Literature DB >> 26057528

Cup Implant Planning Based on 2-D/3-D Radiographic Pelvis Reconstruction-First Clinical Results.

Steffen Schumann, Yoshinobu Sato, Yuki Nakanishi, Futoshi Yokota, Masaki Takao, Nobuhiko Sugano, Guoyan Zheng.   

Abstract

GOAL: In the following, we will present a newly developed X-ray calibration phantom and its integration for 2-D/3-D pelvis reconstruction and subsequent automatic cup planning. Two different planning strategies were applied and evaluated with clinical data.
METHODS: Two different cup planning methods were investigated: The first planning strategy is based on a combined pelvis and cup statistical atlas. Thereby, the pelvis part of the combined atlas is matched to the reconstructed pelvis model, resulting in an optimized cup planning. The second planning strategy analyzes the morphology of the reconstructed pelvis model to determine the best fitting cup implant.
RESULTS: The first planning strategy was compared to 3-D CT-based planning. Digitally reconstructed radiographs of THA patients with differently severe pathologies were used to evaluate the accuracy of predicting the cup size and position. Within a discrepancy of one cup size, the size was correctly identified in 100% of the cases for Crowe type I datasets and in 77.8% of the cases for Crowe type II, III, and IV datasets. The second planning strategy was analyzed with respect to the eventually implanted cup size. In seven patients, the estimated cup diameter was correct within one cup size, while the estimation for the remaining five patients differed by two cup sizes.
CONCLUSION: While both planning strategies showed the same prediction rate with a discrepancy of one cup size (87.5%), the prediction of the exact cup size was increased for the statistical atlas-based strategy (56%) in contrast to the anatomically driven approach (37.5%). SIGNIFICANCE: The proposed approach demonstrated the clinical validity of using 2-D/3-D reconstruction technique for cup planning.

Entities:  

Mesh:

Year:  2015        PMID: 26057528     DOI: 10.1109/TBME.2015.2441378

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Fully automatic reconstruction of personalized 3D volumes of the proximal femur from 2D X-ray images.

Authors:  Weimin Yu; Chengwen Chu; Moritz Tannast; Guoyan Zheng
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-04-02       Impact factor: 2.924

2.  A novel technology for 3D knee prosthesis planning and treatment evaluation using 2D X-ray radiographs: a clinical evaluation.

Authors:  Guoyan Zheng; Hagen Hommel; Alper Akcoltekin; Benedikt Thelen; Jan Stifter; Geert Peersman
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-05-21       Impact factor: 2.924

3.  Surgical resection and reconstructive techniques using autologous femoral head bone-grafting in treating partial acetabular defects arising from primary pelvic malignant tumors.

Authors:  Wei Sun; Pengfei Zan; Xiaojun Ma; Yingqi Hua; Jiakang Shen; Zhengdong Cai
Journal:  BMC Cancer       Date:  2019-10-18       Impact factor: 4.430

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.