-13       Impact factor: 2.352

4.  National survey on internal quality control for tumour markers in clinical laboratories in China.

Authors:  Wei Wang; Kun Zhong; Shuai Yuan; Falin He; Yuxuan Du; Zhehui Hu; Zhiguo Wang
Journal:  Biochem Med (Zagreb)       Date:  2018-04-15       Impact factor: 2.313

  4 in total