Manuela Kunz1, Steven M Devlin2, Mark B Hurtig3, Stephen D Waldman4, John F Rudan5, Davide D Bardana5, A James Stewart1. 1. School of Computing, Queen's University, Kingston, Ontario, Canada. 2. Department of Mechanical and Materials Engineering, Queen's University, Kingston, Ontario, Canada. 3. Ontario Veterinary College, University of Guelph, Guelph, Ontario, Canada. 4. Department of Mechanical and Materials Engineering, Queen's University, Kingston, Ontario, Canada ; Department of Chemical Engineering, Queen's University, Kingston, Ontario, Canada. 5. Department of Surgery, Queen's University, Kingston, Ontario, Canada.
Abstract
OBJECTIVE: Autologous osteochondral cartilage repair is a valuable reconstruction option for cartilage defects, but the accuracy to harvest and deliver osteochondral grafts remains problematic. We investigated whether image-guided methods (optically guided and template guided) can improve the outcome of these procedures. DESIGN: Fifteen sheep were operated to create traumatic chondral injuries in each knee. After 4 months, the chondral defect in one knee was repaired using (a) conventional approach, (b) optically guided method, or (c) template-guided method. For both image-guided groups, harvest and delivery sites were preoperatively planned using custom-made software. During optically guided surgery, instrument position and orientation were tracked and superimposed onto the surgical plan. For the template-guided group, plastic templates were manufactured to allow an exact fit between template and the joint anatomy. Cylindrical holes within the template guided surgical tools according to the plan. Three months postsurgery, both knees were harvested and computed tomography scans were used to compare the reconstructed versus the native pre-injury joint surfaces. For each repaired defect, macroscopic (International Cartilage Repair Society [ICRS]) and histological repair (ICRS II) scores were assessed. RESULTS: Three months after repair surgery, both image-guided surgical approaches resulted in significantly better histology scores compared with the conventional approach (improvement by 55%, P < 0.02). Interestingly, there were no significant differences found in cartilage surface reconstruction and macroscopic scores between the image-guided and the conventional surgeries.
OBJECTIVE:Autologous osteochondral cartilage repair is a valuable reconstruction option for cartilage defects, but the accuracy to harvest and deliver osteochondral grafts remains problematic. We investigated whether image-guided methods (optically guided and template guided) can improve the outcome of these procedures. DESIGN: Fifteen sheep were operated to create traumatic chondral injuries in each knee. After 4 months, the chondral defect in one knee was repaired using (a) conventional approach, (b) optically guided method, or (c) template-guided method. For both image-guided groups, harvest and delivery sites were preoperatively planned using custom-made software. During optically guided surgery, instrument position and orientation were tracked and superimposed onto the surgical plan. For the template-guided group, plastic templates were manufactured to allow an exact fit between template and the joint anatomy. Cylindrical holes within the template guided surgical tools according to the plan. Three months postsurgery, both knees were harvested and computed tomography scans were used to compare the reconstructed versus the native pre-injury joint surfaces. For each repaired defect, macroscopic (International Cartilage Repair Society [ICRS]) and histological repair (ICRS II) scores were assessed. RESULTS: Three months after repair surgery, both image-guided surgical approaches resulted in significantly better histology scores compared with the conventional approach (improvement by 55%, P < 0.02). Interestingly, there were no significant differences found in cartilage surface reconstruction and macroscopic scores between the image-guided and the conventional surgeries.
Chondral and osteochondral lesions in the articular surface of the knee are injuries frequently encountered in clinical practice.[1,2] Because of the inadequate healing response of cartilage,[3,4] defects of a critical size may lead to osteoarthritis if untreated.[5] Most commonly used treatment options for defects include microfracture,[6-10] autologous chrondocyte implantation (ACI),[11-14] and autologous osteochondral transplantation, also known as mosaic arthroplasty.[15-19]In a prospective randomized clinical study, Gudas et al.[20] found significantly better clinical outcomes and histology using autologous osteochondral transplantation compared with microfracture. The study found excellent or good postoperative results in 96% of the patients treated with autologous osteochondral transplantation compared with 52% of the patients treated by microfracture. Horas et al.[21] concluded, from a study of 40 patients, that autologous osteochondral transplantation resulted in a faster recovery than ACI. This result is in contrast to a study conducted by Bentley et al.,[22] who found excellent or good results in 88% of the ACI patients compared with 69% for autologous osteochondral transplantation patients. Dozin et al.,[23] on the other hand, found no difference between both techniques in a randomized trial. Recent reviews of randomized and controlled trials[24,25] to compare the effectiveness of different cartilage repair methods concluded that, at this point, there is insufficient evidence concerning the relative effectiveness of ACI or autologous osteochondral transplantation.Various studies have demonstrated the importance of creating a congruent, continuous joint surface using autologous osteochondral transplantation to optimize outcomes.[26-28] Donor sites accessibility and the variation in the radius of the femoral condyle curvature[29] make re-creation of a congruent joint surface challenging when using multiple small grafts. Sanders et al.[30] found in 2-week postoperative MRI evaluations of 21 autologous osteochondral transplantation patients that only one patient had surface congruency, whereas 16 patients had mild, 2 patients moderate, and 1 patient marked surface incongruency over the defect. This raises the question of whether improved intraoperative methods that help the surgeon to achieve higher accuracy in harvesting and delivery of grafts might improve the outcome of autologous osteochondral transplantation procedures. Koulalis et al.[31] compared the outcome of optoelectronically navigated procedures versus freehand autologous osteochondral transplantation procedures in three cadaveric knees and found improved accuracy in the navigated procedures for the perpendicularity of graft removal and placement, as well as for the depth of graft placement. A limitation of this study was the use of in vitro specimens: the authors could not evaluate whether this improved accuracy influenced the clinical outcome.Although the use of optoelectronic technology for image-guided knee applications has had gratifying results, this technology has some drawbacks: additional technical equipment (optoelectronic camera, PC) is required in the operating theater and an intraoperative registration process is required to find the correspondence between the image data and the patient. To overcome these drawbacks, recent research in the area of image-guided surgeries has used patient-specific templates.[31-34] The idea is to build custom surgical templates based on a three-dimensional reconstruction of the patient’s specific anatomical structures. In the trial presented here, we investigated the application of two image-guided systems for autologous osteochondral transplantation, one system using optoelectronic tracking and the other system using patient-specific templates, and compared the short-term clinical outcome for both systems with the conventional freehand method. The purpose of the pilot study presented here was to investigate whether image-guided methods can help improve the outcome of autologous osteochondral transplantation procedures.
Methods
Fifteen mature sheep were randomly assigned to one of three treatment groups. For each sheep, one treatment and one control knee was randomly assigned. All sheep underwent an initial computer tomography (CT) arthrogram (LightSpeed Plus, GE Healthcare, Waukesha, WI) in helical mode, with a slice thickness of 0.625 mm at 140 kpV, followed by a procedure to create a traumatically inducted cartilage defect in the medial condyle of both knees (cartilage defect surgery). During a second procedure 4 months later, one of three repair procedures was performed (cartilage repair surgery). Three months after the repair surgery, the sheep were euthanized. Both the treatment knee and the control knee were harvested and the outcomes were evaluated. This study was performed with approval from the University Animal Care Committee at Queen’s University and the University of Guelph.The cartilage defect surgery was performed through a 2-cm infrapatellar arthrotomy using spring-loaded impactor to create chronic chondral defects on anterior central weightbearing region of the medial femoral condyles in both knees of each sheep as per previous reports.[35] After routine closure of the arthrotomy, sheep were recovered and allowed exercise in large pens. These 4.5 to7 mm diameter injuries increased in size over 3 months, resulting in an irregularly shaped chondral lesion that was debrided to a minimum 7 mm diameter full thickness chondral defect in a second surgery 4 months later (7 mm is reported as the minimum defect dimension that sheep are incapable of repairing without intervention[36]). Reconstruction was performed with one of the three following techniques: (a) conventional freehand technique, (b) optically guided technique, and (c) template-guided technique. All surgeries were carried out by the same surgeon who was experienced in autologous osteochondral transplantations. The autologous osteochondral transplantation system from Smith and Nephew Endoscopy (Mosaicplasty, Andover, MA) was used for all surgeries. For all procedures, a medial parapatellar arthrotomy was performed and the patella was luxated laterally to expose the donor sites in the medial and lateral trochlear ridges as well as the medial femoral condyle recipient site.
Conventional Surgical Technique
The conventional osteochondral grafting technique was performed as described by Hangody and Kárpáti.[37] During the surgery, 4.5 mm osteochrondral grafts were harvested from the axial aspect of the medial trochlear ridge for transplantation into the medial condyle. The surgeon determined the location of donor and recipient site at the time of the surgery, optimizing the fit and congruency by eye.
Optically Guided Technique
The optically guided procedure consisted of preoperative planning and intraoperative guidance. Prior to the surgery, a CT arthrogram scan for the treatment knee was obtained. All scans were performed after the injection of an iodinated contrast material and were obtained with a LightSpeed Plus (GE Healthcare, Waukesha, WI) in helical mode, with a slice thickness of 0.625 mm at 140 kpV. Three-dimensional (3D) surface models for bone and cartilage were created using the commercial software package Amira (Visage Imaging Inc., Carlsbad, CA).Custom-made surgical planning software for osteochondral grafting was developed. The 3D surface models as well as the CT dataset were loaded into the software and displayed (). The operator created a surgical plan consisting of a set of osteochondral grafts (“plugs”) positioned over the defect site. The 3D position and orientation of each plug, as well as its shape (diameter, height, and surface slope), were chosen by the operator to best reconstruct the desired articular surface at the defect site ().
Figure 1.
Preoperative planning for image-guided procedures: (a) Planning of position and orientation of harvest and delivery site for virtual plug; (b) Creation of virtual cartilage/bone plugs.
Preoperative planning for image-guided procedures: (a) Planning of position and orientation of harvest and delivery site for virtual plug; (b) Creation of virtual cartilage/bone plugs.For each plug, a harvest location was chosen to best match the shape of the plug. The plugs could be rotated axially so that the sloped surface at the harvest site could be made to match the sloped surface at the defect site. The operator validated the surgical plan by superimposing the plugs on the 3D models and by superimposing the plugs on three orthogonal slices of the CT dataset ().A Polaris optoelectronic tracker (Northern Digital, Waterloo, Canada) was installed in the operating theatre () and a tracking sensor was rigidly attached to the femur (). Tracking sensors were attached to conventional harvest chisels and drill guides. A special retractable attachment was required for the harvest chisel because the heavy impacts made to the chisel would dislodge a conventionally attached sensor ().
Figure 2.
Optically guided surgical technique. (a) Installation of optoelectronic camera (i) and PC (ii) in the operating room; (b) Computer-guidance display for navigation of tools; (c) Attachment of optoelectronic sensors to femur; (d) Attachment of optoelectronic sensor to harvest chisel. The specially designed attachment consisted of two pieces: one fixed to the chisel and one free to move from the fixed piece in a translation parallel to the chisel axis. On a blow to the chisel, the latter piece would briefly move away from the fixed piece and then return to its original position as shown in the right image.
Optically guided surgical technique. (a) Installation of optoelectronic camera (i) and PC (ii) in the operating room; (b) Computer-guidance display for navigation of tools; (c) Attachment of optoelectronic sensors to femur; (d) Attachment of optoelectronic sensor to harvest chisel. The specially designed attachment consisted of two pieces: one fixed to the chisel and one free to move from the fixed piece in a translation parallel to the chisel axis. On a blow to the chisel, the latter piece would briefly move away from the fixed piece and then return to its original position as shown in the right image.A registration was made between the sheep femur and the 3D bone model of the femur using a combined pair-point and surface matching algorithm.[38] Using visual feedback from the computer-guidance system, the surgeon used a tracked pointing device to locate the planned harvest site of a plug and, using a sterile pen, marked an axial rotation reference on the cartilage surface of this plug. This mark allowed the surgeon to keep track of the rotation of the plug between harvesting and delivery. Using visual and numerical feedback on the display, the surgeon positioned and oriented the harvesting chisel on the cartilage according to the preoperative plan (). The surgeon then drove the chisel into the cartilage and bone until the guidance display indicated that the correct depth was reached. Then the graft was harvested.After each graft was harvested the surgeon positioned and aligned, in a similar manner, the tracked drill guide over the planned recipient site and the recipient hole was drilled. The depth of the hole was navigated using the conventional depth indicator at the drill bit. The harvested plug was inserted into the drill guide in such a way that the rotation mark of the plug was aligned with the calibrated up-direction of the drill guide. Using the visual feedback of the guidance system, the drill guide was then axially rotated until the planned rotational position of the graft was reached and the graft was carefully inserted into the recipient hole. This procedure was repeated for each planned graft.
Template-Guided Technique
The template-guided procedure consisted of preoperative planning, template construction, and intraoperative guidance. The surgery was planned identically to the optically guided procedure.A set of individualized templates was built for each knee, containing one “marking guide,” one “harvesting guide,” and one “delivery guide” for each planned plug. The underside of each template was shaped to exactly match part of the surface of the knee (), using the information from the prerepair CT arthrogram. By this means, the planned position of the template could be correctly reproduced intraoperatively by adjusting the position of the template until an exact fit with the cartilage surface was achieved. Each template was built out of thermo-plastic acrylonitrile butadiene styrene (ABS) on a rapid prototyping machine (dimension SST; Statasys Inc., Eden Prairie, MN).
Figure 3.
Patient-specific instrument guides for harvesting graft (a) and delivery of graft (b). (i) Mirror image of cartilage surface allowed precise fit of guide to the articular surface. (ii) Planned position and orientation of harvest and delivery graft. (iii) Guidance cylinder for instrument. (iv) Pin holes to fixate guide to knee. (v) Rotation mark. (vi) Slots to irrigate the hole during drilling. To navigate depth of instrument insertion, length of the guidance cylinder was defined based on planned graft height (c).
Patient-specific instrument guides for harvesting graft (a) and delivery of graft (b). (i) Mirror image of cartilage surface allowed precise fit of guide to the articular surface. (ii) Planned position and orientation of harvest and delivery graft. (iii) Guidance cylinder for instrument. (iv) Pin holes to fixate guide to knee. (v) Rotation mark. (vi) Slots to irrigate the hole during drilling. To navigate depth of instrument insertion, length of the guidance cylinder was defined based on planned graft height (c).The marking guide was designed to fit into the femoral patella groove and contained, for each plug, a hole at the planned harvest site of the plug. Each hole had on its circumference a small indicator bump; the surgeon would draw a radial line on the cartilage surface at the location of the indicator. The line allowed the axial rotation of the plug to be tracked.The harvesting guide was designed to fit into the femoral patella groove and contained, for each plug, a guidance cylinder for the harvesting chisel (). The height of each cylinder was chosen to stop the chisel after the chisel had been inserted to the planned depth ().Each delivery guide fit to the medial femoral condyle at the location of the defect and contained a single guidance cylinder. The conventional drill guide fit into the guidance cylinder to guide the drill bit during drilling and to guide the plug during delivery.Rotation marks at the guidance holes () ensured that the harvested plug was delivered with the correct rotational alignment with respect to the plug axis.After the conventional incision was made, the marking guide was positioned on the knee and a rotation reference mark was made for each plug. The harvesting guide was placed on the knee and fixated with two 2-mm Kirscher wires. Using the guidance cylinders, all plugs were harvested and stored in numbered containers (). For each plug, the length of the plug was verified using a conventional ruler. The harvesting guide was removed.
Figure 4.
Intraoperative use of patient-specific instrument guide. (Left) The harvesting guide was placed on the knee and harvest chisel inserted; (Right) The recipient guide was placed on the knee and the drill guide was inserted.
Intraoperative use of patient-specific instrument guide. (Left) The harvesting guide was placed on the knee and harvest chisel inserted; (Right) The recipient guide was placed on the knee and the drill guide was inserted.For each plug in sequence, one delivery guide was placed on the knee (). The delivery hole was drilled. The depth marking on the Mosaicplasty drill bit was used to determine the depth of the hole. Then the plug was inserted into the drill sleeve and the rotation mark on the plug aligned with a corresponding rotation mark at the guidance cylinder. Finally, the plug was pushed through the drill sleeve into the delivery hole and the delivery guide for that plug was removed.
Postoperative Assessments
All sheep were recovered from anesthesia and had restricted exercise in small pens for 3 weeks followed by unrestricted movement in larger pens for the 3-month recovery period. At the end of the study, the sheep were euthanized with an overdose of pentobarbital and the hind limbs harvested for assessments. CT arthrograms were repeated in the reconstructed joints. The joints were then dissected carefully and photodocumented. The following criteria were recorded from each joint:
Shape of articular surface reconstruction
Immediately after the surgery, the surgeon documented the result in surgical notes, describing the congruency of each plug to the surrounding surface at four points on the circumference of the plug.
Weight, pain, and lameness
After the surgery, each sheep was followed daily for 3 months. Weight, pain, and lameness were documented. Pain was graded on a scale of 1 to 3 as a combination of lameness, respiration, attitude, and appetite. Lameness was graded on a scale of 1 to 5, with 1 being “weight bearing but slight limp” and 5 being “not weight bearing, leg lifted or cannot get up.”
Macroscopic International Cartilage Repair Society (ICRS) score after healing
All sheep were euthanized 3 months postoperatively and both knees were harvested and dissected. The joints were photographed and examined macroscopically using the ICRS Macroscopic Score.[39] The scoring was done by one observer who was blinded to the treatment method used for repair.
Shape of articular surface after healing
Three-dimensional models for bone and for cartilage were created from CT images before injury and 3 months postreconstruction using the commercial software package Mimics (Materialise, Leuven, Belgium). Using the Iterative Closest Point algorithm,[40] the posthealing bone model was registered to the predefect bone model. The resulting transformation was applied to align the posthealing cartilage model with the pristine articular surface of the predefect scan. The root mean square (RMS) error between both surfaces over the defect was calculated.
Subchondral bone cyst formation after healing
After harvesting the treatment knee, a MicroCT (GE LOCUS Explore) with a voxel size of 0.095 mm3 was performed. Using the Mimics software, the cysts in the medial condyle were segmented and the volume of these cysts determined.
Histological measures after healing
Immediately after harvesting, imaging, and macroscopic evaluation of the knees, the following tissue samples were obtained for histological evaluation: synovial membrane intercondylar area, medial aspect; osteochondral blocks from the medial femoral condyle; the tibia plateau; and the medial trochlea. All samples were stored in formalin, decalcified in formic acid, and embedded in paraffin blocks from which 6-µm-thick sections cut. Sections were stained with hematoxylin and eosin (H&E) and safranin-O/fast green. Sections from the repair site were examined by two independent reviewers using the ICRS II histological scoring system consisting of 14 parameters.[41] This system is an integrated evaluation of tissue and cell morphology with emphasis on restoration of normal cartilage and subchondral bone plate architecture as well as integration of the grafts and intergraft repair tissue with the surrounding host tissue.
Statistical Analyses
Statistical analysis was performed using the software package Analyse-It (Analyse-It Software Ltd., Leeds, UK). A non-paired Student t test was used to evaluate significant differences between all three groups for parametric tests. For nonparametric score results, differences were evaluated by the Mann-Whitney U test. For all tests, P < 0.05 was considered statistically significant.
Results
For all 15 sheep, the cartilage defect and cartilage repair surgery was successfully performed. There was one case of superficial wound infection in the conventional group after the reconstructive surgery, which was treated successfully with antibiotics. For one sheep in the optically guided group, a mechanical lameness due to an intermittently luxated patella was diagnosed 1 week following the cartilage repair surgery. shows photographs for three knees (one from each group) during different steps of our study and evaluation.
Figure 5.
Examples for outcomes for conventional technique (top row), optically guided technique (middle row), and template-guided technique (bottom row). For each technique, four images demonstrate the state at different points in the study. The left column shows the defect before the repair surgery; the second column shows the area over the defect direct after the repair surgery; the third column shows the articular surface 3 months posthealing after harvesting of the knee; and the last column shows the stained histology image.
Examples for outcomes for conventional technique (top row), optically guided technique (middle row), and template-guided technique (bottom row). For each technique, four images demonstrate the state at different points in the study. The left column shows the defect before the repair surgery; the second column shows the area over the defect direct after the repair surgery; the third column shows the articular surface 3 months posthealing after harvesting of the knee; and the last column shows the stained histology image.
Shape of Articular Surface Reconstruction
shows the percentage of recessed and proud plug surface for all three groups, as determined from the intraoperative notes immediately after surgery. The percentage of proud surface was significantly smaller (P < 0.02) for the template-guided group (1.4 ± 3.1%) compared with the conventional group (31.2 ± 22.4%). The difference between the optically guided group (7.8 ± 7.8%) and the conventional group was not significant (P = 0.06), but the borderline P value suggested a trend toward smaller values for the optically guided group. The percentage of recessed surface had no significant differences between conventional (0.0 ± 0.0%), optically guided (27.1 ± 23.1%), and template-guided (6.2 ± 13.9%) groups.
Figure 6.
Results for surface congruency of articular surface postrepair. Values are displayed as average and 95% confidence interval.
Results for surface congruency of articular surface postrepair. Values are displayed as average and 95% confidence interval.
Weight, Pain, and Lameness
There was no significant difference in weight, pain, or lameness between the conventional group and the computer-assisted groups. But within the computer-assisted groups, the average duration of pain was significantly greater (P < 0.04) for the template-guided group (6.2 ± 2.6 days) compared with the optically guided group (2.4 ± 2.3 days). The conventional group pain duration was 4.0 ± 4.1 days.The intensity of the pain for the template-guided group (2.6 ± 1.4) was significantly greater (P < 0.035) than that of the optically guided group (1.0 ± 0.8). The conventional group pain intensity was 2.0 ± 2.1.
Macroscopic ICRS Score after Healing
No significant difference was found between any groups in the 3-month postrepair macroscopic evaluation ().
Table 1.
Three-Month Postoperative Macroscopic Scores (Mean ± Standard Deviation)
Three-Month Postoperative Macroscopic Scores (Mean ± Standard Deviation)ICRS = International Cartilage Repair Society.
Shape of Articular Surface after Healing
The articular surface over the defect in the posthealing CT scan was compared to the corresponding (pristine) articular surface in the predefect scan. We found an RMS error of 0.33 ± 0.10 mm for the conventional group, 0.44 ± 0.14 mm for the optically guided group, and 0.29 ± 0.25 mm for the template-guided group. No significant differences between the three groups were found.
Subchondral Bone Cyst Formation after Healing
shows the cyst volumes in the medial condyle 3 months posthealing. The cyst volume for the template-guided group (51 ± 47 mm3) was significant smaller (P < 0.02) than that of the conventional group (173 ± 76 mm3). No significant difference was found with the optically guided group (98 ± 144 mm3).
Figure 7.
Results for cysts volume in medial condyle 3 months postsurgery. Values are displayed as average and 95% confidence interval.
Results for cysts volume in medial condyle 3 months postsurgery. Values are displayed as average and 95% confidence interval.
Histological Measures after Healing
shows the histology scores for three areas (the medial condyle, the tibial plateau, and the surrounding tissue) and for the three groups. Error bars show the 95% confidence interval. For the medial condyle, the treatment effect was significant better (P < 0.02) in the optically guided group than in the conventional group. Also, the treatment effect in the template-guided group was significant better (P < 0.01) than in the conventional group. For the tibial condyle, the treatment effect was significantly better (P < 0.035) in the template-guided group than in the conventional group. No significant differences were found for the histology score of the surrounding tissue for all three groups. A significant linear correlation was found (linearity fit P < 0.004) between the ICRS II treatment effect for the medial condyle and the intraoperative estimated percentage of proud reconstructed articular surface for all 15 sheep ().
Figure 8.
Results for treatment effect using the ICRS II histology score 3 months postsurgery. Values are displayed as average and 95% confidence interval. ICRS = International Cartilage Repair Society.
Figure 9.
Linear correlation between the percentage of reconstructed articular surface that was proud and the ICRS II treatment effect for the medial condyle for all 15 sheep. ICRS = International Cartilage Repair Society.
Results for treatment effect using the ICRS II histology score 3 months postsurgery. Values are displayed as average and 95% confidence interval. ICRS = International Cartilage Repair Society.Linear correlation between the percentage of reconstructed articular surface that was proud and the ICRS II treatment effect for the medial condyle for all 15 sheep. ICRS = International Cartilage Repair Society.
Discussion
The primary finding was that both image-guided techniques had a significantly better treatment effect than did the conventional surgical technique.The template-guided technique resulted in significantly better surface congruency than the conventional technique. We also saw a trend toward better surface congruency using the optically guided method compared with the conventional technique. Interestingly, our results did not show a correlation between the articular surface congruency immediately after the operation and the macroscopic appearance of the cartilage at 3 months postoperatively. Similarly, no correlation was found between the articular surface congruency immediately after the operation and the surface congruency measured on the CT scan taken 3 months postoperatively. The lack of correlation suggests that in this time the proud plugs subsided and recessed plugs filled in. This is consistent with observations from other studies.[28]A secondary finding is the significant correlation between the articular surface congruency immediately after the operation and the histology of the cartilage 3 months postoperatively. The results showed that proud plugs are associated with poorer healing in the short term. The poorer healing could be the effect of peak pressure on these grafts. Koh et al.[27] observed in an in vitro study striking increase in peak pressures when the plug was proud. This increased pressure may cause overload in the tissue and may damage the cartilage. The reason that we observed this poorer healing of the cartilage only in the histology and not in the macroscopic appearance could be that our follow-up period of only 3 months was too short to alter the macroscopic appearance of the cartilage.Our results did not show a correlation between recessed plugs and short-term cartilage healing. This is in agreement with a study conducted with rabbits,[42] in which marginally recessed plugs did not adversely affect outcomes. Also, during a sheep study,[43] it was observed that plugs recessed up to 1 mm had a good survival rate. However, our healing results are limited to a short-term period; the longer term effects of recessed plugs were not studied.The surgeon found that it was difficult to hold the position and orientation of the tool according to the optoelectronically guided computer display during impacts. This was likely because the hardness of the sheep bone required more manipulation to harvest osteochondral plugs than would be necessary in normal human bone. On the other hand, the template-guided technique provided mechanical support to the surgical tools because the template was stabilized in the registered position using wires. This might explain the better surface reconstruction for the template-guided group. We speculate that the problem of holding the optically guided chisel will not arise in humanpatients unless the subchondral bone was sclerotic. We did not find any significant differences in ICRS II score in the surrounding cartilage between the template-guided and conventional techniques, which suggests that the placement and fixation of the templates did not have a negative effect on the tissue in the first 3 months postoperatively.The 3-month post-reconstruction micro-CT images revealed a significantly larger cyst volume in the conventional group than in the template-guided group. Subchondral cysts can result in serious complications, such as a collapse of the graft into the recipient hole. Cyst creation during healing might be larger for the conventional technique because of the significantly higher percentage of proud plugs found in the conventional technique. This is consistent with results from Pearce et al.,[28] who observed in a study significantly more cyst creation with plugs placed proud with respect to the articular surface than with flush placed plugs. Those authors suggested that micromotion of the plugs could lead to synovial fluid penetrating normal subchondral bone, which in turn might predispose the development of subchondral cysts. Tytherleigh-Strong et al.[44] also discussed the possibility of synovial fluid penetration into the gap between the graft and the surrounding, normal subchondral bone to create subchondral cysts. The template-guided technique may provide a more tightly fitting plug, reducing the penetration of synovial fluid, because the drill guide is held more rigidly inside the template during the preparation of the recipient site. With the conventional and optoelectronically guided techniques, the tools are handheld, without external support, and can result in a hole that is less cylindrical.We found no significant difference in measures of pain, lameness, weight, and macroscopic scores after healing between the optically guided and conventional techniques. This suggests that the invasive attachment of the sensors to the femur was well tolerated. However, we found a significant increase of the length and intensity of postoperative pain in the template-guided group compared with the optically guided group. This difference might be explained by the reduced invasiveness of the optically guided technique: although the arthotomies in both groups were the same, the insertion of the template required more soft tissue retraction, which could have caused more capulitis and synovitis.
Conclusions
The small number of sheep limited the statistical power of the measures obtained from the study. Nonetheless, statistical significance was found for a number of important measures. This is, to our knowledge, the first in vivo study to investigate the clinical outcome of image-guided autologous osteochondral transplantation in comparison to the conventional surgical method. The planning for the image-guided techniques required that an operator use a computer interface to place virtual plugs on a model of the patient’s bone and cartilage. The planning process took 30 to 45 minutes per procedure and required that the surgeon estimate, on the computer screen, the desired 3D articular cartilage surface over the defect. To improve the planning process, we developed, subsequent to this study, fully automatic planning methods.In conclusion, this in vivo animal study has shown that image-guided techniques produce better morphological and healing outcomes for autologous osteochondral transplantation compared with conventional surgical techniques. Further studies are necessary to confirm that this short-term improvement will translates to a better long-term clinical outcome. However, we believe that the results of this study show a great promise that computer-assisted mosaic arthroplasty can improve the clinical outcome not only in an animal model but also in patients.