| Literature DB >> 26024887 |
Keren Kang1,2, Emmanuel E Dzakah3, Yongping Huang4, Mingquan Xie5, Xiaochun Luo6, Wenmei Li7, Jihua Wang8.
Abstract
BACKGROUND: The low sensitivity and specificity of Plasmodium falciparum diagnostic tests pose a serious health threat to people living in endemic areas. The objective of the study was to develop a rapid assay for the detection of histidine-rich protein 2 (HRP2) of P. falciparum in whole blood by immunofluorescence chromatographic technology.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26024887 PMCID: PMC4475284 DOI: 10.1186/s12936-015-0740-1
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 2.979
Fig. 1FinecareTM FIA meter. The Finecare™ FIA Meter is a portable instrument for fluorescence detection of various analytes in blood or urine. The meter equipped with a built-in test cassette holder and printer
Screening of Plasmodium falciparum control sample panel
| Samples | Derivation of sample | Results | S/CO |
|---|---|---|---|
| Benin I 200 | Benin, Africa | Positive | 3.14 |
| Benin I 2000 | Benin, Africa | Positive | 9.12 |
| Santa Lucia 200 | El Salvador, Central America | Positive | 3.06 |
| Santa Lucia 2000 | El Salvador, Central America | Positive | 6.57 |
| Nigeria XII 200 | Nigeria, Africa | Positive | 3.52 |
| Nigeria XII 2000 | Nigeria, Africa | Positive | 7.76 |
| FC27/A3 200 | Papua New Guinea, Australia | Positive | 2.83 |
| FC27/A3 2000 | Papua New Guinea, Australia | Positive | 5.98 |
| PH1 200 | Philippines, Southeast Asia | Positive | 2.54 |
| PH1 2000 | Philippines, Southeast Asia | Positive | 4.72 |
S/CO indicates the ratio of detected value to the cut-off value. S/CO < 1 is negative, S/CO ≧ 1 is positive. The greater the S/CO values, the higher the HRP2 concentration
Fig. 2Testing of negative and positive HRP 2 samples. a and b Negative HRP2 sample test. This represents the test cassette in fluorescence excitation mode in the testing of a negative HRP2 sample (a). The absence of no target antigen in the sample is represented by the “T zone” having no excitation peak and the “C zone” showed a fluorescence peak (b). c and d Positive HRP2 sample test. The test cassette showed band detections at both T and C zones for positive samples (c) and the excitation in these regions are detected by the FIA and represented by the two peaks at T and C zones
Precision performance of the immunofluorescence test kit
| Time | Lot 1 | Lot 2 | Lot with lot CV (%) | ||
|---|---|---|---|---|---|
| Result | S/CO | Result | S/CO | ||
| 1 | Positive | 1.75 | Positive | 1.81 | 6.04 |
| 2 | Positive | 1.69 | Positive | 1.89 | |
| 3 | Positive | 1.60 | Positive | 1.98 | |
| 4 | Positive | 1.70 | Positive | 1.91 | |
| 5 | Positive | 1.73 | Positive | 1.88 | |
| 6 | Positive | 1.69 | Positive | 1.95 | |
| 7 | Positive | 1.78 | Positive | 1.85 | |
| 8 | Positive | 1.66 | Positive | 1.87 | |
| 9 | Positive | 1.77 | Positive | 1.71 | |
| 10 | Positive | 1.64 | Positive | 1.82 | |
| Within lot CV (%) | 3.39 | 4.10 | |||
S/CO indicates the ratio of detected value to the cut-off value. S/CO < 1 is negative, S/CO > 1 is positive. The greater the S/CO values, the higher the HRP2 concentration
Performance of test kit on interfering substances
| Interfering substance | Conc. | Cases | Negative reference samples | Positive reference samples | ||
|---|---|---|---|---|---|---|
| Criterion | Results | Criterion | Results | |||
| Bilirubin | 2 g/L | 20 | 10/10−/− | 10/10−/− | 10/10+/+ | 10/10+/+ |
| Triglycerides | 30 g/L | 20 | 10/10−/− | 10/10−/− | 10/10+/+ | 10/10+/+ |
| Cholesterol | 15 g/L | 20 | 10/10−/− | 10/10−/− | 10/10+/+ | 10/10+/+ |
| Rheumatoid factor | 320 IU/ml | 20 | 10/10−/− | 10/10−/− | 10/10+/+ | 10/10+/+ |
10/10−/− indicates that all negative samples were tested negative for ten repeated times, and 10/10+/+ indicates that all positive samples were tested positive for ten repeated times
Cross-reactivity determination
| Sample type | Results | |
|---|---|---|
| Malaria samples |
| Negative (4/4−/−) |
|
| Negative (10/10−/−) | |
|
| Negative (10/10−/−) | |
| Hepatitis B virus positive specimens | Negative (10/10−/−) | |
| Hepatitis C virus positive specimens | Negative (10/10−/−) | |
| Human Immunodeficiency Virus positive specimens | Negative (10/10−/−) | |
|
| Negative (10/10−/−) | |
4/4−/− indicates that four cases of P. malariae specimens were tested negative; 10/10−/− indicates that ten cases corresponding cross-reactive specimens were tested negative
Stability test and analysis
| Days | Internal quality control panel | |||||
|---|---|---|---|---|---|---|
| Positive reference samples (low concentration) | Precision reference samples | Negative reference samples | Positive reference samples | |||
| L1 | L2 | L3 | (J) | N1 to N10 | P1 to P10 | |
| 1 | − | + | + | 10/10+/+ | 10/10−/− | 10/10+/+ |
| 7 | − | + | + | 10/10+/+ | 10/10−/− | 10/10+/+ |
| 15 | − | + | + | 10/10+/+ | 10/10−/− | 10/10+/+ |
| 22 | − | + | + | 10/10+/+ | 10/10−/− | 10/10+/+ |
| 30 | − | + | + | 10/10+/+ | 10/10−/− | 10/10+/+ |
“+” indicates positive; “-” indicates negative
Fluorescent HRP2 diagnostic test and reference kit in the screening of clinical samples
| Developed kit | Total | |||
|---|---|---|---|---|
| Positive | Negative | |||
| Reference kit | Positive | 360 | 0 | 360 |
| Negative | 3 | 800 | 803 | |
| Total | 363 | 800 | 1,163 | |
Compared with the reference kit, the positive coincidence rate of the two kits is 100 %(360/360), and the negative coincidence rate is 99.62 % (800/803)