Modeling blurring effects due to continuous gantry rotation: Application to region of interest tomography.

Jeroen Cant1, Willem Jan Palenstijn2, Gert Behiels3, Jan Sijbers2.   

Abstract

PURPOSE: Projections acquired with continuous gantry rotation may suffer from blurring effects, depending on the rotation speed and the exposure time of each projection. This leads to blurred reconstructions if conventional reconstruction algorithms are applied. In this paper, the authors propose a reconstruction method for fast acquisitions based on a continuously moving and continuously emitting x-ray source. They study the trade-off between total acquisition time and reconstruction quality and compare with conventional reconstructions using projections acquired with a stepwise moving x-ray source.
METHODS: The authors introduce the algebraic reconstruction technique with angular integration concept, which models the angular integration due to the relative motion of the x-ray source during the projection.
RESULTS: Compared to conventional reconstruction from projections acquired with pulsed x-ray emission, the proposed method results in substantially improved reconstruction quality around the center of rotation. Outside this region, the proposed method results in improved radial resolution and a decreased tangential resolution. For a fixed reconstruction quality of this region of interest, the proposed method enables a lower number of projections and thus a faster acquisition.
CONCLUSIONS: The modeling of the continuous gantry rotation in the proposed method substantially improves the reconstruction quality in a region of interest around the rotation center. The proposed method shows potential for fast region of interest tomography.

Entities:  

Mesh:

Year:  2015        PMID: 25979069     DOI: 10.1118/1.4914422

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  High-Fidelity Modeling of Detector Lag and Gantry Motion in CT Reconstruction.

Authors:  Steven Tilley; Alejandro Sisniega; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Conf Proc Int Conf Image Form Xray Comput Tomogr       Date:  2018-05

2.  Effect of optical blurring of X-ray source on breast tomosynthesis image quality: Modulation transfer function, anatomical noise power spectrum, and signal detectability perspectives.

Authors:  Changwoo Lee; Jongduk Baek
Journal:  PLoS One       Date:  2022-05-19       Impact factor: 3.240

3.  Temporal sparsity exploiting nonlocal regularization for 4D computed tomography reconstruction.

Authors:  Daniil Kazantsev; Enyu Guo; Anders Kaestner; William R B Lionheart; Julian Bent; Philip J Withers; Peter D Lee
Journal:  J Xray Sci Technol       Date:  2016       Impact factor: 1.535

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.