| Literature DB >> 26077585 |
Ole Fröbert1, Fredrik Calais1, Stefan K James2, Bo Lagerqvist2.
Abstract
BACKGROUND: The clinical effect of thrombus aspiration in ST-elevation myocardial infarction may depend on the type of aspiration catheter and stenting technique. METHODS ANDEntities:
Keywords: angioplasty; myocardial infarction; stenting; thrombus aspiration
Mesh:
Year: 2015 PMID: 26077585 PMCID: PMC4599530 DOI: 10.1161/JAHA.114.001755
Source DB: PubMed Journal: J Am Heart Assoc ISSN: 2047-9980 Impact factor: 5.501
Baseline and Procedural Characteristics According to Aspiration Device
| Missing Information | Terumo Eliminate | Medtronic Export | Vascular Solutions Pronto | |
|---|---|---|---|---|
| N | 1748 | 1291 | 380 | |
| Age (y), mean (±SD) | (0) | 67.1 (11.6) | 66.4 (11.2) | 65.8 (11.6) |
| Female sex, n (%) | (0) | 452 (25.9) | 330 (23.5) | 98 (25.8) |
| Diabetes mellitus, n (%) | (0.5) | 229 (13.1) | 158 (12.2) | 41 (10.8) |
| Current smoker | (5.5) | 535 (30.6) | 386 (29.9) | 101 (26.6) |
| Previous myocardial infarction, n (%) | (1.3) | 197 (11.3) | 140 (10.8) | 42 (11.1) |
| Previous PCI, n (%) | (0) | 160 (9.2) | 125 (9.7) | 35 (9.2) |
| Previous coronary artery bypass grafting, n (%) | (0) | 38 (2.2) | 20 (1.5) | 8 (2.1) |
| On warfarin prior to PCI, n (%) | (0) | 31 (1.8) | 22 (1.7) | 5 (1.3) |
| Heparin prior to PCI, n (%) | (0) | 840 (48.1) | 397 (30.8) | 146 (38.4) |
| Symptom to PCI time (h), median (IQR) | (0) | 3.11 (2.08 to 5.67) | 3.08 (2.00 to 5.35) | 3.75 (2.09 to 6.39) |
| Diagnostic ECG to PCI time (h), median (IQR) | (5.8) | 1.12 (0.83 to 1.55 | 1.08 (0.77 to 1.55) | 1.15 (0.88 to 1.85) |
| Killip class ≥2, n (%) | (1.0) | 97 (5.5) | 75 (5.8) | 53 (13.9) |
| Radial access, n (%) | (0) | 1349 (77.2) | 750 (58.1) | 300 (78.9) |
| TIMI flow grade 0 or 1 before PCI, n (%) | (12.0) | 1382 (79.1) | 1019 (79.0) | 263 (71.9) |
| Bivalirudin during PCI, n (%) | (0) | 1503 (86.0) | 1012 (78.4) | 197 (51.8) |
| Glucoprotein IIb/IIIa inhibitor during PCI, n (%) | (0) | 215 (12.3) | 221 (17.1) | 135 (35.5) |
ECG indicates electrocardiogram; PCI, percutaneous coronary intervention; TIMI, thrombolysis in myocardial infarction.
Baseline characteristics that differ at P<0.05.
Figure 1A, Kaplan–Meier curves showing cumulative probability of all-cause mortality up to 1 year in patients randomized to thrombus aspiration. Outcomes associated with different thrombus aspiration catheters are shown. B, Cumulative probability of reinfarction until 1 year. C, Cumulative probability of stent thrombosis until 1 year.
Aspiration Device and Outcome at 1 Year
| N (%) | All-Cause Death (%) | Hospitalization for a New Myocardial Infarction (%) | Stent Thrombosis (%) | |||||
|---|---|---|---|---|---|---|---|---|
| Terumo Eliminate | 1748 (48.3) | 5.4 | 2.5 | 0.9 | ||||
| Medtronic Export | 1291 (35.7) | Unadjusted | 5.0 | HR 0.93 (0.68 to 1.28), | 2.6 | HR 1.07 (0.68 to 1.68), | 0.6 | HR 0.72 (0.30 to 1.69), |
| Adjusted with missing imputation | HR 0.96 (0.69 to 1.34), | HR 0.98 (0.62 to 1.56), | NA | |||||
| Vascular Solutions Pronto | 380 (10.5) | Unadjusted | 4.5 | HR 0.87 (0.49 to 1.39), | 3.2 | HR 1.27 (0.67 to 2.40), | 0.5 | HR 0.6 (0.14 to 2.63), |
| Adjusted with missing imputation | HR 1.04 (0.61 to 1.78), | HR 1.29 (0.66 to 2.53), | NA | |||||
The Terumo Eliminate catheter was used as reference for hazard ratios (HR) and 95% CI. NA indicates not applicable.
Figure 2A, Kaplan–Meier curves showing cumulative probability of all-cause mortality up to 1 year in patients treated with direct stenting and randomized to thrombus aspiration (TA) and percutaneous coronary intervention (PCI) or only PCI. B, Cumulative probability of reinfarction until 1 year. C, Cumulative probability of stent thrombosis until 1 year.
Figure 3A, Kaplan–Meier curves showing cumulative probability of all-cause mortality up to 1 year in patients stented with drug-eluting stents and randomized to thrombus aspiration (TA) and percutaneous coronary intervention (PCI) or only PCI. B, Cumulative probability of reinfarction until 1 year. C, Cumulative probability of stent thrombosis until 1 year.
Figure 4A, Kaplan–Meier curves showing cumulative probability of all-cause mortality to 1 year in patients treated with postdilatation and randomized to thrombus aspiration (TA) and percutaneous coronary intervention (PCI) or only PCI. B, Cumulative probability of reinfarction until 1 year. C, Cumulative probability of stent thrombosis until 1 year.