Literature DB >> 26033921

The Development of a Continuous Intravascular Glucose Monitoring Sensor.

Barry C Crane1, Nicholas P Barwell2, Palepu Gopal3, Mannam Gopichand4, Timothy Higgs5, Tony D James6, Christopher M Jones2, Alasdair Mackenzie2, Krishna Prasad Mulavisala3, William Paterson2.   

Abstract

BACKGROUND: Glycemic control in hospital intensive care units (ICU) has been the subject of numerous research publications and debate over the past 2 decades. There have been multiple studies showing the benefit of ICU glucose control in reducing both morbidity and mortality. GlySure Ltd has developed a glucose monitor based on a diboronic acid receptor that can continuously measure plasma glucose concentrations directly in a patient's vascular system. The goal of this study was to validate the performance of the GlySure CIGM system in different patient populations.
METHODS: The GlySure Continuous Intravascular Glucose Monitoring (CIGM) System was evaluated in both the Cardiac ICU (33 patients) and MICU setting (14 patients). The sensor was placed through a custom CVC and measured the patients' blood glucose concentration every 15 seconds. Comparison blood samples were taken at 2 hourly then 4 hourly intervals and measured on a YSI 2300 STAT Plus or an i-STAT.
RESULTS: Consensus error grid analysis of the data shows that the majority of the data (88.2% Cardiac, and 95.0% MICU) fell within zone A, which is considered to be clinically accurate and all data points fell within zones A and B. The MARD of the Cardiac trial was 9.90% and the MICU trial had a MARD of 7.95%. Data analysis showed no significant differences between data generated from Cardiac and MICU patients or by time or glucose concentration.
CONCLUSIONS: The GlySure CIGM System has met the design challenges of measuring intravascular glucose concentrations in critically ill patients with acceptable safety and performance criteria and without disrupting current clinical practice. The accuracy of the data is not affected by the patients' condition.
© 2015 Diabetes Technology Society.

Entities:  

Keywords:  boronic acid; continuous glucose; intravascular

Mesh:

Substances:

Year:  2015        PMID: 26033921      PMCID: PMC4525654          DOI: 10.1177/1932296815587937

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  26 in total

Review 1.  Preventing complications of central venous catheterization.

Authors:  David C McGee; Michael K Gould
Journal:  N Engl J Med       Date:  2003-03-20       Impact factor: 91.245

2.  Could susceptibility to low hematocrit interference have compromised the results of the NICE-SUGAR trial?

Authors:  George S Cembrowski; David V Tran; Linda Slater-Maclean; Dat Chin; R T Noel Gibney; Michael Jacka
Journal:  Clin Chem       Date:  2010-04-29       Impact factor: 8.327

3.  Blood glucose control in 2010: 110 to 150 mg/dL and minimal variability.

Authors:  Jean-Louis Vincent
Journal:  Crit Care Med       Date:  2010-03       Impact factor: 7.598

4.  Tight glucose control in critically ill patients: should glucose meters be used?

Authors:  David B Sacks
Journal:  Clin Chem       Date:  2009-06-25       Impact factor: 8.327

5.  Variability of blood glucose concentration and short-term mortality in critically ill patients.

Authors:  Moritoki Egi; Rinaldo Bellomo; Edward Stachowski; Craig J French; Graeme Hart
Journal:  Anesthesiology       Date:  2006-08       Impact factor: 7.892

6.  No cerebrocervical venous congestion in patients with multiple sclerosis.

Authors:  Florian Doepp; Friedemann Paul; José M Valdueza; Klaus Schmierer; Stephan J Schreiber
Journal:  Ann Neurol       Date:  2010-08       Impact factor: 10.422

7.  Modular fluorescence sensors for saccharides.

Authors:  S Arimori; M L Bell; C S Oh; K A Frimat; T D James
Journal:  Chem Commun (Camb)       Date:  2001-09-21       Impact factor: 6.222

8.  Understanding low sugar from NICE-SUGAR.

Authors:  Craig French; Forbes McGain
Journal:  N Engl J Med       Date:  2012-12-20       Impact factor: 91.245

9.  Effect of an intensive glucose management protocol on the mortality of critically ill adult patients.

Authors:  James Stephen Krinsley
Journal:  Mayo Clin Proc       Date:  2004-08       Impact factor: 7.616

10.  Intensive insulin therapy and pentastarch resuscitation in severe sepsis.

Authors:  Frank M Brunkhorst; Christoph Engel; Frank Bloos; Andreas Meier-Hellmann; Max Ragaller; Norbert Weiler; Onnen Moerer; Matthias Gruendling; Michael Oppert; Stefan Grond; Derk Olthoff; Ulrich Jaschinski; Stefan John; Rolf Rossaint; Tobias Welte; Martin Schaefer; Peter Kern; Evelyn Kuhnt; Michael Kiehntopf; Christiane Hartog; Charles Natanson; Markus Loeffler; Konrad Reinhart
Journal:  N Engl J Med       Date:  2008-01-10       Impact factor: 91.245

View more
  19 in total

1.  Autoregressive Modeling of Drift and Random Error to Characterize a Continuous Intravascular Glucose Monitoring Sensor.

Authors:  Tony Zhou; Jennifer L Dickson; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2017-07-14

Review 2.  Designing the Glucose Telemetry for Hospital Management: From Bedside to the Nursing Station.

Authors:  Medha Satyarengga; Tariq Siddiqui; Elias K Spanakis
Journal:  Curr Diab Rep       Date:  2018-08-29       Impact factor: 4.810

3.  In Silico Testing of an Artificial-Intelligence-Based Artificial Pancreas Designed for Use in the Intensive Care Unit Setting.

Authors:  Leon DeJournett; Jeremy DeJournett
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

4.  Glucose Management Technologies for the Critically Ill.

Authors:  Pedro D Salinas; Carlos E Mendez
Journal:  J Diabetes Sci Technol       Date:  2019-01-13

Review 5.  Vascular Glucose Sensor Symposium: Continuous Glucose Monitoring Systems (CGMS) for Hospitalized and Ambulatory Patients at Risk for Hyperglycemia, Hypoglycemia, and Glycemic Variability.

Authors:  Jeffrey I Joseph; Marc C Torjman; Paul J Strasma
Journal:  J Diabetes Sci Technol       Date:  2015-06-15

Review 6.  Molecular Probes, Chemosensors, and Nanosensors for Optical Detection of Biorelevant Molecules and Ions in Aqueous Media and Biofluids.

Authors:  Joana Krämer; Rui Kang; Laura M Grimm; Luisa De Cola; Pierre Picchetti; Frank Biedermann
Journal:  Chem Rev       Date:  2022-01-07       Impact factor: 60.622

7.  Continuous Glucose Monitors and Automated Insulin Dosing Systems in the Hospital Consensus Guideline.

Authors:  Rodolfo J Galindo; Guillermo E Umpierrez; Robert J Rushakoff; Ananda Basu; Suzanne Lohnes; James H Nichols; Elias K Spanakis; Juan Espinoza; Nadine E Palermo; Dessa Garnett Awadjie; Leigh Bak; Bruce Buckingham; Curtiss B Cook; Guido Freckmann; Lutz Heinemann; Roman Hovorka; Nestoras Mathioudakis; Tonya Newman; David N O'Neal; Michaela Rickert; David B Sacks; Jane Jeffrie Seley; Amisha Wallia; Trisha Shang; Jennifer Y Zhang; Julia Han; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2020-09-28

8.  Glucose-sensitive silicone hydrogel contact lens toward tear glucose monitoring.

Authors:  Ramachandram Badugu; Edward Albert Reece; Joseph R Lakowicz
Journal:  J Biomed Opt       Date:  2018-05       Impact factor: 3.170

9.  Continuous Glucose Monitoring Measures Can Be Used for Glycemic Control in the ICU: An In-Silico Study.

Authors:  Tony Zhou; Jennifer L Dickson; Geoffrey M Shaw; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2017-11-06

Review 10.  Self and directed assembly: people and molecules.

Authors:  Tony D James
Journal:  Beilstein J Org Chem       Date:  2016-03-01       Impact factor: 2.883

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.