Expulsion of Nova-T380, Multiload 375, and Copper-T380A contraceptive devices inserted during cesarean delivery.

Ahmed Ragab1, Hossam O Hamed2, Mohamed A Alsammani3, Hend Shalaby4, Hanan Nabeil4, Rafik Barakat4, Ahmed N Fetih5.   

Abstract

OBJECTIVE: To compare the expulsion rate of Nova-T380, Multiload 375, and Copper-T380A intrauterine contraceptive devices (IUCDs) inserted during cesarean delivery.
METHODS: A comparative randomized study was conducted between January 1, 2013, and June 30, 2014, in three maternity centers in Egypt and Saudi Arabia. All women scheduled for an elective cesarean and accepting intraoperative insertion of an IUCD were randomly allocated to receive the Nova-T380 (group 1), Multiload 375 (group 2), or Cu-T380A (group 3) using a computer-generated table. Researchers and participants were not masked to the type of IUCD. Follow-up was for 1 year. The primary outcome was IUCD expulsion (complete or partial [i.e. displacement]).
RESULTS: Each group contained 40 participants. At 1 year, expulsion had been reported for 5 (13%) women in group 1, 2 (5%) in group 2, and 6 (15%) in group 3 (P>0.05 for all). The frequency of displacement was significantly lower in group 2 (5 [13%] participants) than in group 1 (15 [38%]; P=0.001) and group 3 (14 [35%]; P=0.008).
CONCLUSION: Despite a comparable risk of expulsion following IUCD insertion during cesarean delivery, the Multiload 375 device showed the lowest risk of displacement.
Copyright © 2015 International Federation of Gynecology and Obstetrics. Published by Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Cesarean delivery; Copper-T380A; Intrauterine contraceptive device; Multiload 375; Nova-T380; Postpartum insertion

Mesh:

Year:  2015        PMID: 25975871     DOI: 10.1016/j.ijgo.2015.03.025

Source DB:  PubMed          Journal:  Int J Gynaecol Obstet        ISSN: 0020-7292            Impact factor:   3.561


  6 in total

1.  Expulsion of intrauterine devices after postpartum placement by timing of placement, delivery type, and intrauterine device type: a systematic review and meta-analysis.

Authors:  Sarah H Averbach; Yokabed Ermias; Gary Jeng; Kathryn M Curtis; Maura K Whiteman; Erin Berry-Bibee; Denise J Jamieson; Polly A Marchbanks; Naomi K Tepper; Tara C Jatlaoui
Journal:  Am J Obstet Gynecol       Date:  2020-03-03       Impact factor: 8.661

2.  A Decision Analysis Model of 1-Year Effectiveness of Intended Postplacental Compared With Intended Delayed Postpartum Intrauterine Device Insertion.

Authors:  Sarita Sonalkar; Tegan Hunter; Elizabeth P Gurney; Arden McAllister; Courtney A Schreiber
Journal:  Obstet Gynecol       Date:  2018-11       Impact factor: 7.661

3.  Intrauterine Device Expulsion After Postpartum Placement: A Systematic Review and Meta-analysis.

Authors:  Tara C Jatlaoui; Maura K Whiteman; Gary Jeng; Naomi K Tepper; Erin Berry-Bibee; Denise J Jamieson; Polly A Marchbanks; Kathryn M Curtis
Journal:  Obstet Gynecol       Date:  2018-10       Impact factor: 7.661

Review 4.  Intra-cesarean insertion and fixation of frameless intrauterine devices.

Authors:  Ateş Karateke; Abdulkadir Turgut; Özkan Özdamar; Dirk Wildemeersch
Journal:  Turk J Obstet Gynecol       Date:  2017-03-15

Review 5.  Insertion of intrauterine devices after cesarean section: a systematic review update.

Authors:  Norman D Goldstuck; Petrus S Steyn
Journal:  Int J Womens Health       Date:  2017-04-18

6.  Assessment of a Novel Fixation Method of a Frameless Intrauterine Contraceptive Device Inserted during Cesarean Delivery as a Means of Preventing Displacements and Expulsions: A Prospective Observational Study.

Authors:  Hazal Kutlucan; Recep Onur Karabacak; Stefanie De Buyser; Ahmet Erdem; Nuray Bozkurt; Erhan Demirdağ; Dirk Wildemeersch
Journal:  Life (Basel)       Date:  2022-01-07
  6 in total