Literature DB >> 26044535

The taper disaster--how could it happen?

Michael M Morlock1.   

Abstract

Corrosion of metallic implants in contact with body fluids is unavoidable, especially at interfaces where movement occurs or in gaps. Corrosion became clinically relevant with the introduction of large modular metal-on-metal total hip joint articulations (MoM THA) early in the 21st century. This review attempts to summarise the scientific knowledge about taper problems available at the time of introduction of these bearings, why this "disaster" could happen. It is speculated that changes to the taper connection made in the 1990s to increase the range of motion with small heads (28 and 32 mm) reduced the mechanical strength of this connection, which did not matter for small heads. With the use of large and very large metal heads in MoM articulations, which have a larger lever arm and can generate high friction in unfavourable situations, suddenly the taper interface exhibited corrosion problems on a previously unknown scale. It is speculated that due to the higher mechanical loading with larger heads, the taper connection became less forgiving with respect to assembly conditions, contamination, manufacturing tolerances and other factors, which are yet not known. Since no major clinical problems had been reported before the introduction of these bearings and the pre-clinical testing was very successful, the disaster took its course. The patient-implant-surgeon system is a very complex intrinsically hazardous system. Pre-clinical testing addresses few and defined factors and such, good results cannot be directly transferred to the clinical reality. A controlled stepwise introduction of innovations is required.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26044535     DOI: 10.5301/hipint.5000269

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  11 in total

1.  Revision for taper corrosion at the head-neck junction: pearls and pitfalls.

Authors:  Brian J McGrory; Brigham R McKenney
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

Review 2.  Corrosion of the Head-Stem Taper Junction-Are We on the Verge of an Epidemic?: Review Article.

Authors:  Michael Morlock; Dennis Bünte; Julian Gührs; Nicholas Bishop
Journal:  HSS J       Date:  2016-09-20

Review 3.  [Endoprostheses in the elderly : Biomaterials, implant selection and fixation technique].

Authors:  M M Morlock; M Jäger
Journal:  Orthopade       Date:  2017-01       Impact factor: 1.087

4.  Fretting-corrosion in hip taper modular junctions: The influence of topography and pH levels - An in-vitro study.

Authors:  Dmitry Royhman; Robin Pourzal; Deborah Hall; Hannah J Lundberg; Markus A Wimmer; Joshua Jacobs; Nadim J Hallab; Mathew T Mathew
Journal:  J Mech Behav Biomed Mater       Date:  2021-03-12

5.  Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability.

Authors:  A R MacLeod; N P T Sullivan; M R Whitehouse; H S Gill
Journal:  Bone Joint Res       Date:  2016-08       Impact factor: 5.853

Review 6.  What the Surgeon Can Do to Reduce the Risk of Trunnionosis in Hip Arthroplasty: Recommendations from the Literature.

Authors:  Claude B Rieker; Peter Wahl
Journal:  Materials (Basel)       Date:  2020-04-21       Impact factor: 3.623

7.  Blood cytokine pattern and clinical outcome in knee arthroplasty patients: comparative analysis 5 years after standard versus "hypoallergenic" surface coated prosthesis implantation.

Authors:  Peter Thomas; Philipp Hisgen; Hartmuth Kiefer; Ulf Schmerwitz; Andreas Ottersbach; Dominique Albrecht; Burkhard Summer; Christian Schinkel
Journal:  Acta Orthop       Date:  2018-10-29       Impact factor: 3.717

8.  Pseudotumours, cobalt and clinical outcome in small head metal-on-metal versus conventional metal-on-polyethylene total hip arthroplasty.

Authors:  Hugo C van der Veen; Inge Hf Reininga; Wierd P Zijlstra; Martijn F Boomsma; Sjoerd K Bulstra; Jos Jam van Raay
Journal:  Hip Int       Date:  2019-03-06       Impact factor: 2.135

9.  Fatal vascular complication due to metallosis.

Authors:  Miguel García-Gimeno; Ignacio Puyuelo Jarne; Salvador Tagarro-Villalba; Borja Antón Capitán; Consuelo Casco Aguilar; Emma González-González
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-04-15

10.  Model-Based Roentgen Stereophotogrammetric Analysis to Monitor the Head-Taper Junction in Total Hip Arthroplasty in Vivo-And They Do Move.

Authors:  Jing Xu; Robert Sonntag; J Philippe Kretzer; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

View more