Impact of the Static and Radiofrequency Magnetic Fields Produced by a 7T MR Imager on Metallic Dental Materials.

Kenta Oriso1, Takuya Kobayashi, Makoto Sasaki, Ikuko Uwano, Hidemichi Kihara, Hisatomo Kondo.   

Abstract

PURPOSE: We examined safety issues related to the presence of various metallic dental materials in magnetic resonance (MR) imaging at 7 tesla.
METHODS: A 7T MR imaging scanner was used to examine 18 kinds of materials, including 8 metals used in dental restorations, 6 osseointegrated dental implants, 2 abutments for dental implants, and 2 magnetic attachment keepers. We assessed translational attraction forces between the static magnetic field and materials via deflection angles read on a tailor-made instrument and compared with those at 3T. Heating effects from radiofrequency during image acquisitions using 6 different sequences were examined by measuring associated temperature changes in agarose-gel phantoms with a fiber-optic thermometer.
RESULTS: Deflection angles of the metallic dental materials were significantly larger at 7T than 3T. Among full metal crowns (FMCs), deflection angles were 18.0° for cobalt-chromium (Co-Cr) alloys, 13.5° for nickel-chromium (Ni-Cr) alloys, and 0° for other materials. Deflection angles of the dental implants and abutments were minimal, ranging from 5.0 to 6.5°, whereas the magnetic attachment keepers were strongly attracted to the field, having deflection angles of 90° or more. Increases in temperature of the FMCs were significant but less than 1°C in every sequence. The dental implant of 50-mm length showed significant but mild temperature increases (up to 1.5°C) when compared with other dental implants and abutments, particularly on sequences with high specific absorption rate values.
CONCLUSION: Although most metallic dental materials showed no apparent translational attraction or heating at 7T, substantial attraction forces on the magnetic attachment keepers suggested potential risks to patients and research participants undergoing MR imaging examinations.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25994037     DOI: 10.2463/mrms.2014-0122

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  8 in total

1.  Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3.0-T MRI: Author response.

Authors:  Serkan Görgülü; Simel Ayyıldız; Kıvanç Kamburoğlu; Sıla Gökçe; Tuncer Ozen
Journal:  Dentomaxillofac Radiol       Date:  2015-10-15       Impact factor: 2.419

Review 2.  [Safety of implants in high field and ultrahigh field MRI].

Authors:  O Kraff; H H Quick
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

3.  [Metal objects of the head and neck region in magnetic resonance imaging : Survey among radiologists].

Authors:  Ulrike Kielburg; Felix H Blankenstein
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

4.  [Comparison of the two magnetic resonance deartifact techniques in imaging of different porcelain-fused-to-metal crowns].

Authors:  Wen-Jin Li; Jin-Liang Niu; Li Zhu; Yu Wang; Ying An; Shuang-Yuan Zhang
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2019-02-01

5.  Magnetization and demagnetization of magnetic dental attachments in a 3-T MRI system.

Authors:  Norio Hayashi; Akio Ogura; Toshio Tsuchihashi; Daisuke Takahashi; Tsuyoshi Matsuda; Shinya Seino; Tsukasa Doi
Journal:  Radiol Phys Technol       Date:  2017-04-27

6.  Patient safety incidents in dentomaxillofacial imaging: reported adverse events from Hospital District Helsinki and Uusimaa and the City of Helsinki during 2012-2017.

Authors:  Marianne Suuronen; Taina Autti; Lasse Lehtonen
Journal:  Oral Radiol       Date:  2022-05-25       Impact factor: 1.852

7.  Microleakage beneath orthodontic brackets in high field magnetic resonance imaging (MRI) AT 1.5 & 3 Tesla.

Authors:  Esra Bolat Gümüş; Samed Şatir; Alper Kuştarci
Journal:  Dentomaxillofac Radiol       Date:  2022-02-09       Impact factor: 3.525

8.  Assessment of magnetic field interactions and radiofrequency-radiation-induced heating of metallic spinal implants in 7 T field.

Authors:  Itsuko Tsukimura; Hideki Murakami; Makoto Sasaki; Hirooki Endo; Daisuke Yamabe; Ryosuke Oikawa; Minoru Doita
Journal:  J Orthop Res       Date:  2017-03-08       Impact factor: 3.494

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.