Eiji Kobayashi1, Mamoru Kakuda2, Yusuke Tanaka2, Akiko Morimoto2, Tomomi Egawa-Takata2, Shinya Matsuzaki2, Toshihiro Kimura2, Yutaka Ueda2, Kiyoshi Yoshino2, Kiyokazu Nakajima3, Tadashi Kimura2.
Abstract
AIM: Poor visualization of the operative field due to an obscured camera lens is a problem frequently encountered while performing laparoscopic surgery. Little has been published about the prevention of lens obstruction specifically due to a contaminated camera port used for laparoscopic surgery. The aim of our study is to develop a new device, the Endowiper, for cleaning the laparoscopic port. MATERIALS: The new cleaner for the port's valve is made from rolled gauze. To simulate a surgical environment in the laboratory, we have used pseudo-blood to smudge the port's valve.Entities:
Keywords: Camera port cleaner; Clear surgical image; Hindrance during surgery; Laparoscopic surgery; Surgical device
Mesh:
Year: 2015 PMID: 26091981 PMCID: PMC4710664 DOI: 10.1007/s00464-015-4229-3
Source DB: PubMed Journal: Surg Endosc ISSN: 0930-2794 Impact factor: 4.584
Fig. 1A 12 mm Endowiper, B the device introduced into a 12-mm port, C 5 mm Endowiper, D the device introduced into a 5-mm port
Fig. 2Contaminated port valve by pseudo-blood
Fig. 3Port valve cleaning devices for 12 mm port (A) and 5 mm port (B). Image A shows an Endowiper, a sterile gauze tightly wrapped around an endoscopic dissecting cramp, a small sterile gauze grasped by an endoscopic dissecting cramp, and a sterile swab for a 12-mm port valve cleaner. Image B shows an Endowiper and a sterile swab for a 5-mm port valve cleaner
Fig. 4Camera images captured after cleaning the port valve. A Completely clear image; cleaning was judged effectively B blurry, obscured, or otherwise not completely clear image; cleaning was deemed ineffective
Clearance rate and trouble rate for four cleaning devices
| Characteristics | 5 mm trocar | 12 mm trocar | ||||||
|---|---|---|---|---|---|---|---|---|
| N | (%) | Clearance rate (%) | Trouble rate (%) | N | (%) | Clearance rate (%) | Trouble rate (%) | |
| All | 40 | 100 | 240 | 100 | ||||
| Type of device | ||||||||
| Endowiper | 20 | 50.0 | 85.0 | 0.0 | 60 | 25.0 | 83.3 | 0.0 |
| Gauze | N/A | N/A | N/A | N/A | 60 | 25.0 | 56.7 | 1.7 |
| Small gauze | N/A | N/A | N/A | N/A | 60 | 25.0 | 36.7 | 15.0 |
| Sterized swab | 20 | 50.0 | 20.0 | 30.0 | 60 | 25.0 | 40.0 | 90.0 |
|
|
|
|
|
| ||||
Odds ratios for clarity associated with type of device
| Variables | 5 mm trocar | 12 mm trocar | ||
|---|---|---|---|---|
| OR | (95 % CI) | OR | (95 % CI) | |
| Type of device | ||||
| Endowiper | 22.67 | (4.34–117.47) | 3.82 | (1.63–8.94) |
| Gauze | N/A | 1.00 | ||
| Small gauze | N/A | 0.51 | (0.25–1.05) | |
| Sterized swab | 1.00 | 0.44 | (0.21–0.92) | |
CI confidence interval, NA not applicable, OR odds ratio