Literature DB >> 26044840

Evaluation of functionality and biological response of the multilayer flow modulator in porcine animal models.

Sherif Sultan1, Edel P Kavanagh, Niamh Hynes, Edward B Diethrich.   

Abstract

BACKGROUND: This study outlines the use of non-aneurysmal porcine animal models to study device functionality and biological response of the Multilayer Flow Modulator (MFM) (Cardiatis, Isnes, Belgium), with an emphasis on preclinical device functionality and biological response characteristics in an otherwise healthy aorta.
METHODS: Twelve animals were implanted with the study device in the abdominal aorta, in 6 animals for 1 month and 6 animals for 6 months. Upon completion of the study period, each animal underwent a necropsy to examine how the implanted device had affected the artery and surrounding tissue. Neointima and stenosis formation were recorded via morphometry, and endothelialization via histopathological analysis.
RESULTS: The MFM devices were delivered to their respective implantation sites without difficulty. Six of the implanted stents were oversized with percentages ranging from 2.6% to 18.8%. Statistical analysis was carried out and showed no significance between the regular sized stent group and oversized stent group for neointimal area (P=0.17), neointimal thickness (P=0.17), and percentage area stenosis (P=0.65). Histopathological findings showed in most areas flattened endothelium like cells lined the luminal surface of the neointima. Scanning electron microscopy also showed the devices were well tolerated, inciting only a minimal neointimal covering and little fibrin or platelet deposition. Neointimal thickness of 239.7±55.6 μm and 318.3±130.4 μm, and percentage area stenosis of 9.6±2.6% and 12.6±5% were recorded at 1 and 6 months respectively. No statistical differences were found between these results.
CONCLUSION: The MFM devices were delivered to their respective implantation sites without difficulty and incited little neointimal and stenosis formation in the aorta, affirming its functionality and biocompatibility.

Entities:  

Mesh:

Year:  2015        PMID: 26044840

Source DB:  PubMed          Journal:  Int Angiol        ISSN: 0392-9590            Impact factor:   2.789


  3 in total

1.  Multilayer flow modulator enhances vital organ perfusion in patients with type B aortic dissection.

Authors:  Farhad Rikhtegar Nezami; Lambros S Athanasiou; Junedh M Amrute; Elazer R Edelman
Journal:  Am J Physiol Heart Circ Physiol       Date:  2018-08-10       Impact factor: 4.733

2.  Packing Technique with or without Remodeling for Endovascular Coil Embolization of Renal Artery Aneurysms: Safety, Efficacy and Mid-Term Outcomes.

Authors:  Grégory Secco; Olivier Chevallier; Nicolas Falvo; Kévin Guillen; Pierre-Olivier Comby; Christiane Mousson; Nabil Majbri; Marco Midulla; Romaric Loffroy
Journal:  J Clin Med       Date:  2021-01-17       Impact factor: 4.241

3.  Endovascular management of renal artery aneurysms using the multilayer flow modulator.

Authors:  Sherif Sultan; Mahmoud Basuoniy Alawy; Rita Flaherty; Edel P Kavanagh; Mohamed Elsherif; Ala Elhelali; Florian Stefanov; Violet Lundon; Niamh Hynes
Journal:  Open Heart       Date:  2016-03-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.