A moving blocker-based strategy for simultaneous megavoltage and kilovoltage scatter correction in cone-beam computed tomography image acquired during volumetric modulated arc therapy.

Luo Ouyang1, Huichen Pam Lee1, Jing Wang2.   

Abstract

PURPOSE: To evaluate a moving blocker-based approach in estimating and correcting megavoltage (MV) and kilovoltage (kV) scatter contamination in kV cone-beam computed tomography (CBCT) acquired during volumetric modulated arc therapy (VMAT). METHODS AND MATERIALS: During the concurrent CBCT/VMAT acquisition, a physical attenuator (i.e., "blocker") consisting of equally spaced lead strips was mounted and moved constantly between the CBCT source and patient. Both kV and MV scatter signals were estimated from the blocked region of the imaging panel, and interpolated into the unblocked region. A scatter corrected CBCT was then reconstructed from the unblocked projections after scatter subtraction using an iterative image reconstruction algorithm based on constraint optimization. Experimental studies were performed on a Catphan® phantom and an anthropomorphic pelvis phantom to demonstrate the feasibility of using a moving blocker for kV-MV scatter correction.
RESULTS: Scatter induced cupping artifacts were substantially reduced in the moving blocker corrected CBCT images. Quantitatively, the root mean square error of Hounsfield units (HU) in seven density inserts of the Catphan phantom was reduced from 395 to 40.
CONCLUSIONS: The proposed moving blocker strategy greatly improves the image quality of CBCT acquired with concurrent VMAT by reducing the kV-MV scatter induced HU inaccuracy and cupping artifacts.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  CBCT; Hounsfield unit accuracy; Megavoltage scatter correction; Moving blocker; VMAT

Mesh:

Year:  2015        PMID: 26026484      PMCID: PMC4508229          DOI: 10.1016/j.radonc.2015.04.015

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  31 in total

1.  Improved scatter correction using adaptive scatter kernel superposition.

Authors:  M Sun; J M Star-Lack
Journal:  Phys Med Biol       Date:  2010-10-28       Impact factor: 3.609

2.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  A quality assurance program for image quality of cone-beam CT guidance in radiation therapy.

Authors:  Jean-Pierre Bissonnette; Douglas J Moseley; David A Jaffray
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

4.  Scatter correction for cone-beam CT in radiation therapy.

Authors:  Lei Zhu; Yaoqin Xie; Jing Wang; Lei Xing
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

5.  A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiation.

Authors:  Marcel van Herk; Lennert Ploeger; Jan-Jakob Sonke
Journal:  Radiother Oncol       Date:  2011-09-15       Impact factor: 6.280

6.  Extraction of primary signal from EPIDs using only forward convolution.

Authors:  V N Hansen; W Swindell; P M Evans
Journal:  Med Phys       Date:  1997-09       Impact factor: 4.071

7.  Towards the clinical implementation of iterative low-dose cone-beam CT reconstruction in image-guided radiation therapy: cone/ring artifact correction and multiple GPU implementation.

Authors:  Hao Yan; Xiaoyu Wang; Feng Shi; Ti Bai; Michael Folkerts; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  A megavoltage scatter correction technique for cone-beam CT images acquired during VMAT delivery.

Authors:  C J Boylan; T E Marchant; J Stratford; J Malik; A Choudhury; R Shrimali; J Rodgers; C G Rowbottom
Journal:  Phys Med Biol       Date:  2012-05-23       Impact factor: 3.609

9.  A study on adaptive IMRT treatment planning using kV cone-beam CT.

Authors:  George X Ding; Dennis M Duggan; Charles W Coffey; Matthew Deeley; Dennis E Hallahan; Anthony Cmelak; Arnold Malcolm
Journal:  Radiother Oncol       Date:  2007-08-20       Impact factor: 6.280

Review 10.  Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review.

Authors:  Kavitha Srinivasan; Mohammad Mohammadi; Justin Shepherd
Journal:  Pol J Radiol       Date:  2014-07-03
View more
  1 in total

1.  Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study.

Authors:  Jihun Kim; Ki Chang Keum; Ho Lee; Chae-Seon Hong; Kwangwoo Park; Jin Sung Kim
Journal:  Radiat Oncol       Date:  2020-09-25       Impact factor: 3.481

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.