Literature DB >> 25983310

A randomized study of remote monitoring and fluid monitoring for the management of patients with implanted cardiac arrhythmia devices.

Lars Lüthje1, Dirk Vollmann2, Joachim Seegers2, Christian Sohns2, Gerd Hasenfuß2, Markus Zabel2.   

Abstract

AIMS: Only limited comparative data exist on the benefits of fluid monitoring (FM) combined with remote monitoring (RM) regarding morbidity and mortality of heart failure (HF) patients. This prospective single-centre randomized pilot study aimed to estimate the influence of RM in combination with FM on HF hospitalizations as well as ventricular tachyarrhythmias and mortality. METHODS AND
RESULTS: Patients with standard indication for implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy and defibrillator were implanted with devices capable of RM and FM, and were followed for 15 months. Subjects were randomly allocated to RM including OptiVol and predefined management of alerts (remote group), or standard in-office visits every 3 months (control group). A total of 176 patients (77% male; 66 ± 12 years; left ventricular ejection fraction (LVEF) 32 ± 11%; ischemic cardiomyopathy 50%; CRT device 50%; primary prevention 85%) were analysed. Cox proportional hazard analysis on the time to first HF-related hospitalization showed a hazard ratio of 1.23 [0.62-2.44] (P = 0.551) favouring the control group. In the remote group, 13 patients (15%) experienced ICD shocks vs. 10 patients (11%) in the control group (P = 0.512). The average time to first ICD shock was 212 ± 173 days in the remote arm and 212 ± 143 days in the control arm (P = 0.994). The Kaplan-Meier estimate of mortality after 1 year was 8.6% (eight deaths) in the remote group vs. 4.6% in the control group (six deaths; P = 0.502).
CONCLUSION: In a single-centre randomized pilot study of RM in combination with FM, no significant influence on HF-related hospitalizations, ICD shocks, or mortality was found. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2015. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Cardiac decompensation; Fluid monitoring; Heart failure; OptiVol fluid index; Remote monitoring; Telemedicine

Mesh:

Year:  2015        PMID: 25983310     DOI: 10.1093/europace/euv039

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  13 in total

Review 1.  Remote monitoring of implantable cardiac devices in heart failure patients: a systematic review and meta-analysis of randomized controlled trials.

Authors:  Sultan Alotaibi; Jaime Hernandez-Montfort; Omar E Ali; Karim El-Chilali; Bernardo A Perez
Journal:  Heart Fail Rev       Date:  2020-05       Impact factor: 4.214

Review 2.  [Device-based remote monitoring : Current evidence].

Authors:  David Duncker; Roman Michalski; Johanna Müller-Leisse; Christos Zormpas; Thorben König; Christian Veltmann
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2017-08-15

Review 3.  Remote Monitoring in Heart Failure: the Current State.

Authors:  Rajeev C Mohan; J Thomas Heywood; Roy S Small
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-03

4.  Does remote patient monitoring reduce acute care use? A systematic review.

Authors:  Monica L Taylor; Emma E Thomas; Centaine L Snoswell; Anthony C Smith; Liam J Caffery
Journal:  BMJ Open       Date:  2021-03-02       Impact factor: 2.692

5.  mHealth education interventions in heart failure.

Authors:  Sabine Allida; Huiyun Du; Xiaoyue Xu; Roslyn Prichard; Sungwon Chang; Louise D Hickman; Patricia M Davidson; Sally C Inglis
Journal:  Cochrane Database Syst Rev       Date:  2020-07-02

6.  Bioimpedance Alerts from Cardiovascular Implantable Electronic Devices: Observational Study of Diagnostic Relevance and Clinical Outcomes.

Authors:  Christophe Jp Smeets; Julie Vranken; Jo Van der Auwera; Frederik H Verbrugge; Wilfried Mullens; Matthias Dupont; Lars Grieten; Hélène De Cannière; Dorien Lanssens; Thijs Vandenberk; Valerie Storms; Inge M Thijs; Pieter M Vandervoort
Journal:  J Med Internet Res       Date:  2017-11-23       Impact factor: 5.428

Review 7.  Clinical outcomes of digital sensor alerting systems in remote monitoring: a systematic review and meta-analysis.

Authors:  Fahad M Iqbal; Kyle Lam; Meera Joshi; Sadia Khan; Hutan Ashrafian; Ara Darzi
Journal:  NPJ Digit Med       Date:  2021-01-08

Review 8.  Factors influencing the effectiveness of remote patient monitoring interventions: a realist review.

Authors:  Emma E Thomas; Monica L Taylor; Annie Banbury; Centaine L Snoswell; Helen M Haydon; Victor M Gallegos Rejas; Anthony C Smith; Liam J Caffery
Journal:  BMJ Open       Date:  2021-08-25       Impact factor: 2.692

9.  Prediction of heart failure hospitalizations based on the direct measurement of intrathoracic impedance.

Authors:  Michael R Zile; Vinod Sharma; Catalin F Baicu; Jodi Koehler; Anthony S Tang
Journal:  ESC Heart Fail       Date:  2020-08-13

Review 10.  Remote Monitoring of Implantable Cardioverter-Defibrillators, Cardiac Resynchronization Therapy and Permanent Pacemakers: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2018-10-24